Hygiene Equipment
Designed for People
6
Annual maintenance and inspection:
- Annual maintenance and inspection must be performed by Qualified staff
authorized by the manufacturer or together with qualified electrician and plumber.
- Repeat Before every use, Daily maintenance and Monthly maintenance.
- Inspection according to local requirements of the electrical installation, power cord,
earth cables and equipment must be performed by a qualified electrician.
- Inspection of the plumbing installation, flexible connection hoses, shower hoses,
vacuum breakers, check valves etc. must be performed by a qualified plumber.
Replace if necessary.
- Make a scalding test by shutting OFF the cold-water supply. The thermostatic mixer
shall shut OFF if only HOT water is supplied.
If it doesn´t shut OFF the mixer or cartridge must be replaced.
Autofill:
- If your TR900 is equipped with an Autofill function perform a function test
according to the instructions in the Operating manual. See table of content in the
Operating manual and read the Autofill chapter to see how it works.
Cleaning function:
- If your TR900 is equipped with Cleaning function perform a function test. Start the
cleaning function by turning the key on the side of the back cover counterclockwise
and make sure that cleaning detergent and water is mixed properly. Turn the key
on the Clockwise to turn it OFF.
- Verify that the check valve works OK.
Warning:
Never use the cleaning function on the patient!