Surface Concept HVPS Manuel utilisateur

19“ HVPS Detector Supply
1 Channel
(Release R020)
Manual

2
User Manual for the
19” HVPS Detector Supply 1 Channel
Release: R020
Manual Version 2.0
Printed on 2020-11-13
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
Surface Concept GmbH
Am Sägewerk 23a
55124 Mainz
Germany
phone: +49 6131 62716 0
fax: +49 6131 62716 29
email: info@surface-concept.de
web: www.surface-concept.de
All rights reserved. No part of this manual may
be reproduced without the prior permission
of Surface Concept GmbH.

3
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
1 Table of Contents
1 Table of Contents ......................................................................................................................3
2 Introduction................................................................................................................................5
2.1 General Information..........................................................................................................5
2.2 Safety Instructions .............................................................................................................5
2.3 General Overview...............................................................................................................6
3 Introduction................................................................................................................................7
3.1 Initial Inspection.................................................................................................................7
3.2 Installation............................................................................................................................7
4 Device Layout & Operation ...................................................................................................9
4.1 Device Layout.....................................................................................................................9
4.2 General Device Operation........................................................................................... 10
4.3 Operation of a MCP/CEM Detector.......................................................................... 11
4.4 Schematic Layout of the 19”HVPS Detector Supply R020.............................. 12
5 List of Figure ........................................................................................................................... 13

419” HVPS Detector Supply R020 Manual | Surface Concept GmbH
This side has been left blank on purpose.

5
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
2 Introduction
2.1 General Information
This manual is intended to assist users in the installation, operation and maintenance of Release Version
020 of the 19” HVPS Detector Supply. It is divided into 5 chapters.
2.2 Safety Instructions
Please read this manual carefully before performing any electrical or electronic operations and strictly
follow the safety rules given within this manual.
The following symbols may appear throughout the manual:
The“caution symbol”marks warnings, which are given to prevent an accidental damaging
of the device. Do NOT ignore these warnings and follow them strictly. Otherwise no
guarantee is given for arose damages.
Note
The “note symbol” marks text passages that contain important information/hints about
the operation of the device. Follow this information to ensure a proper operation of the
device.
The “high voltage symbol” marks warnings, given in context with the description of the
operation/use of high voltage supplies and/or high voltage carrying parts.
Hazardous voltages are present that can cause serious or fatal injuries. Therefore only
persons with the appropriate training are allowed to carry out the installation, adjustment
and repair work.

619” HVPS Detector Supply R020 Manual | Surface Concept GmbH
2.3 General Overview
The Surface Concept 19” HVPS Detector Supply R020 is a 1 channel high voltage module especially
designed for the Surface Concept 19”Basic Unit, a modular supply system.
The 19” HVPS Detector Supply R020 is laid out for the operation of e.g. MCP based detectors with metal
anode or CEM detectors. It provides one single high voltage with a positive polarity. It also provides the
capability to oat on an external reference potential.
The 19” HVPS Detector Supply R020 comes with a maximum output voltage of +3000V and can oat on
voltages of up to +/- 1000V. Check the specication sheet of your specic 19”HVPS Detector Supply R020
for further specic information.
The device can produce lethal high voltages of up to several kV. Hazardous voltages are
present, therefore only persons with the appropriate training are allowed to carry out
the installation, adjustment and repair work.
Do not open the power supply, while it is in operation. Hazardous voltages are present.
In case that the device must be opened, turn o the device rst AND pull out the power
plug.

7
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
3 Introduction
3.1 Initial Inspection
Visual inspection of the system is required to ensure that no damage has occurred during shipping. If
there are any signs of damage, please contact SURFACE CONCEPT immediately. Please check the delivery
according to the packing list (see Table 1) for completeness.
• 1x 19“ HVPS-HV Module R019
• 1x SHV Termination Plug
Table1: Packing list for the 19“ Basic Unit R019.
3.2 Installation
The general connection scheme of the 19”HVPS-HC Module R020 is as follows:
Figure 1: Connection scheme of the 19” HVPS Detector Supply R020 (the detector ange is given for illustration only.
Please check the manual of your MCP/CEM detector for the correct ange layout and feedthrough assignment).

819” HVPS Detector Supply R020 Manual | Surface Concept GmbH
Finish the complete cabling of the 19” HVPS Detector Supply R020 before switching on
the 19” Basic Unit and switch o the device rst before performing any changes to the
cabling.
• Install the 19”HVPS Detector Supply R020 into a free slot of the 19”Basic Unit (if not already installed).
• High voltage output is provided to the SHV socket named“MCP-B”.
• The socket named“MCP-F”connects the external reference voltage which is given to the device via the
“HV Ref. In” input to the front side of the MCP stack/ CEM of the detector
• Use SHV cables to connect the outputs of the 19”HVPS Detector Supply R020 to the corresponding HV
input/s of the MCP/ CEM detector (see your MCP/ CEM detector manual for more details).
• Use the SHV termination plug to terminate the “HV Ref. In” input when not working with an external
reference voltage.
Note
In cases that no reference voltage is applied to the device, the termination plug must be
used to ground the reference input of the 19”HVPS Detector Supply R020. With a missing
reference potential the device is not providing any output voltage.

9
Manual 19“ HVPS Detector Supply R020
19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
4 Device Layout & Operation
4.1 Device Layout
The layout of the 19” HVPS-HC Module R020 is given below.
1. SHV connector for output of operation voltage for a
MCP/CEM detector.
2. SHV connector for output of reference potential to
the front side of a MCP/CEM detector.
3. SHV connector for input of an external reference
potential (max. +/- 1,000V, check with your MCP
or CEM detector manual for details on maximum
reference potentials).
Figure 2: Layout of the 19”HVPS Detector Supply R020.
The 19” HVPS Detector Supply R020 is specied for the operation with an external
reference voltage of maximum +/- 1,000 V (see no. 3 in Figure 2).
Please also respect the corresponding specications for the maximum voltage for the
MCP front potential of your MCP/CEM detector (see the manual and the specication
sheet of your MCP/CEM detector for further details).

10 19” HVPS Detector Supply R020 Manual | Surface Concept GmbH
4.2 General Device Operation
After switching on the 19” Basic Unit, its display shows the “Surface Concept” animated logo, while the
device is scanning for the 19” HVPS Detector Supply R020 and its specic settings. This can take up to
several seconds. If the 19” Basic Unit is ready for operation, it switches into the standby mode and shows
an empty mask for the voltage adjustment (see Figure 3).
Push the“Start/Standby” button in the lower left
corner of the display to switch on the high
voltage.
Alternatively one can press the“Channel”control
knob.
After switching on the high voltage the device
is in the so called operation mode. In operation
mode the display shows the name of the selected
channel in the top line (in this case“HV 1”) as well
as the output voltage of that channel.
The dierent HV channels can be selected by
turning the “Channel” control knob.
Figure 3: Standby mode.
Figure 4: Operation mode.
Figure 5: Operation mode – voltage adjustment.
The “Adjust” control knob is used to adjust the
output voltage. Turn the “Adjust” control knob
clockwise/counterclockwise to increase/decrease
the value of the output voltage in a step width as
dened in the line “edit step”.
The line “set value” displays the nominal value
for the output voltage as adjusted by the user.
Voltage adjustment can only be made in this line.
The line “actual” displays the actual value for the
output voltage on the output connector as measured by the device.The device always regulates the actual
value of the output voltage to t to the nominal value as set by the user. Hereby the voltage measurement
is always a relative measurement between the two HV outputs of one channel. Additional reference
voltages (e.g. in oating operation) are not measured and therefore are also not displayed (see the manual
of the specic module for further details). Turn the“Adjust”control knob clockwise/counterclockwise while
pushing it to increase/decrease the step width in the line “edit step”.
Push the„Start/Standby“-button in the lower left corner of the display again to switch back to
the “Standby” mode.
Autres manuels pour HVPS
1
Table des matières
Autres manuels Surface Concept Capteur de sécurité

Surface Concept
Surface Concept Delayline DLD 8080 Manuel utilisateur

Surface Concept
Surface Concept MCPD25 Manuel utilisateur

Surface Concept
Surface Concept DLD6060-8S Manuel utilisateur

Surface Concept
Surface Concept DLD4040 Manuel utilisateur

Surface Concept
Surface Concept MCPD18 Manuel utilisateur

Surface Concept
Surface Concept CEM 4230 Manuel utilisateur

Surface Concept
Surface Concept MCPD40 Manuel utilisateur

Surface Concept
Surface Concept Delayline DLD 8080 Manuel utilisateur