Model 4021
-6-
5. Maintenance
Store the meter within the rated storage temperature (from -20 to +70oC). When the front panel or the case is cleaned, use soft cloth
dipped with cleaner liquid. Do not use organic solvent like benzene or paint thinner as they may deform or discolor the case.
6. Calibration
In order to maintain long term accuracy, periodical calibration at an interval of about one year is recommended. Make a calibration in
the ambient condition of 23oC±5oC, 75%RH or less.
6.1 Calibration of Thermocouple input
Prepare the standard voltage generator, cold contact circuit (ice-bath), and standard thermocouple for calibration.
To enter the calibration mode, turn the power on with pushing M switch.
Connect the voltage generator and the CJS thermocouple as shown in the Figure-1.
Turn the power on with pushing the M switch.
Hold down the M switch for over five seconds.
After the lump test (all LED light up), display shows the sensor number, and then proceeds to the
Calibration mode with “
” message.
Press the M switch to enter ZERO and MAX calibration.
Display shows the measuring value. Resolution is 0.1oC for 100oC or lower, and 1oC for more than
100oC.
-ZERO calibration
Set the voltage generator to 0.00mV as shown in the Figure-1.
Press the >switch to momentarily memorizes the ZERO value. NOTE1
- MAX calibration
Set the voltage generator to the maximum voltage.
Press the Λ switch to momentarily memorize
the MAX value. NOTE1
Press the M switch to enter GAIN calibration
of the thermocouple in the ice-bath.
Display shows the measuring value. NOTE2
Set the voltage generator to 500.0mV as shown in the Figure-1.
Press the Λ switch to momentarily memorize the calibration value.
Press the M switch to enter ZERO calibration of the thermocouple in the ice-bath.
Connect the voltage generator, the CJS thermocouple, and the ice-bath as shown in the Figure-2.
Display shows the measuring value. NOTE2
Set the voltage generator to 0.00mV.
Press the >switch to momentarily memorizes the ZERO value.
Press the M switch to show “
”
Press the Λ switch to be complete your calibration. Display flashes to confirm your calibration.
Do not turn the power off to confirm your calibration.
An error message “
” will be displayed if you calibrate out of range.
Power OFF
Power OFF
No. Sensor MAX
K 950℃39.314mV
J 950℃54.956mV
T 400℃20.872mV
NOTE1: An error message “
” will be displayed
if you calibrate out of range.
NOTE2: Resolution is 0.1mV for 100mV or lower,
and 1mV for more than 100mV.
4
2
+
1
NC
53
−
CJS
CU+ CU−
generator
Standard voltage
Gray
Purple
3
2
+
1
NC
4
CJS
5
−
+−Gray
Purple
generator
Standard voltage
ice water 0℃
Standard
Thermocouple
CU−
CU+
Fig.‑1 Fig.‑2