MPM380 Piezoresistive Pressure Transducer
6
b. Match the orientation groove on the plug core to the protuberance on the socket lightly;
c. Screw the screw cap with fingers, then finish the whole connection.
Note:
1. When weld wires, please do not take too much Tin. Cover wires with plastic casing to
prevent short.
2. Please use cable ф5~6mm to connect, this would make seal reliably. Please do not
connect with 2 single wires to connect, for the vapor would go into the socket through
the cable to cause short and corrode.
7. Calibration
We have checked the transducer according to product standard and user’s requirements
when it is delivered from our factory. The user could check as the following steps:
7.1 Zero output check
Install the transducer onto pressure gauge, the pressure is zero. Then do electric
connection as item 6.3 and power on.
The available zero output error is 0±2mV.
7.2 FS output check
a. Calibrate pressure gauge, make the pressure on the gauge is the same as transducer
FS pressure.
b. transducer FS output should confirm to regulated requirements.
c. If the user wants to check the output at other pressure point, please make the pressure
on the gauge be equal with the measuring pressure.
7.3 Condition
Please fulfill the following conditions before check:
a. please be sure transducer’s installation and electric connection confirm to this
operation manual;
b. please be sure the accuracy of pressure gauge is higher min. 1~2 class to transducer;
c. please check the transducer at standard conditions.
8. Operation, Maintenance and Identification
8.1Operation
Please install the transducer as manual indicates, then operate.
Before operation, please be sure the installation and electric connection are correct.
Power on and operate.
The transducer will work once it is powered on, but the output signal will be more stable
and reliable after 30 minutes.
8.2 Maintenance
Please pay attention to the following contents for everyday maintenance:
a. Please check the electric connection often, and make sure it is reliable.
b. Please do not poke or pressure the diaphragm with metal wire or sharp objects.
c. Please clean flush diaphragm with impregnant, and please do not peel off or knock off