1 0 MA100/1 SA01 Datum 09.05.2000 Art.Nr. 80336 Z.Nr. 8699098 Änd.Stand -/-
lowstheprogramming of a
calculating factor.
FAC: Calculating factor: used to
obtainfor exampleanangle
display. Basis is the maximal
possibleresolutionof1/1000
mm. The calculation factor
whichhastobeprogrammed
=measuring
range to be displayed / total
workingrange [1/1000mm].
Example:
anglemeasurement
on a circular disk with a dis-
play range of 0 to180°; dis-
playin1/100°; circumference
of the circular disk 942,48
mm;hencetotalworkingran-
ge471,240 mm;
FAC=18000/471240=0,03820
REF: Absolutereferencepointfor
the measuring system. This
value is determined by cali-
brating the system according
to chapter 9.
DEC.: Decimal point: modification of
the decimal point position to
determinetheresolution, eg.
afterprogrammingthecalcu-
lation factor. Up to 5 digits
after the comma are possible.
DIR: Direction:countingdirection
of the measuring system
"UP" upward
"DOWN" downward
P KEY: Time for which P-key needs
to be pressed to enter the
programmingmode.
RESET: Reset enable: reset to refe-
rence value via key .
"ON" Reset function on
"OFF" Reset function off
ABS/REL: incrementalmeasurementen-
able.
"ON" functionon
"OFF" function off
REF.EN.: Ref.valueinput enable:
to enter / change calibration
value.
"ON" Referencevaluechangeon
"OFF" Referencevaluechangeoff
UNITS: Unit of measure selection.
BRIGHT.: LEDdisplaybrightnesstoggle.
USL: upperlimitingvalue(contactA1)
LSL: lowerlimitingvalue(contactA2)
LANGUAGE: Language:Tochoose thelan-
guagedisplayed
"GER" German
"ENG" English
CONTROL: NurfürServicezwecke
CALIBR.: Fine calibration of the sensor.
Only has to be carried out be-
fore the first use of the sy-
stem or if system compon-
entsareexchanged.
9. Input mode
Resetfunctionviakeyboard
• Presskey tosetthedisplaytothereference
value.
Precondition: Parameter'Resetenable' (RE-
SET) in programming mode must be program-
med to "ON", but unit must not be left in
programming mode (see chapter 7 'To leave
programmingmode').
Resetfunctionviareferenceinput
By short activation (ground switching) of the
inputRESETeg.byapushbutton orby usinga
proximity switch. (Closing contact function)
Incrementalmeasurement
Press key for activation.
• The display is zeroed.
•Sign blinks.
•Press key to deactivate. The absolute
measuringvalue is displayedagain.
Precondition:Menupoint'Incrementalmeasu-
rementenable'(ABS/REL)inprogrammingmode
mustbeprogrammedto"ON",butunitmustnot
beleftinprogrammingmode(seechapter7‚'To
leaveprogrammingmode').
Referencevaluemodification
Forreferencevaluealteration,keys +must
bepressedsimultaneously.
The display then shows the reference value,
which can be changed via the two arrow keys.