10
Modules
5.1 Module 1: "standard"
This module must always be configured first and must be used only once.
Output data:
a: Pre-selection heating [P001]: 0 – off, 1 – on
b: Stand by [P002]: 0 – off, 1 – on
c: Error acknowledgement [P009]: only when a malfunction is present to 1, otherwise
0
d: Pumping station [P010]: 0 – off, 1 – on
e: Motor [P023]: 0 – off, 1 – on
f: Operating mode pump [P026]: 0 – final rotation speed mode, 1 – rotation speed set-
ting mode (see pos. h:)
g: Run-up time monitoring [P004]: 0 – off, 1 – on
h: Set rotation speed as a % of the final rotation speed [rounded P707], only then
valid in rotation speed setting mode (f: is switched to 1)
i: Rotation speed switch point as a % of the final rotation speed [P701]
Input data:
a: Unit remote controlled [P300]: 0 – no, 1 – yes
b: Switchpoint attained [P302]: 0 – no, 1 – yes
c: Set rotation speed attained [P306]: 0 – no, 1 – yes
d: Pump accelerates [P307]: 0 – no, 1 – yes
e: TMS controller responded [P333]: 0 – no, 1 – yes
f: Actual rotation speed as a % of the final rotation speed
g: Error/Warning number (low byte): 0 – no error, 1...999 – error 1...999, 1001...1999 –
warning 1...999
h: Error/Warning number (high byte)
i: Oil deficiency
j: Excess temperature [P304, P305]: 0 – no, 1 – yes
Configuration data:
C1h, 02h, 83h, 01h
PLEASE NOTE
☞
The fields "h" and "i" must always be filled with valid values (for example 50), even though
these functions are not used.
Byte 012
Bit 76543210
abcdefg - h i
Byte 012 3
Bit 76543210 765 4... 0
abcde - - - f g i j - h