Sterisil AC Manuel utilisateur

719 622 7200 Sterisil.com
STERISIL®AC SYSTEM
INSTALLATION AND
OPERATING MANUAL

GP6002.B
Sterisil, Inc.
835 S. Hwy 105 Ste. D
Palmer Lake, CO USA 80133
719 622 7200
NEED HELP OR HAVE QUESTIONS?
Our technical support staff is available and
ready to help. Please call, email, or visit us
online for additional support.
sterisil.com
TABLE OF CONTENTS
Introduction 3
Sterisil®Ac Features 4
Installation Details 5
Installation Instructions 6
Maintenance 11
Troubleshooting and FAQ 13

3
GP6002.B
INTRODUCTION
The ease of use and confidence that Sterisil®has brought to dental unit water line treatment now comes to
distilled and deionized water. Engineered for use with your office’s table top autoclaves and instrument washers,
the Sterisil®Ac produces distilled quality water with <10 ppm Total Dissolved Solids (TDS). Audible and visual
alarms make cartridge replacement worry-free and the Sterisil®Ac requires only minutes of maintenance. Its
space-saving, compact design allows for easy installation even under sinks.
This manual will show you how to install and maintain your new Sterisil®Ac unit and answer a few frequently
asked questions that you may have. For additional support, please contact Sterisil®at (719) 622-7200 or email

4GP6002.B
STERISIL®AC FEATURES
SIMPLE
• Compact design for easy sterilization center or under-sink installation.
• Optional shorter cartridges for even smaller spaces.
• Automatic shut off valves for easy, spill-free cartridge replacement.
• Optional filler wand for easy tabletop autoclave filling.
• Sliding mount for easy access.
• Metal enclosure protects internal components.
EFFECTIVE
• Produces distilled quality water with <10 ppm TDS for use with table top autoclaves.
• Water Quality Monitor samples the water purity in real-time.
• Audible and visual alarms alert you when cartridges need to be replaced.
• Dispensing faucet with LED monitoring light.
• Backup filters supply your office with water for your table top autoclaves even after audible and visual
alarms trigger.
• Designed for offices with source water containing <150 ppm TDS.
• Sterisil®ISO 13485 certification ensures that your new Sterisil®Ac unit has been tested for performance
and reliability.

5
GP6002.B
STERISIL®AC SYSTEM
SPECIFICATIONS
Pressure Min/Max 40 PSI/125 PSI
Temp. Min/Max 45°F/ 100°F
Tubing (PEX, PUR) 1/4” OD
Electrical Outlets 3 110V AC, 3 AMP
Fittings Quick Connects
INSTALLATION DETAILS
UNPACKING
Carefully unpack the contents of the box. The box will include the Sterisil®Ac System, Installation Manual,
Warranty Registration Card, large bag, small bag, a white table top faucet and 1/4” tubing
Check to make certain there was no damage during shipment. If damage is evident, contact the shipping
company immediately.
TOOLS
Required tools vary but typically include a sharp razor, knife, or tube cutter, measuring tape, drill, assorted bits
and screws, screwdrivers, wrenches, level, and clean rags.
LOCATION
Generally, the Ac System is installed centrally within the dental office in the equipment room, sterilization center,
hallway, or laboratory. If the hinged mounting bracket is selected, the typical mounting location would be on the
wall near the source water supply. If the sliding bracket is selected, the typical mounting location is under a sink
near the source water supply.
TUBING & CONNECTIONS
Copper piping should not be used anywhere downstream from the system. However, copper piping may be used
from the source water to the System itself. Make sure cuts on the tubing are square and clean to prevent leaks.

6GP6002.B
INSTALLATION INSTRUCTIONS
1. Unpack box and take note of the Sterisil®Ac specifications and dimensions before selecting a desired
installation site.
2. Prepare installation site.
a. Ensure there is one 110 V AC electrical outlet within 6 ft and a water supply line within 10 ft of the desired
installation site.
3. Install the mounting plate.
a. Test the fit of the mounting plate at the desired location by placing it on the wall.
b. Ensure that the mounting plate is level. Then mark each of the holes indicated below with a pen or pencil.
c. Using a ¼” drill bit, drill holes at the points you marked in step b.
d. Insert the drywall anchors into the holes and gently hammer them all the way in.
e. Place the mounting plate over the holes and using the Phillips #10 x 1” sheet metal screws, screw the
mounting plate into position.
4. Remove the Ac unit from the packaging and slide it onto the rails on the mounting plate as shown below. (Do
not install cartridges yet.)

7
GP6002.B
5. Install dispensing faucet.
a. Drill a 5/8” hole, ensuring holes are on sink surface at the desired location, and that LED data wires fit
without pinching during installation.
b. Insert the faucet assembly into the hole as shown.
c. Tighten faucet in place and install quick connect fitting.
6. Establish Sterisil®Ac plumbing connections.
a. Ensure that the municipal cold water supply line to your sink is turned off. From the municipal cold water
supply line, install a tee valve for under-sink installation (supplied by customer) and reduce to ¼” via
either quick connect fitting or compression fitting.

8GP6002.B
b. Insert the white ¼” (red in the diagram below) PEX tubing supplied with your Ac unit into the fitting and
ensure a secure fit.
c. Run the ¼” white tubing from the municipal cold water supply to the Ac unit and cut the tubing, leaving
enough slack to allow the system to slide out from the cabinet. Firmly insert the tubing into the red port
labeled “SOURCE WATER IN” on the side of the unit.
d. Run another length of ¼” white tubing to the quick connect fitting on the dispensing faucet. Cut the
tubing leaving enough slack to allow the system to slide out from the cabinet.

9
GP6002.B
7. Connect electrical and data lines.
a. Connect the black power cord to the 110V AC power outlet.
NOTE: The red LED on the faceplate of the unit may turn on and the alarm may sound after initially
plugging in the unit. If so, unplug the power cord and wait to plug in until after cartridges are installed
and water is turned on to the unit. b.
b. Connect the RJ12 data line from the Ac unit to the data union included with the unit. Connect the other
end of the data union to the RJ12 data line from the faucet.

10 GP6002.B