Sentry DCCP Reboiler Manuel d'instructions

Original Instructions
Installation, Operation
&Maintenance Manual
Sentry DCCP Reboiler
Single Line Conditioning Module
S-SW-IOM-00255-17 07-19

2 Sentry Equipment Corp
Do not install, maintain, or operate this equipment without reading,
understanding, and following the appropriate Sentry Equipment Corp
instructions. Otherwise, injury, damage, or both may result.
Copyright
© 2019 by Sentry Equipment Corp. All rights reserved. All product and company names
are property of their respective owners. This document contains proprietary information.
No part of this document may be photocopied or reproduced without the prior written
consent of Sentry Equipment Corp.
Limit of Liability
Sentry Equipment Corp, its employees, agents, and the authors and contributors to this
document specically disclaim all liabilities and warranties, express or implied (including
warranties of merchantability and tness for a particular purpose), for the accuracy,
currency, completeness, and/or reliability of the information contained herein and/or
for the tness for any particular use and/or for the performance of any material and/or
equipment selected in whole or part with the user of/or in reliance upon information
contained herein. Selection of materials and/or equipment is at the sole risk of the user of
this publication.
Note
The information contained in this document is subject to change without notice.
Table of Contents
Safety Information ......................... 3
General Safety Precautions ................. 4
General Description ........................ 5
Specications .............................. 6
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operation.................................. 10
Spare Parts List............................. 13
Drawings .................................. 14
Standard Warranty ......................... 17
Customer Support.......................... 17

Sentry DCCP Reboiler 3
Safety Information
Please read the entire manual before attempting to unpack, set up, or operate this
product. Pay careful attention to all Warnings, Cautions, and Notes. Failure to do so could
result in serious personal injury and/or equipment damage.
Use of Hazard Information
If multiple hazards exist, the signal word corresponding to the greatest hazard shall be
used.
Denitions
DANGER indicates a hazardous situation
which, if not avoided, will result in death or
serious injury.
WARNING indicates a hazardous situation
which, if not avoided, could result in death or
serious injury.
CAUTION, used with the safety alert symbol,
indicates a hazardous situation which, if not
avoided, could result in minor or moderate
injury.
NOTICE is used to address practices not
related to personal injury.
ÂNOTE
Information that requires special emphasis.
ÂTIP
Alternate techniques or clarifying information.
SHALL:
This word is understood to be mandatory.
SHOULD:
This word is understood to be advisory.
This symbol, when noted on a product, indicates a potential hazard that could
cause serious personal injury and/or death. The user should reference this
instruction manual for operation and/or safety information.
This symbol, when noted on the product, identies the location of the
connection for protective earth (ground).
Electrical equipment marked with this symbol may not be disposed of in
European public disposal systems. In conformity with European local and
national regulations, European electrical equipment users must now return old or
end-of-life equipment to the manufacturer for disposal at no charge to the user.

4 Sentry Equipment Corp
General Safety Precautions
Product Selection, Installation, and Use
Improper selection, installation, or use can cause personal injury or property damage. It is solely
the responsibility of users, through their own analysis and testing, to select products suitable
for their specic application requirements, ensure they are properly maintained, and limit their
use to their intended purpose.
Follow proper local, state, and federal regulations for proper installation and operational
requirements.
Always use caution and common sense when working with any chemical. Read the product
label and Safety Data Sheets (SDS) carefully and follow the instructions exactly.
Potential Equipment Hazards
Hot surfaces! This equipment may have very hot surfaces. If an operator contacts a hot surface,
injury may occur. Use protective clothing to prevent injury. If other equipment comes in contact
with a hot surface, damage to the equipment may occur. Ensure the area around this equipment
is kept clear to prevent this damage from occurring.
High pressures! This equipment may contain uids at very high pressures. Prior to installing,
removing, or maintaining this equipment, ensure that the equipment is isolated from all
connecting piping, the equipment is depressurized, the contents have been drained, and the
equipment is cool.

Sentry DCCP Reboiler 5
General Description
Do not install or use this equipment in any manner other than that which is specied in this
manual, otherwise the protection provided by this equipment could be impaired.
The Sentry® DCCP Reboiler Degassed Cation Conductivity Panel is designed to optimize
the analytical results for degassed cation conductivity of high purity water. Any other
use, or use not mentioned here, that is incompatible with the technical specications is
deemed inappropriate. The operator is solely responsible for any damage arising from
such use.
Other prerequisites for appropriate use include:
Observing the instructions, notes and requirements set out in this or updated
instruction manuals.
Observing all local safety regulations.
Observing all warnings and cautions in the documentation regarding all products
used in this system.
Observing the prescribed environmental and operational conditions.
Observing chemical compatibility with all wetted materials.
The Sentry DCCP Reboiler is available in various congurations with and without
conductivity cells and cation exchange column depending on each customer’s individual
requirements. Panels without the conductivity cells before the reboiler and without the
exchange column are designed to be used in conjunction with an analysis panel which
contains these components.
Regenerative Cooling Option
The Sentry DCCP reboiler has the option for designated cooling water or regenerative
cooling. With the regenerative cooling option, cooling water is not required. Instead, the
sample ow path is altered so that the sample cools itself while running through the DCCP
panel. The overall dimensions of the two versions are the same.
For best performance when using regenerative cooling, sample temperatures should not
exceed 113°F (45°C) or the analyzer sensor temperature limit designated in the analyzer
manual). Sample ow should not exceed 150 cc/min when using regenerative cooling.
ÂNOTE
Sentry Equipment recommends using a temperature-controlled cooling water source for sample
cooling, controlled to 77 F (25 C) for best CO2 removal performance. This recommendation is
consistent with applicable EPRI, ASME and ASTM guidance. If regenerative cooling is selected
the user should understand the DCC analyzer temperature compensation performance and
limits.
Please refer to the P&ID Layouts of both designs below.

6 Sentry Equipment Corp
Figure 1. Standard P&ID Figure 2. Regenerative Cooling Option
Specications
Dimensions
Folded Backplate: 36 in x 17.5 in x 14.9 in (91.2 cm x 44.5 cm x 37.8 cm)
Flat Backplate: 32 in x 22 in x 11.2 in (81.3 cm x 55.9 cm x 28.4 cm)
Weight
varies by model
61 lb (28 kg) for model DCCP-013100 (pictured on cover)
Pressure rating
100 psig (7 barg)
Temperature rating
130°F (54°C) with RC-100 before degas
Sample ow
100–150 cc/minute
Sample inlet/outlet connections
1/4 inch tube tting
from CE-2
CC
to CE-3
DCC
SC-1
HTR-1
6
TICTE
1
3
CE
DCC
CC
2
CE
RC
HTR-1
SC-1
1
CE
SC
1
TIC
1
TE
2
1
CIT
4
3
3
CIT
FICV
SC
2
CIT
CC
1
6

Sentry DCCP Reboiler 7
Sample cooler type
DTC-DCCP
Cooler cooling water connections
1/4 inch tube tting
Cooling water consumption
0.2 gpm (0.8 Lpm) at 22°C (71.6°F)
Input power
DCCP-x1xxx: 110–120 VAC, 1 phase, 50/60 Hz
DCCP-x2xxx: 220–240 VAC, 1 phase, 50/60 Hz
Power consumption
1.6 KVA
Cell holders (optional)
3/4 inch FNPT, 316 SS
Resin column (optional)
Sentry RC-100
Indoor/outdoor use
Altitude: Up to 6,561 ft (2,000 m)
Ambient temperature: 36°F to 122°F (2°C to 50°C)
Relative humidity: 0% to 85% RH
Pollution degree 2
Overvoltage Category II
Main supply voltage uctuations up to ±10% of the nominal voltage
Installation
Sharp Edges! Wear gloves when lifting and carrying the panel as edges could be sharp.
To ensure proper alignment and safe installation, lifting and carrying should be completed by at
least two people.
This instrument should be installed by qualied technical personnel to ensure adherence to all
applicable codes.
Use proper connection sizing to avoid an unsafe condition. This product comes standard with
1/4" compression ttings for the sample inlet/outlet, and 1/4" compression connections for the
cooling water inlet/outlet. Never attempt to install a metric equivalent tube, pipe, or ttings into
these connection points.
Lift unit by the stainless steel backplate. Do not lift using mounted components as these could
break, damaging the equipment and possibly causing personal injury.

8 Sentry Equipment Corp
Choosing a Mounting Location
Before installation, select a mounting location that provides the following:
Floor or wall must have adequate structural capability.
Anchor location and design is adequate.
Operation and maintenance space is sucient.
Cooling water is available.
Floor drain is accessible if required.
Mounting surfaces are at/level and at appropriate height.
There is adequate light for operation and maintenance.
For optimum equipment reliability and installation, the area should:
Be free from mechanical vibration.
Be free from corrosive media.
Have a normal ambient temperature near 75-80°F (24-27°C) with minimal variation.
Be free from strong heat radiation.
Provide adequate support and anchoring to minimize gravity and thermal loads on the
connections and to minimize any vibration eect.
If installing over drywall, the panel must be secured to the wood studs using 3/8 inch
(M10) hardware.
Connecting Sample Lines
Improper connection sizing will result in an unsafe condition.
Improper cell holder locations can adversely aect the operation of the unit.
Connect the sample inlet and outlet at the bottom of the panel. Both connections are
1/4" tube compression ttings. Clean and deburr all connecting pipe and tubing before
installing.
ÂNOTE
Samples supplied to the DCCP should be cooled to 77°F (25°C) and pressure controlled by either
a backpressure regulator or head cup to maintain a constant sample ow rate.
If you have purchased the DCCP without factory mounted cell holders, refer to Figure 3
for proper cell mounting and drain piping congurations. Cell holder(s) must be installed
in an orientation that maintains the water level in the degas chamber near the cross over
section of the degas chamber. To accomplish this, the sample outlet from the panel must
be even with the cross over section of the degas chamber. It is important to follow these
instructions to obtain accurate conductivity measurements.

Sentry DCCP Reboiler 9
Connecting Cooling Water
Improper connection sizing will result in an unsafe condition.
Connect the cooling water inlet and outlet piping (3/8”) to the bottom of the panel.
Connecting Electrical Wiring
Connect plant electrical supply wiring of ample
rating as indicated on the Electrical Wiring
Diagram (see drawing in back).
Supply wire must be between AWG 12
(3.31mm2) and AWG 4 (21.15mm2).
Supply wire must be rated to 230°F (110°C) or
higher
1
Sample Inlet
Sample Outlet
Cooling Water Inlet
Cooling Water Outlet
17 1/2 [444]
without analyzers
24 1/2 [622]
with analyzers
0 [0]
0 [0]
36 [914]
Heater
Sample
Cooler
Resin Column
Flow Indicating
Control Valve
Conductivity Cell
Conductivity Cell
Conductivity Cell
Analyzer (by customer)
Analyzer (by customer)
Temperature
Indicating Controller
Condensate Drain

10 Sentry Equipment Corp
The free length of supply wire inside the electrical enclosure must be at least 6 inches
(150 mm).
When connecting external supply voltage to the electrical enclosure, installer must
select appropriate hardware to maintain NEMA 4x rating.
Panel must be connected to a properly grounded electrical supply.
An electrical disconnect must be provided for the panel:
– It must be suitably located (e.g., close to the panel) and easily reached.
– It must be marked as the disconnecting device for the panel.
Operation
Unit will be hot during normal operation.
Starting Up
1. Establish cooling water ow.
2. Initiate sample ow through the reboiler.
3. Adjust sample ow to 100–150cc/min.
4. Turn the heater power switch SW-1, to “ON.”
5. Verify that the indicator lamplight and the display on the controller are functioning.
6. Allow sample to ow through the unit to stabilize (about 10 to 15 minutes).
ÂNOTE
The indicator light cycles on and o with the heater.
Adjusting the Controller
The controller has been preprogrammed at the factory. If additional trimming of the
controller is required, the heat range setting should be between 223°F and 226°F (106°C
and 108°C) based on sample ow rate.
When the heater and ow rate are set properly, condensed water should trickle out from
the condensate drain.
If no condensate is seen, reduce the sample ow rate or increase the temperature
setpoint until the rate of condensation is constant.
If droplets of hot boiling water are seen venting from the condensate drain, either
reduce the temperature setpoint or increase the ow rate.
Adjustments should be within the specied ranges for ow rate and temperature
Table des matières
Autres manuels Sentry Unité de contrôle
Manuels Unité de contrôle populaires d'autres marques

Festo
Festo Compact Performance CP-FB6-E Manuel de la liste des pièces

Elo TouchSystems
Elo TouchSystems DMS-SA19P-EXTME Manuel utilisateur

JS Automation
JS Automation MPC3034A Manuel utilisateur

JAUDT
JAUDT SW GII 6406 Series Guide rapide

Spektrum
Spektrum Air Module System Manuel utilisateur

BOC Edwards
BOC Edwards Q Series Manuel utilisateur