Electrical connection (4-20 mA current output)
The 2-wire 4-20 mA current loop combines the transmitter
output signal and supply voltage in one cable where the
current consumption represents the measurement signal.
The 4-20 mA signal complies with the NAMUR NE 43
standard.
A current loop resistor () value of 250 ohms is commonly
used and will provide a 1-5 VDC across the resistor. The
current loop resistor value can be chosen freely up to a
maximum of 800 Ohm, provided that the minimum and
maximum supply voltage range is respected.
INFORMATION: The loop current resistor should
have a low temperature drift coefficient to ensure
best measurement performance.
Calculation of current ():
Calculation of voltage ():
Calculation of resistor ():
Connector pinout and cable wiring (4-20 mA current output)
Connector pinout 4 pin DIN175301-803A
Connector pinout 4 pin M12, IEC61076-2-101
Cable (Color code DIN41700)
Signal to pressure conversion (4-20 mA current output)
The transmitter with 4-20 mA current output is available with
different pre-configured output scaling.
The current signal () can be converted to pressure using the
following linear expression:
Current to pressure conversion:
Calculation of constants:
+
Where and
are the minimum and maximum
pressure, respectively; and are the minimum and
maximum current, respectively.
ATTENTION! The PCM-1 has a built-in self-diagnostic
and in case of product failure the current output will
provide a fault signal in the range of 3.2 - 3.6 mA in
compliance with the NAMUR NE43 standard. Replace
the unit in case of receiving a fault signal.
User configuration of current output scaling
The S4-Connect™ interface enables flexible user configuration
of the analog current output. The minimum and maximum
output current and pressure can be set to any value within the
valid range for the transmitter. For more information refer to
the full operating instructions or visit www.sens4.com/s4-
connect.html
Measuring range (0 bar to full-scale)
Measuring range (0 psi to full-scale)
15, 20, 30, 50, 100, 200, 250, 500 psi
Ceramic membrane absolute or gauge
Output resolution (Voltage)
Output resolution (Current)
External current shunt resistor
Solid state relay contact rating
250 mA, 50 VDC / VAC peak
Operating ambient temperature
Storage ambient temperature
Maximum media pressure (2)
Protection rating, EN 60529/A2:2013
Power consumption (Voltage version)
Power consumption (Current version)
Reverse polarity protection
100 mA (Thermal recoverable)
Connector DIN 175301-803A
Connector M12 IEC 61076-2-101
PA Nylon, Nickel plated Zinc alloy
Sensor diaphragm (media wetted)
Process leak tightness (ISO 27895:2009)
EN61000-6-2, EN 61000-6-3
Pressure directive 97/23/CE
Dimensions
P/N: PCM-1-xxxxxx1
M12 IEC 61076-2-101
P/N: PCM-1-xxxxxx2
DIN 175301-803A
P/N: PCM-1- xxxxxx3
P/N: PCM-1- xxxxxx4
P/N: PCM-1- xxxxxx5
Cable with flying leads
All dimensions are in mm unless otherwise stated –General tolerance ISO 2768-1 M
3D step files are available online at: www.sens4.com/support.html
Maintenance
Maintenance is not required during the lifecycle of this product.
The calibration may shift during the life-time and re-calibration
can be performed by the user. Refer to PCM-1 configuration and
operating manual.
The PCM-1 can be user configured, calibrated and tested using
the S4-Connect™ USB adapter.
Return
Before returning a product to Sens4 proper return forms and a
return materials authorization (RMA) must be filled out. The
RMA procedure can be found on:
www.sens4.com/support.html
Disposal in the European Union
At the end of life of this product, it must be
disposed according to the European Directive
2012/19/EU (WEEE). This product should not be
mixed with general household waste.
WARNING! If the product has been exposed to human
or environmental hazards materials during its use,
ensure proper decontamination before disposal.
For proper treatment, recovery and recycling, please take this
product to designated collection points. Please contact your
local authority for further details of your nearest designated
collection point.
For questions regarding disposal please contact your dealer or
Sens4 for further information.
Declaration of Conformity
This declaration of conformity has been made in accordance with EN ISO/IEC 17050-
1:2010
Sens4 A/S
Nordre Strandvej 119G
3150 Hellebaek
Denmark
We hereby declare under our sole responsibility that the following products:
Complies with the requirements of following relevant European Union harmonization
directive:
Electromagnetic Compatibility (EMC) Directive 2014/30/EU
RoHS Directive EU 2015/863
Conformity is assessed in accordance to the following standards:
Product family standard, Measurement, control and
laboratory equipment
Technical documentation for the assessment of
electrical and electronic products with respect to the
restriction of hazardous substances
Ole Wenzel, Chief Executive Officer
This declaration of conformity is available online at: www.sens4.com/doc.htm
Document number: DOC-PCM-1-01, Revision: A
Warranty and disclaimer
Sens4 warrants this product to be free from defects in
materials and workmanship for a period of 24 months from
the date of the delivery.
Warranty does not cover:
1. Mechanical or corrosive damage to the sensor
diaphragm
2. Physical or deposition contamination of the sensor
diaphragm
3. Damage caused by shipping
4. Normal wear and tear
5. Incorrect use or installation
6. Operation beyond the published design limits
Sens4 is not liable for any claims arising from improper use,
incorrect installation or use with gases or liquids not
compatible with the media wetted materials described in the
specifications table. Sens4 is not liable for loss of profits or
revenue, overheads, loss of data, reinstallation costs, damage
to other equipment or any incidental or consequential
damages of any nature.
The Standard Terms and Conditions can be found on
www.sens4.com and shall apply to the sales contract and
use of this product.
Sens4 A/S | Ndr. Strandvej 119G | 3150 Hellebaek | Denmark