Quick Guide
Vaisala VaiNet Wireless Data Logger
RFL100

PUBLISHED BY
Vaisala Oyj
Vanha Nurmijärventie 21, FI-01670 Vantaa, Finland
P.O. Box 26, FI-00421 Helsinki, Finland
+358 9 8949 1
Visit our Internet pages at www.vaisala.com.
© Vaisala 2020
No part of this document may be
reproduced, published or publicly
displayed in any form or by any means,
electronic or mechanical (including
photocopying), nor may its contents be
modified, translated, adapted, sold or
disclosed to a third party without prior
written permission of the copyright holder.
Translated documents and translated
portions of multilingual documents are
based on the original English versions. In
ambiguous cases, the English versions are
applicable, not the translations.
The contents of this document are subject
to change without prior notice.
Local rules and regulations may vary and
they shall take precedence over the
information contained in this document.
Vaisala makes no representations on this
document’s compliance with the local
rules and regulations applicable at any
given time, and hereby disclaims any and
all responsibilities related thereto.
This document does not create any legally
binding obligations for Vaisala towards
customers or end users. All legally binding
obligations and agreements are included
exclusively in the applicable supply
contract or the General Conditions of Sale
and General Conditions of Service of
Vaisala.
This product contains software developed
by Vaisala or third parties. Use of the
software is governed by license terms and
conditions included in the applicable
supply contract or, in the absence of
separate license terms and conditions, by
the General License Conditions of Vaisala
Group.
This product may contain open source
software (OSS) components. In the event
this product contains OSS components,
then such OSS is governed by the terms
and conditions of the applicable OSS
licenses, and you are bound by the terms
and conditions of such licenses in
connection with your use and distribution
of the OSS in this product. Applicable OSS
licenses are included in the product itself
or provided to you on any other applicable
media, depending on each individual
product and the product items delivered
to you.

Table of contents
English.............................................................................................................................................5
Deutsch........................................................................................................................................ 23
Français.........................................................................................................................................41
Español.........................................................................................................................................59
Português.................................................................................................................................... 77
日本語...............................................................................................................95
中文................................................................................................................ 113
3

4 M211822EN-F

Setting up RFL100
Start the installation of RFL100 by performing the appropriate setup procedure:
• If you are connecting 1 probe, see Setup with 1 probe (page 5).
• If you are connecting 2 temperature probes using the probe splitter accessory, see Setup
with 2 temperature probes (page 7).
When you switch on an RFL100 that has not been accepted to any viewLinc
system yet, it starts to scan for VaiNet access points that are in installation mode.
RFL100 will connect to the access point with the best signal strength, and wait to
be accepted by the administrator of the viewLinc Enterprise Server. Make sure the
access point(s) are installed and in installation mode before starting to set up
RFL100 data loggers.
Setup with 1 probe
1
2
3
1Power switch.
2Transport protection cap. Remove
after installation is complete.
3Latch of the battery cover.
1. If any cable is connected to the service port of the data logger, disconnect it.
2. Open the battery cover of the data logger.
3. Make sure the power switch is in the O position.
4. If a probe is not already connected to the data logger, connect it now:
• To connect a probe directly to the data logger, first align the orientation mark on the
probe with the line above the probe connector. Then push the probe in all the way, do
not rotate.
• To connect a probe using an extension cable, use an accessory cable that has been
designed for RFL100. Insert the cable connector in the same way as a probe, and then
connect the probe to the cable. Note that HMP110 and HMP110T probes must always
be connected using a cable as they do not mechanically lock into the data logger
when the battery cover is closed.
5. Move the power switch to the On position.
5
ENGLISH

6. Close the battery cover of the data logger. Push the latch down until you hear a click. If
the cover does not close easily, push the probe (or extension cable) in and try again.
7. Look at the display and wait for the data logger to start up. Then verify the status:
• Data logger shows the word NEW on each aected channel when it detects that a
probe has been changed 1). The text is shown for a few seconds, after which the
display shows the measurement reading(s).
Figure 1 Detection of a new RH + T probe
• If dashes "- - -" are shown instead of measurement readings, check that the probe is
properly connected. It is possible to disconnect the probe or cable by pulling it with
the battery cover open.
• Battery indicator should show full batteries .
• There should be no error codes shown. If there are, see Error codes (page 19).
1) This feature is included in RFL100 firmware version 1.2.0 and newer.
6 M211822EN-F

Setup with 2 temperature probes
1
23
5 6
7
4
1Power switch
2Probe splitter
3Latch of the battery cover
4Instruction label with short version of this setup instruction
5Probe connectors of the probe splitter
6Temperature probes
7Numbering labels for TMP115 probes (included in the probe mounting accessories)
• Two temperature-only probes of following models: HMP110T, HMP115T and
TMP115 (any combination)
• Probe splitter (Vaisala item code CBL210834)
• Optional: extension cable(s), to be connected between the probe splitter and
probe
• Compatible firmware and software versions: at least RFL100 firmware version
1.2.0, AP10 firmware version 3.0, and viewLinc 5.0.2.
You must follow the procedure below to set up RFL100 with the
probe splitter. The procedure includes steps to label the probes to make it easy
to identify the channel assignment later. You can also use the info mode of the
RFL100 to check which probe serial number is connected to which channel.
For more information on the info mode, see RFL100 User Guide (M211861EN).
CAUTION!
Make sure the components of your viewLinc Monitoring System are
updated to the required firmware and software levels before setting up any
RFL100 data loggers for two-probe operation.
CAUTION!
7
ENGLISH

1. If any cable is connected to the service port of the data logger, disconnect it.
2. Open the battery cover of the data logger.
3. Make sure the power switch is in the O position.
4. If the probe splitter is not yet connected to the data logger, connect it:
a. If a probe or cable is currently connected to the probe connector, pull it straight out
without rotating it.
b. Align the orientation mark on the probe splitter with the line above the probe
connector. Push the probe splitter straight in all the way, do not rotate.
5. Connect the first temperature probe to the probe splitter, using an extension cable if
necessary. The first connected probe will be assigned to channel 1.
6. Move the power switch to the On position.
7. Look at the display and wait for the data logger to start up. You should see the word NEW
on channel 1 for a few seconds, after which it is replaced by the temperature reading.
8. Attach a label with number 1 to the probe you just connected. If you are using TMP115
probes, use the included probe labels. Connect the label to the thin cable between the
probe body and the sensor tip.
9. Move the power switch to the O position.
10. Connect the second temperature probe to the probe splitter, using an extension cable if
necessary.
11. Move the power switch to the On position.
12. Look at the display and wait for the data logger to start up. You should see the word NEW
on channel 2 for a few seconds, after which it is replaced by the temperature reading.
13. Attach a label with number 2 to the probe you just connected.
8 M211822EN-F

14. Look at the display:
• If dashes "- - -" are shown instead of measurement readings, check that the probe and
probe splitter are properly connected. It is possible to disconnect the probe splitter
accidentally by pulling it with the battery cover open.
• Battery indicator should show full batteries .
• There should be no error codes shown. If there are, see Error codes (page 19).
15. Close the battery cover of the data logger. Push the latch down until you hear a click. If
the cover does not close easily, push the probe splitter in and try again.
Once the data logger has been set up in two-probe mode, it will require a factory
reset to allow it to be set up in one-probe mode again.
For the factory reset procedure, see RFL100 User Guide (M211861EN).
9
ENGLISH

Mounting RFL100
A B
DC
68 [2.65]
mm
[in]
Figure 2 RFL100 mounting methods
AMounting with screws. Screws and wall plugs are included with the data logger.
BMounting with cable ties. Cable ties are included with the data logger.
CMagnetic mounting (with optional magnetic mounting bracket)
DMounting with a hook (hook not included)
10 M211822EN-F
Autres manuels pour VaiNet RFL100
4
Table des matières
Langues :
Autres manuels Vaisala Enregistreur de données

Vaisala
Vaisala HUMICAP HMT140 Manuel utilisateur

Vaisala
Vaisala CWL100 Manuel utilisateur

Vaisala
Vaisala VaiNet RFL100 Manuel utilisateur

Vaisala
Vaisala HUMICAP HMT140 Manuel utilisateur

Vaisala
Vaisala VaiNet RFL100 Manuel utilisateur

Vaisala
Vaisala HUMICAP HMT140 Manuel utilisateur

Vaisala
Vaisala MRG112 Manuel d'utilisation et d'entretien

Vaisala
Vaisala VaiNet RFL100 Manuel utilisateur