SUTO S605 Manuel

English
Instruction and Operation Manual
S605
Portable Breathing Air Quality Analyzer

Dear Customer,
Thank you for choosing our product.
Before starting up the device, please read the manual in full and observe
the operating instructions stated. The manufacturer cannot be held liable
for any damage which occurs as a result of non-observance or non-
compliance with this manual.
Should the device be tampered with in any manner other than a
procedure which is described and specified in the manual, the warranty
is void and the manufacturer is e empt from liability.
The device is designed e clusively for the described application.
We offer no guarantee for the suitability for any other purpose. We are
also not liable for consequential damage resulting from the delivery,
capability or use of this device.
2 S605

Table o Contents
1 Safety Instructions......................................................................5
2 Registered Trademarks................................................................8
3 Application.................................................................................8
4 Features....................................................................................8
5 Technical Data............................................................................9
5.1 General Data........................................................................9
5.2 Operating Conditions.............................................................9
5.3 Measurement Data..............................................................10
6 Dimensional Drawings................................................................12
7 Installation on Site....................................................................13
7.1 Compressed Air Inlet...........................................................14
7.2 Compressed Air Connections.................................................15
7.2.1 Installation Requirements...............................................15
7.2.2 Installation...................................................................16
7.3 Electrical Connectors............................................................17
7.4 Electrical Connections..........................................................17
7.4.1 RS-485 Networking (Modbus/RTU)...................................17
7.4.2 TCP/IP Networking (Modbus/TCP)....................................18
7.4.3 Connect with a PC or an OTG USB Stick............................18
7.5 Water Filling Connector........................................................18
7.6 Calibration Connector ..........................................................19
8 Setup and Configuration.............................................................20
9 Operation.................................................................................21
9.1 Main Screen........................................................................23
9.1.1 Quick Buttons and Icons.................................................23
9.1.2 Icons in the Status Bar...................................................23
9.2 Value View..........................................................................24
9.3 Graphic View......................................................................24
9.4 Menu.................................................................................25
9.5 Configure S4A Remote Connection.........................................27
9.6 SIM Card Requirement for 4G/LTE Option...............................27
9.6.1 Data Traffic...................................................................27
9.6.2 PIN Code Protection.......................................................27
10 Guided Measurement...............................................................28
10.1 Steps for Guided Measurement............................................30
10.2 Reports for Guided Measurements.......................................32
S605 3

11 Firmware Updates...................................................................33
12 Calibration..............................................................................35
12.1 Calibration for DP Sensor and Oil Sensor...............................35
12.2 Calibration for O2, CO, CO2 Sensor......................................35
12.2.1 Calibration Gases.........................................................36
12.2.2 Calibration Procedure...................................................36
13 Optional Accessories................................................................42
14 Maintenance...........................................................................42
15 Disposal or Waste....................................................................42
16 Warranty................................................................................42
17 Appendi - Modbus Holding Register Table..................................43
17.1 Modbus Interface...............................................................43
17.2 Modbus Register Table........................................................44
4 S605

1 Safety Instructions
1 Sa ety Instructions
Please check i this instruction manual matches the
product type.
Please observe all notes and instructions indicated in this
manual. It contains essential information which must be
observed before and during installation, operation and
maintenance. Therefore this instruction manual must be read carefully
by the technician as well as by the responsible user / qualified
personnel.
This instruction manual must be available at the operation site of the
flow sensor at any time. In case of any obscurities or questions,
regarding this manual or the product, please contact the manufacturer.
WARNING!
Compressed air!
Any contact with quickly escaping air or bursting
parts o the compressed air system can lead to
serious injuries or even death!
•Do not e ceed the ma imum permitted pressure range (see
sensors label and/or the specifications written in this manual).
•Only use pressure tight installation material.
•Avoid that persons get hit by escaping air or bursting parts of the
instrument.
•The system must be pressure less during maintenance work.
WARNING!
Voltage used or supply!
Any contact with energized parts o the product
may lead to an electrical shock which can lead to
serious injuries or even death!
•Consider all regulations for electrical installations.
•The system must be disconnected from any power supply during
maintenance work.
•Do not e ceed the permitted operating parameters as specified in
the data sheet, this manual or written on the instrument.
•Make sure the product is operated in its permitted limitations.
S605 5

1 Safety Instructions
•Do not e ceed or undercut the permitted storage and operation
temperature, pressure and humidity.
•The product should be maintained and calibrated frequently, at
least annually. For more information, please contact the customer
support.
General sa ety instructions
•It is not allowed to use the product in e plosive areas.
•Please observe the national regulations before/during installation
and operation.
Remarks
•It is not allowed to disassemble the product.
•Always check the compressed air connectors in terms of stability
and tightness before any operation.
ATTENTION!
Measurement values can be a ected by mal unction!
The product must be installed properly and requently
maintained, otherwise it may lead to wrong
measurements, which can lead to wrong results.
•The instrument must be operated within or better than
compressed air quality Class 4.4.4 according to the ISO8573-1. If
the air quality cannot reach the specified class, the device may
show wrong measurement values or even may get damaged.
•Do not e ceed the ma imum operation temperature or pressure.
•Avoid condensation inside the instrument caused by the supplied
air or gas because it harms the instrument and affects the
accuracy.
•It is recommended that before any operation you e amine the
instrument in terms of the measurement accuracy and deviations.
To do this, you may use high efficiency filters for oil vapor and
particle, and the instrument should show low values close to zero.
Additionally, it is recommended you e amine the dew point
measurement to ensure it is working correctly. To do this, you may
use a dry gas with a defined dew point value which should be
reached within several minutes. Or you may use any reference
meter. Only after doing these e aminations, you can rely on the
measurement results after the instrument have been used for
other measurement cycles.
6 S605

1 Safety Instructions
•Always perform a visual inspection in prior to every measurement
before connecting compressed air. If any loose objects are found,
do not connect the compressed air because it may lead to a
damage or injuries.
Storage and transportation
•Make sure that the transportation temperature is between
-10... 50°C.
•For transportation it is recommended to use the packaging which
comes with the sensor.
•Please make sure that the storage temperature of the sensor is
between -10... 50°C.
•Avoid direct UV and solar radiation during storage.
•For the storage the humidity must be < 90%, no condensation.
ATTENTION!
Equipment may get damaged!
Please make sure, that your measuring point is
ree o excessive contamination/dirt. This should
maintained be ore every measurement.
•Observe the measuring point always before measurement if it is
free of contamination like water drops, oil drops or other rough
contaminants.
•Should water hit the inner electronics, the senors could be
seriously damaged.
•Any rough contamination of particles, water, oil or other impurities
may damage the instrument or affect the accuracy.
•Check your measurement point with the enclosed test kit, by
purging air in prior to the use of the instrument through the test
kit. A clean tissue may help to identify oil, water or any other
lubricants in the air.
S605 7

2 Registered Trademarks
2 Registered Trademarks
SUTO®Registered trademark of SUTO iTEC
MODBUS®Registered trademark of the Modbus Organization,
Hopkinton, USA
3 Application
The S605 Portable Breathing Air Quality Analyzer measures the content
of O2, CO, CO2, H2O and Oil vapor according to the breathing air purity
standards. It monitors the contaminant concentration to ensure that the
breathing air quality complies with international standards such as
EN 12021 or CFSR 1910.134(d).
The S605 is mainly used in an industrial environment. It can not be used
in hazardous or e plosive environment and it not suited for permanent
outside applications.
4 Features
•All-in-one device measures O2, CO, CO2, H2O, Oil, and pressure.
•Portable design makes it to be carried with one hand.
•The device has a DC power supply and additionally a battery which
can bridge power failures.
•High resolution 5” color touchscreen display.
•Software guided measurement.
•USB port for transferring data to a USB-OTG memory.
•Modbus/TCP and Modbus/RTU interfaces for communication.
•Data logger with the capacity of 30-million values.
•Only one gas inlet for all parameters.
•0.4...1.5 MPa(g) pressure at inlet; higher pressure is available
with e ternal pressure reducer.
•With optional 4G modem feature (P/N: A1670) and dedicated S4A
software, you can check the online measurement values, read out
the logging files remotely.
8 S605

5 Technical Data
5 Technical Data
5.1 General Data
Display Touchscreen, Size: 5“,
Resolution: 800 480 pi
Power supply Mains supply adapter (AC/DC)
Input: 100 ... 240 VAC, 50/60 Hz, 1.8 A
Output: 20 VDC, 3.25 A, 65 W ma
Internal battery 7500 mAh
Rechargeable for nearly 3.5 hours operation
Charging time about 1.5 hours
Process connection Micro quick connector, full pass through,
male (1.5 m hose with coupling included)
Interface USB
M12 (Modbus/RTU)
Ethernet (Modbus/TCP)
USB-A for 4G/LTE modem (optional)
Data logger Internal, 30-million values
Housing material PC + ABS, Aluminum
Protection class IP54 (cover closed)
Dimension 470 365 181 mm
Approvals FCC, EC
Weight 11 kg
5.2 Operating Conditions
Measuring medium Compressed breathing air
Humidity of the medium < 40%, non condensation
Temperature of the medium 0°C ... +45°C
Operation pressure 0.4 ... 1.5 MPa(g)
Higher pressure needs an e ternal
pressure reducer.
Storage & Transport temperature -10°C ... + 50°C
Ambient temperature 0°C ... +50°C
S605 9

5 Technical Data
Ambient humidity 0 ... 90% rH
Sampling flow rate 6 LPM@4 MPa(g), depends on input
pressure
Sampling rate 1 sample/sec
5.3 Measurement Data
Oil Vapor Sensor (only or S605-I version)
Measuring range 0.001... 5.000 mg/m³
(Based on 1000 hPa(a), 20°C, 0%
relative humidity)
Accuracy 5% of reading ± 0.003 mg/m³
Resolution 0.001 mg/m³
Sensor principle Photo ionization detector
UV lamp lifetime 6,000 working hours or 1 year,
whichever comes first
Oil Mist and Particle Sensor (only or S605-C version)
Measuring range 0.0 ... 5.0 mg/m³
(Based on 1000 hPa(a), 20°C, 0%
relative humidity)
Accuracy 15% of reading ± 0.1 mg/m³
Resolution 0.1 mg/m³
Sensor principle Oil mist and particle sensor
O2 Sensor
Measuring range 0 … 25%
Accuracy ±(1% of reading +0.05%)
Resolution 0.1%
Sensor principle Optical o ygen sensor
CO Sensor
Measuring range 0 … 20 ppm
Accuracy ±(5% of reading +1 ppm)
Resolution 0.1 ppm
Sensor principle Electrochemical sensor
CO2 Sensor
10 S605
Table des matières
Autres manuels SUTO Instrument de mesure
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur