Perfect Prime TC0520 K Manuel utilisateur

Datalogger Thermometer
TC0520
Instruction Manual

CONTENTS
P. 1
1. General Description
2. Safety Information
3. Features
4. Specications
5. Symbol Denition and Button Location
6. Button Instructions
6.1 Power ON/OFF button
6.2 Backlight button
6.3 Data-Hold button
6.4 REC button
6.5 MEM button
6.6 RECALL Saved Reading Button
6.7 MAX/MIN/AVG Button
6.8 Selecting Temperature Units
6.9 SETUP Button
7. Operating Instructions
7.1 Setup Options
7.2 Menu Item
7.3 Menu Description
7.3.2 Select thermocouple type
7.3.3 Setting interval time for data storing
7.3.4 Set oset to compensate for probe errors
7.3.5 Set alarm point (only for channel T1)
7.3.6 Set T1-T2 subtraction mode
7.3.7 Set auto power o time
7.3.8 Set system clock
7.4 Clearing Data Logger Records
7.5 Clearing Instant Read-out Memory
7.6 Connecting To A computer
8. Power Operation
8.1 Battery Replacement
9. Maintenance
10. Testlink SE-520 Software
........................................................................ 2
......................................................................... 2
........................................................................................ 3
............................................................................ 4 - 5
......................................... 6 - 7
................................................................... 8 - 10
.................................................................. 8
.......................................................................... 8
........................................................................ 8
................................................................................. 8
............................................................................... 9
....................................................... 9
................................................................... 9
........................................................ 10
............................................................................ 10
........................................................... 11 - 16
............................................................................ 11
............................................................................... 11
...................................................................... 12
.................................................... 12
...................................... 13
................................ 13
...................................... 14
.................................................. 14
....................................................... 15
................................................................. 15
...................................................... 16
................................................ 16
.......................................................... 16
........................................................................... 17
.................................................................. 17
................................................................................ 18
....................................................... 19 - 28

1. GENERAL DESCRIPTION/
2. SAFETY INFORMATION
P. 2
Thank you for choosing our data logger thermometer. To ensure the
safety and the best performance of this instrument, we recommend you
to read and follow the manual carefully before operation. Data can be
stored in the meter or directly saved on a computer through PC interface.
Recorded data can be further processed by PC computer.
Read the following safety information carefully before attempting to
operate or service the meter. Use the meter only as specied in this
manual; otherwise, the protection provided by the meter may be
impaired.
ENVIRONMENT CONDITIONS
• Altitude up to 2000 meters
• Relatively humidity: 90% max
• Operation ambient temperature: 0°C to 50°C
MAINTENANCE & CLEARING
• Repairs or servicing not covered in this manual should only be
performed by qualied personnel.
• Periodically wipe the case with a dry cloth. Do not use abrasives or
solvents on this instrument.
SAFETY SYMBOLS
Comply with EMC
When servicing, use only specied replacement parts.
SAFETY INFORMATION
GENERAL DESCRIPTION

3. FEATURES
P. 3
• Four-channel inputs
• Support K, J, E, T type thermocouple
• Temperature alarm function
• Fast response and sampling rate
• 16,000 data logger records each channel
• Instant recall function
• USB PC interface with Windows software included
• Adjustable auto power o timer

P. 4
4. SPECIFICATIONS
Measurement range
Resolution
Accuracy
Temperature coefficient
K: -200°C ~1372°C (- 328°F ~ 2501°F)
J: -200°C~1000°C (- 328°F ~ 1832°F)
E: -200°C~750°C (- 328°F ~ 1382°F)
T: -200°C~400°C (-328 °F ~ 752°F)
0.1°C < 600°C / 0.1°F < 1000°F,
1°C ≥600°C / 1°F ≥1000°F
Accuracy is specied for ambient
temperatures between 18°C (64°F) and
28°C (82°F). The specications do not
include thermocouple sensor error.
±(0.1% of reading+0.7°C) ±(0.1% of
reading+1.4°F)
below -100°C(-148°F)
±(0.4% of reading +0.7°C) ±(0.4% of
reading +1.4°F)
0.01% of reading + 0.05°C per
°C (<18°C or >28°C)

4. SPECIFICATIONS
P. 5
Sample rate
Battery type
Battery lifetime
Operation temperature
Operation humidity
Storage temperature
Storage humidity
Dimensions/ Weight
Standard Accessories
1 time per second
UM-4 or AAA 1.5V battery x 4
Approx. 100 hours (alkaline battery)
0°C to 50°C (32°F to 122°F)
10 to 90%RH (no condensing)
-20°C to 60°C (-4°F to 140°F)
10 to 75%RH
187mm(L) × 75mm(W) × 29mm(H)/
Approx. 290g
Instruction manual, battery
1.5V AAA x 4 pcs, Windows software,
micro USB cable, K type bead prob
x 2 pcs, carrying case.

5. SYMBOL DEFINITION &
BUTTON LOCATION
P. 6
Battery condition indicator
Minimum indicator
Maximum indicator
Average indicator
Setup option indicator
Auto Power O enabled indicator
Recording data logger indicator
Memory full indicator
Memory group indicator
Recall Memory group reading indicator
Thermocouple type
Temperature reading
Temperature channel
Temperature subtraction mode
Temperature alarm indicator
Hi Temperature alarm indicator
Lo Temperature alarm indicator
Compensate for probe errors
Temperature measurement units
Data hold indication)

5. SYMBOL DEFINITION &
BUTTON LOCATION
P. 7
Thermocouple Input
Display Screen
Power ON/OFF Button)
Back Light Button
MAX MIN AVG Button
DATA HOLD Button
°C , °F Button
2
3
4
5
6
7
1REC Button
RECALL Saved Reading Button
SETUP Button
MEM Button (Set 100 Memory)
USB Interface
Tilt Stand
Battery Compartment
9
10
11
12
13
14
8
1
2
3
5
9
10
4
6
7
8
11
13
12
14

6. BUTTON INSTRUCTIONS
P. 8
6.1 Power ON/OFF Button:
Press the button to turn on the meter. Press and hold button for
3 seconds to turn o.
6.2 Backlight Button:
Press to turn on the LCD backlight. This makes it easier to read in
dark environment. Press again to turn o backlight. The backlight
will be automatically turned o after 30 seconds to save battery power.
6.3 Data-Hold Button :
Press button to freeze the data shown on the LCD screen. Press it
again to exit Data-Hold mode.
Note: When the unit is in the Data-Hold mode, , , , and
buttons are disabled.
6.4 REC Button:
Press button to start recording data. The “ ” symbol will display
on the screen. To stop recording, press button again.
Note: During recording period, most of the buttons are disabled, such
as the , . All settings must be made before starting the data
logger function.
Note: When the memory is full (32,000 recorders), “ ” symbol will
blink on the LCD screen. The data logger stops.
Note: When battery power is low ( “ ” symbol lights up on the
screen), data logger cannot be started. If the battery is running low
during data logging, it will stop recording automatically.

6. BUTTON INSTRUCTIONS
P. 9
6.5 MEM Button:
Press button to save readings on the LCD screen. The“ ”symbol
lights up on the screen for 2 seconds. Press the button again to store
next group readings. The store readings on the LCD screen for 00~99
group.
6.6 RECALL Saved Reading Button:
Press the button to recall saved readings. The “ ”symbol lights
up on the screen.
Note: When the meter is during connection section, this function is
disabled. Press the , , , or button to select MEM group. The
LCD shows “ hour:minute:second ” on the screen for 2 seconds, and
then shows the readings. Press or button to exit RECALL mode.
6.7 MAX/MIN/AVG Button:
Under this mode, the unit simultaneously monitors and stores the
maximum, minimum and average value in the memory. The unit will
keep updating/refreshing the data.
To start:
1. Press button. “ ” symbol lights up on LCD, the reading shows
the maximum data.
2. Press button again to show minimum data; the “ ” symbol
lights up on LCD.
3. Press button again to show average data; the“ ” symbol lights
up on LCD.
4. Press button again, the“ , and ”symbol blinks together,
the readings shows real time data.
Ce manuel convient aux modèles suivants
4
Table des matières
Autres manuels Perfect Prime Enregistreur de données

Perfect Prime
Perfect Prime TC0376 Manuel utilisateur

Perfect Prime
Perfect Prime TH0165 Manuel utilisateur

Perfect Prime
Perfect Prime TC0309 Manuel utilisateur

Perfect Prime
Perfect Prime CO2000 Manuel utilisateur

Perfect Prime
Perfect Prime TC0301 Manuel utilisateur

Perfect Prime
Perfect Prime TC0378 K Manuel utilisateur