Tisch Environmental PUF+ Manuel utilisateur

PUF+
Operator’s Manual
Date: 3/18/2008


Table of Contents
PUF+ NOTATION ...................................................................................................................1
PUF+ CONNECTIONS............................................................................................................ 3
USER INTERFACE ................................................................................................................. 7
MAIN STATUS SCREEN ....................................................................................................... 9
STEP BY STEP GUIDE......................................................................................................... 11
TIMER .................................................................................................................................... 17
DATA ..................................................................................................................................... 21
SETUP .................................................................................................................................... 27
ETI .......................................................................................................................................... 29
CONFIGURATION................................................................................................................ 31
CALIBRATION .....................................................................................................................33
DIAGNOSTICS...................................................................................................................... 39
INFO DISPLAY ..................................................................................................................... 41
DATA FORMATS.................................................................................................................. 43


Tisch Environmental PUF+ Operator’s Manual
1
PUF+ NOTATION
Notations and Measurements: The Table below describes the measurement notion used by
the PUF+ and units used to report those measurements.
Notation Units of
Measurement
Description
Qsys LPM System Flow Rate
Qstd LPM
Flow Rate at Standard Conditions
Reported at 760 mmHg and 25°C
Vstd m3Volume at Standard Conditions
Reported at 760 mmHg and 25°C
Qamb LPM Flow Rate at Ambient Conditions
Vamb m3Volume at Ambient Conditions
Tamb °C Ambient Temperature
Tcjc °C Cold Junction Compensation Temperature (compensates
Ambient Temp)
Pamb mmHg Ambient Pressure (Uncorrected Barometric Pressure)
Pdif inH2O Differential Pressure (Orifice Pressure)
Pcal inH2O Flow Calibrator Pressure


Tisch Environmental PUF+ Operator’s Manual
3
PUF+ CONNECTIONS
The descriptions of the various connections to the PUF+ control unit are described
below.
“Motor”
The “Motor” power connection supplies power to the motor being controlled by the
PUF+. Both Brush-Type and Brushless-Type motor are supported, but only one may be used
at a time. The PUF+ will come configured to match the motor type ordered. THE MOTOR
USED WITH THE PUF+ MUST MATCH THE AC POWER BEING SUPPLIED TO
THE PUF+. WHILE THE PUF+ CONTROLS THE MOTOR SPEED THROUGH THE
EMBEDDED ELECTRONICS, IT CAN NOT CHANGE THE OPERATING VOLTAGE
OR THE FREQUENCY OF THE INCOMING POWER THAT IS SUPPLIED TO THE
MOTOR.
“Power”
The “Power” connection supplies power to the PUF+. It should be connected to the
AC power line. Operation at both 120 Volts AC and 240 Volts AC are supported. The
system is designed to operate at either 50 or 60 Hz without the need to change the
configuration of the unit. However, a jumper inside the PUF+ determines the voltage on
which the unit operates. DO NOT APPLY 240-VOLT POWER TO THE PUF+ WHILE
THE UNIT IS CONFIGURED FOR 120-VOLT OPERATION. THIS WILL DAMAGE
THE UNIT. APPLYING 120-VOLT POWER TO THE PUF+ WHILE THE UNITS IS
CONFIGURED FOR 240-VOLT OPERATION SHOULD NOT DAMAGE THE UNIT,
BUT IS NOT RECOMMENDED OR SUPPORTED.

PUF+ Operator’s Manual Tisch Environmental
4
“AutoCal” Pressure
The “AutoCal” pressure port is used to make the Pcal measurement. It should be
connected to the calibrator during the AutoCal procedure.
“Ambient” Pressure
The “Ambient” pressure port is used to make the Pamb measurement. It should be
left open to atmosphere so that it properly senses the ambient pressure.
“Orifice +” Pressure
The “Orifice +” pressure port is used to make one half of the Pdif measurement. It
should be connected to the upstream port of the orifice (upper most port). Both “Orifice +”
and “Orifice –” must be connected to the orifice for proper flow measurement.
“Orifice –” Pressure
The “Orifice –” pressure port is used to make one half of the Pdif measurement. It
should be connected to the downstream port of the orifice (lower most port). Both “Orifice
+” and “Orifice –” must be connected to the orifice for proper flow measurement.
“Ambient” Temperature
The Ambient Temperature thermocouple port is used to make the Tamb
measurement. A Type K thermocouple must be used. The sensing tip of the thermocouple
should be located so that it receives air flow while the unit is sampling and is always shield
from the sun. By locating the sensor in this manner it can be used to monitor two items of
importance. Most importantly, this will allow for the accurate measurement of the ambient
temperature while the unit is sampling. In addition, when the unit is not sampling this allows
the temperature of the filter to be approximated. This can be of importance when sampling
for compounds that may changes phases (i.e. from solid to gas) due to changes in
temperature. Below, is a picture showing a suitable location for the thermocouple. Note the
thermocouple tip should be located so that it does not come into direct contact with any
surface.

Tisch Environmental PUF+ Operator’s Manual
5

Table des matières
Manuels Unité de contrôle populaires d'autres marques

Festo
Festo Compact Performance CP-FB6-E Manuel de la liste des pièces

Elo TouchSystems
Elo TouchSystems DMS-SA19P-EXTME Manuel utilisateur

JS Automation
JS Automation MPC3034A Manuel utilisateur

JAUDT
JAUDT SW GII 6406 Series Guide rapide

Spektrum
Spektrum Air Module System Manuel utilisateur

BOC Edwards
BOC Edwards Q Series Manuel utilisateur