Evoqua IONPURE IP-LXM04 Manuel

IONPURE®
LX CEDI
Modules
Operation
&
Maintenance
Manual
IP-MAN-LX-1220-EN
Rev 3
December 2020
This manual covers
model numbers:
·IP-LXM - - EU-4
·IP-LXM - - HI-3
·IP-LXM - - X-4
·IP-LXM - - Z-5
Evoqua Water Technologies, LLC
558 Clark Road
Tewksbury, Massachusetts 01876, USA
ionpure@evoqua.com (email)
www.ionpure.com (web)
Visit web site to find local Ionpure contact

IONPURE ®LX CEDI Modules
Page 2 IP-MAN-LX-1220-EN.pdf
TABLE OF CONTENTS
DISCLAIMER STATEMENT 5
PROPRIETARY RIGHTS STATEMENT 5
MANUAL USER'S GUIDE 5
OPERATING MANUAL REVISION HISTORY 6
1. INTRODUCTION 7
1.1. LX Product Family Overview 7
1.2. Using This Manual 8
1.3. Installation Precautions 8
1.4. Operating Precautions 9
1.5. Shutdown Precautions 9
2. PRE-INSTALLATION: PREPARATION & REQUIREMENTS 10
2.1. Tools Required 10
2.2. Unpacking and Moving LX Modules 10
2.3. Electrical and Plumbing Connection Requirements 11
2.4. Operating Requirements 11
2.4.1. Operating Environment 12
2.4.2. Space Requirements 12
2.4.3. Module Orientation 12
2.4.4. DC Electrical Requirements 12
2.4.5. Feed Water Requirements 12
2.4.6. Drain Requirements 13
2.5. Flow Rates and Pressure Drops 13
3. LX MODULE INSTALLATION 14
3.1. Preparation 14

IONPURE ®LX CEDI Modules
Page 3 IP-MAN-LX-1220-EN.pdf
3.2. Checking Tie Bar Torque 15
3.2.1. Tightening End Plate Tie-bar Nuts 15
3.3. Water Connection Configuration and Fittings 16
3.3.1. LX Piping Adapter Sealing Mechanism – EU, X, Z Modules (BSP Male) 16
3.3.2. LX Piping Adapter Sealing Mechanism – HI Modules (BSP Female) 17
3.4. Electrical Connections 18
3.4.1. Electrical Precautions 18
3.5. RO/CEDI System Design Considerations 19
4. PREPARATION FOR STARTUP 20
4.1. Verify Feed Water Meets LX Quality Requirements 20
4.2. Estimate DC Current Required 20
5. START-UP PROCEDURE 21
5.1. Test Interlocks 21
5.2. Startup of LX Modules 21
5.3. Minimum Reject Flow Rate 21
5.4. Water Recovery 21
5.4.1. Example calculation of recovery 22
5.4.2. Example calculation of reject flow 22
6. MAINTENANCE AND TROUBLESHOOTING 23
6.1. General Maintenance Guidelines 23
6.1.1 Operating Data Log Sheet 23
6.1.2 Periodic Maintenance 23
6.2. Approved Cleaning Procedures 23
6.3. Cleaning and Sanitization Prompts 24

IONPURE ®LX CEDI Modules
Page 4 IP-MAN-LX-1220-EN.pdf
6.4. Clean-In-Place (CIP) System Construction 24
6.5. Chemical Cleaning and Sanitization Precautions 26
6.6. Typical Procedure for Cleaning or Sanitization 26
6.6.1 Preparation 26
6.6.2 Cleaning 26
6.6.3 Return to service 27
6.7. Cleaning Solution Recipes 28
6.8. Recommended Grades for Cleaning Chemicals 30
6.9. Hot Water Sanitization (HWS) 31
6.8.1 HWS Phase 1 – Hot Water Introduction 31
6.8.2 HWS Phase 2 – Temperature Hold 31
6.8.3 HWS Phase 3 – Return to service 32
7. TROUBLESHOOTING 33
8. SHUTDOWN AND STORAGE 40
8.1. System Shutdown 40
8.2. Startup After Shutdown 40
8.3. Rebuild or Repair 40
8.4. Disposal 40
APPENDIX A: LX MODULE SPECIFICATIONS 41
APPENDIX B1: LX LAYOUT & ELEVATION DRAWING (EU, X, Z) 43
APPENDIX B2: LX LAYOUT & ELEVATION DRAWING (HI) 44
APPENDIX B3: LX JUNCTION BOX DRAWING (ALL MODELS & SIZES) 45
APPENDIX C: LX ADAPTER LIST 46
APPENDIX D: LX MATERIALS OF CONSTRUCTION & APPROVALS 47

IONPURE ®LX CEDI Modules
Page 5 IP-MAN-LX-1220-EN.pdf
DISCLAIMER STATEMENT
The operation and maintenance manual should provide complete and accurate
information to meet your operating and/or service requirements based on the information
available at the time of publication. The information in this manual may not cover all
operating details or variations or provide for all conditions in connection with installation,
operation and maintenance. Should questions arise which are not answered specifically
in this manual, contact your water system supplier.
IONPURE reserves the right to make engineering refinements that may not be reflected
in these manuals. The material in these manuals is for informational purposes and is
subject to change without notice.
PROPRIETARY RIGHTS STATEMENT
This manual discloses information in which IONPURE has proprietary rights. Neither
receipt nor possession of this manual confers or transfers any right to the client, and by
its retention hereof, the client acknowledges that it will not reproduce or cause to be
reproduced, in whole or in part, any such information except by written permission from
IONPURE. The client shall have the right to use and disclose to its employees the
information contained herein for the purpose of operating and maintaining the IONPURE
equipment, and for no other purpose.
In the event the content of this manual is altered or section/items are omitted during a
reproduction, in whole or in part, and instructions or definitions within the reproduction
result in personal injury to those who follow the altered instructions, the burden of
responsibility for personal injury falls solely on the party who effects the reproduction.
MANUAL USER'S GUIDE
This manual describes the procedures necessary to install, operate, and maintain your
IONPURE Continuous Electrodeionization modules. Please read this manual carefully
before installing and operating your modules. The module warranty may be voided if
installation or operation instructions are not followed correctly.
Notes, Warnings, Cautions are used to attract attention to essential or critical information
in a manual. Warnings and Cautions will appear before the text associated with them,
and notes can appear either before or after the associated text.
NOTE: Notes are used to add information, state exceptions, and point out areas
that may be of greater interest or importance.

IONPURE ®LX CEDI Modules
Page 6 IP-MAN-LX-1220-EN.pdf
Cautions indicate a situation that may cause damage or
destruction of equipment or may pose a long-term health hazard.
Warnings indicate condition, practices, or procedures which
must be observed to avoid personal injury or fatalities.
IONPURE continually strives to provide safe, efficient, trouble-free equipment using the
optimum technology for your application. If problems should develop, IONPURE's
worldwide network of technical support will be available to provide assistance. For
service, sales, parts, or additional manual copies, please visit the website:
www.ionpure.com.
OPERATING MANUAL REVISION HISTORY
Event Date Description
First publication 05-2018 LX Family Operation & Maintenance Manual.
Rev 1 09-2018 Corrections on Table 6-1 (LX-4).
Added cleaning chemical grade suggestions (6.8).
Rev 2 10-2019 2.3 Added metric wire sizes.
3.4 Added terminal strip torque specification and
corrected junction box connection hole size.
Rev 3 12-2020 Updated LX-Z to reflect change from -4 to -5.
Removed obsolete parts from LX adapter list.
Updated L&E drawing in Appendix B1.

IONPURE ®LX CEDI Modules
Page 7 IP-MAN-LX-1220-EN.pdf
1. INTRODUCTION
1.1.LX Product Family Overview
Ionpure LX modules are medium flow rate CEDI devices. Their compact state-of-the-art
design assures ease of installation, maintenance, and service. LX modules are available
in the following sizes:
Size Nominal Flow Description
IP-LXM04 2.0 gpm (0.45 m3/h) 4-cell
IP-LXM10 5.0 gpm (1.14 m3/h) 10-cell
IP-LXM18 9.0 gpm (2.05 m3/h) 18-cell
IP-LXM24 12.5 gpm (2.84 m3/h) 24-cell
IP-LXM30 15.0 gpm (3.4 m3/h) 30-cell
IP-LXM45 22.5 gpm (5.1 m3/h) 45-cell
For more information on the LX module specifications and flow rates, see Section 2.4 and
Appendix A of this Manual.
There are four different types of LX modules (see Appendix D), each of which is available
in the six sizes listed above:
EU-4 For European Pharmacopeia (EP) applications
HI-3 For U.S. Pharmacopeia (USP) applications, hot water sanitizable (HWS)
X-4 For USP, NSF (drinking water) and general use (worldwide) applications
Z-5 For General Industry, with improved Cl2tolerance
Figure 1-1: LX Module (Cathode/Plumbing Side Shown)
HI type (BSP Female) EU / X / Z type (BSP Male)

IONPURE ®LX CEDI Modules
Page 8 IP-MAN-LX-1220-EN.pdf
1.2.Using This Manual
Service technicians should review this manual prior to going to the installation site. It lists
tools and materials needed to install the modules. It also outlines the site information
required to prepare for installation.
NOTE: The warranty may be void if installation or operation instructions contained in this
manual are not followed exactly.
This manual describes the installation, operation, and routine maintenance of the LX
modules. It also contains information on basic troubleshooting (See Section 7).
IONPURE strongly recommends all users read the entire contents of the manual. If the
LX module is not operating properly after going through the basic troubleshooting
exercises, contact your Local Service Provider.
1.3.Installation Precautions
·During operation, the electrode wiring inside the module junction
boxes are at high voltage and present a shock hazard.
THEREFORE, BEFORE TOUCHING THE INSIDE OF THE
JUNCTION BOX, CONFIRM THAT AC POWER TO THE DC
POWER SUPPLY HAS FIRST BEEN DISCONNECTED AND
LOCKED OUT ACCORDING TO STANDARD LOCKOUT/
TAGOUT PROCEDURES.
·To minimize the possibility of electric shock, confirm that all ground
wires are properly connected.
·Do not open the LX module. Opening the module will void the
warranty and cause irreversible damage.
·The module must be operated according to the design
specifications for temperature and humidity.
·Metal piping should never be connected directly to the module.
Non-metallic piping adapters are required at the module inlets and
outlets. It is then allowable to transition to metallic piping.
·Pipe sections prepared for installation must be inspected, and be
free of debris from storage or cutting tool particles. This must be
done before installation.
·Because LX modules have narrow flow distribution channels,
plugging by particles can cause permanent damage. Always
install pressure gauges, sampling ports, sensors, etc. in tee
fittings. Do not drill or tap into piping.

IONPURE ®LX CEDI Modules
Page 9 IP-MAN-LX-1220-EN.pdf
·Always flush out the piping to remove any debris before operating
the LX module.
·Installation of the LX module must be completed in accordance
with the procedures outlined in this manual. If deviations from the
prescribed procedures are deemed necessary to achieve the
desired performance, consult your local Service Provider.
1.4.Operating Precautions
·DO NOT APPLY POWER TO THE LX MODULE UNTIL PROPER
FLOW AND PRESSURE HAVE FIRST BEEN CHECKED AND
VERIFIED.
·Never block off (dead-head) the LX outlets. Dead-heading the
outlets can result in over-pressurization, leading to permanent
damage.
·Do not operate the module under conditions other than those
stated in the module manual. The prescribed feed water
requirements, electrical requirements, and flow configurations,
must be followed at all times. If the feed water quality or the
product water requirements change, contact the IONPURE
Technical Support Department for assistance.
·Once every six months:
oMake sure all wiring connections are tight.
oTest safety interlocks such as flow switches or connections to upstream
equipment.
oCheck torque on tie bar nuts and tighten as required.
1.5.Shutdown Precautions
·Confirm that the pressure in the unit is relieved until all pressures inside the unit are
atmospheric (i.e., all pressure gauges should read zero).
·Drain standing water and valve off or plug all inlets and outlets. This is to minimize
bacteria growth and prevent drying of ion exchange resins during shutdown.

IONPURE ®LX CEDI Modules
Page 10 IP-MAN-LX-1220-EN.pdf
2. PRE-INSTALLATION: PREPARATION & REQUIREMENTS
2.1.Tools Required
·Dolly or forklift to move the module into place
·Cords, cables or straps to secure module to dolly or forklift
·Slip joint pliers, for tightening of plumbing connectors
·Wire cutters/strippers, for wiring connections inside junction box
·Screwdrivers (flat blade and Phillips head) to connect wires at terminal strip
·Adjustable torque wrench with 10-50 ft-lb (14-68 N-m) range, 3/8” (10 mm) drive
·19 mm extra deep socket (Ionpure part number W2T210908)
·19 mm open end wrench
2.2.Unpacking and Moving LX Modules
·After uncrating the module, inspect it for any signs of damage. If damage is apparent,
immediately notify the carrier and your CEDI System Provider.
·The LX modules can weigh 70-330 lbs. (32-150 kg) dry, depending on size, and may
require mechanical assistance for lifting and moving into position. Both LX endplates
have two ¾-10 tapped holes for installation of lifting eyebolts, if desired.
·Ensure that lifting apparatus has appropriate load rating.
Figure 1-2: LX Module Lifting Points (Cathode/Plumbing Side Shown)
¾”-10
tapped hole
¾”-10
tapped hole Eyebolt or handle
(not provided)
Ce manuel convient aux modèles suivants
5
Table des matières
Autres manuels Evoqua Unité de contrôle
Manuels Unité de contrôle populaires d'autres marques

Festo
Festo Compact Performance CP-FB6-E Manuel de la liste des pièces

Elo TouchSystems
Elo TouchSystems DMS-SA19P-EXTME Manuel utilisateur

JS Automation
JS Automation MPC3034A Manuel utilisateur

JAUDT
JAUDT SW GII 6406 Series Guide rapide

Spektrum
Spektrum Air Module System Manuel utilisateur

BOC Edwards
BOC Edwards Q Series Manuel utilisateur