Metal Samples Company MS3500E Manuel utilisateur

MS3500E/3510E
Remote ER Data Logger
MetalSamplesCompany
A Division of Alabama Specialty Products, Inc.
152 Metal Samples Rd., Munford, AL 36268 Phone: (256) 358-4202 Fax: (256) 358-4515
E-mail: [email protected] Internet: www.metalsamples.com

.

Table of Contents
I. Introduction ......................................................................................................................1
A. InstrumentFunctionOverview .................................................................................1
B. InstrumentPartsList................................................................................................ 2
II. Installation ........................................................................................................................3
A. EnclosureandMountingDetails...............................................................................3
B. BatteryInstallation................................................................................................... 4
C.4-20mALoopInstallation ........................................................................................5
D. Probe-InstrumentConnection.................................................................................. 6
E. DataDownloadConnection(s)................................................................................. 7
III. SetupandOperation .........................................................................................................8
A. Main Menu Overview ............................................................................................. 8
B. Program Setup ...................................................................................................... 10
1. Set Time and Date ..................................................................................... 10
2. Enter Probe ID .......................................................................................... 11
3. StartDataLogging..................................................................................... 12
C. Measure ProbeFunction ....................................................................................... 13
D. ViewStored DataFunctions.................................................................................. 14
E. SetReadingIntervalFunction ................................................................................ 15
F. DownloadingData ................................................................................................. 16
1. Upload to PC ............................................................................................ 16
2. Upload to MS1500E ................................................................................. 25
G. SystemShutdownFunction.................................................................................... 26
H. TestMode Function .............................................................................................. 26
I. DataInterpretation ................................................................................................. 27
1. ProbeReading........................................................................................... 27
2. Metal Loss ................................................................................................ 28
3. Corrosion Rate .......................................................................................... 28
4. 4-20mA Output ......................................................................................... 29

IV. ERTechnique-General .................................................................................................. 30
A. Measurement Theory ............................................................................................ 30
B. ERSensingElements ............................................................................................. 32
C. ERElementPerformance ...................................................................................... 34
D. ERProbe Body Design ......................................................................................... 38
AppendixA-MS3500ESpecifications ................................................................................... 41
AppendixB-Testingthe MS3500E/MS3510E with theERMeterProver .............................. 42
Appendix C - MS1500E Select Probe ID............................................................................... 44
AppendixD-ReturninganInstrument for Repair .................................................................... 47
AppendixE-Warranty .......................................................................................................... 49
AppendixF-InstallationDrawingofMS3500EinHazardousLocation ................................... 50

1
I. Introduction
A. InstrumentFunctionOverview
TheMS3500E/3510EDataLoggerisdesignedto provide a continuous record ofcorrosiveactivityin
remotelocationsthatareinfrequentlyvisited, such ascross-countrypipelines,andunmannedproduction
platforms.However,thisunitfindsequalapplicationinlocationsthatareinconvenientordifficultto
accessona regular basis, suchasrefineryoverheads,andflue-gasstacks.
TheMS3500E/3510Eiscompletelyself-containedwiththeonboardbatterysystemsupplyingthetotal
powerrequirementfor operation. It canbeusedwithanyof Metal Samples’ ERprobes,orthoseofall
othermajormanufacturers.Onceinplace,itwillautomaticallyreadtheprobeatcustomer-selected
intervalsandstoretheresultingdataintheunit’sonboardmemory.Theonboardmemorywillcollectup
to3,100datapointsbeforedata download is required. The datacollectionintervalisprogrammablein
hourlyincrements.Evenattheminimumdatacollectioninterval(1hour),datadownloadingneedonlybe
performedeveryninetydays.Withmore practical data collection intervals (8 - 12 hours), the unit may
beunattendedfor as longas6-8months between downloading operations.
Anoptical,infrared,RS232communicationlink is provided for datadownloadtoeitheraPCorMetal
Samples'MS1500E portable ERdatalogger.Downloadeddata may beanalyzed,reviewed,or re-
portedbyconventionalspreadsheet,database, or mathematical software packages.
Theinfraredcommunicationlink isanintegralpartoftheintrinsicallysafedesignoftheMS3500E/
3510E.Opticalinsteadofelectrical,thisuniquefeaturepermitsdatadownloadingwithoutremovingthe
instrumentor“memorymodule” fromthehazardousarea.
AnotheruniquefeatureoftheMS3500E/3510Eis the highlevelofonboardintelligence.Thetwo-line,
20-characterLCDallowsvisualreviewofallhistoricaldatainmemoryandreads directly as “total metal
loss”and/or“averagecorrosionrate”. Thisisincontrasttoothercommerciallyavailableunits that store
anddisplayonlytheraw “resistanceratio”thatrequiresadditionalmanipulationtoprovideusefuland
intelligibledata.
TheLCDscreen,togetherwiththe2-keymembranekeypad,provides a user-friendly,interactive,
promptingsystemthatis used for bothsystemsetupanddatareview.ThismakestheMS3500Ethe
mostadvancedunitofits type on the market.
Byaddinga4-20mAcommunication facility, the MS3500E becomes an MS3510E.Thisallows
continuoustransmissionofdata, via a 4-20mAloop,to a plant computer orcentraldatalogger, for
integrationwithotherreal-timeprocessparameters.This data transmissioncanbeaccomplishedwithout
disruptionoftheunit’s basic logginganddatastorageoperations.The 4-20mAloopextends the capa-
bilitiesoftheunittoincludeconventional,in-plant,real-timedatacommunication.
TheMS3500E/3510Eunitusesastainlesssteel NEMA 4X (IP 66) enclosure, makingitsuitableforuse
inthemostextremeofoutdoorconditions.

2
B.InstrumentPartsList
TheMS3500E/MS3510Eissuppliedwiththefollowingaccessories.
Partno. Qty. Description
ET0133 6 1.5V,AADryCells(Installed)
ET0612 1 4-20maLoop Connector (MS3510EOnly)
IN3500-2 1 OptilinkCable
IN3500-4 1 10 Ft. Probe Cable
IN3500-5 1 MeterProver
IN3500-6 1 SerialPortAdapter
IN3500-7 1 OperationsManual
Uponreceiptoftheinstrument,theusershouldverifythattheabovelistedaccessoriesareincluded.
Anyshortageshouldbereportedimmediately to:
Metal Samples
Phone:(256) 358-4202
Fax:(256) 358-4515
E-mail:[email protected]

3
II. Installation
A.Enclosure&MountingDetails
TheMS3500E/3510Eenclosureisfabricatedin carbon steel with a baked-on,high-buildepoxy
coating.Theunithasahinged door that is secured,whenshut,byscrew-downclampsthatdiscourage
unauthorizedtampering.TheenclosureisconstructedtoNEMA4X(IP66)weatherproofstandards,
makingitsuitableforthe mostrigorousofoutdoorenvironments.
Theoveralldimensionsoftheunitare12-1/4"(height),9"(width),4-3/8"(depth),makingitaconve-
nientsizetomounteveninthemostconfinedoflocations.Forthebestresults, the instrument should be
mountedwithintenfeetoftheprobelocation.Tofacilitatethis,theunit hasaversatilemountingflange
withfour.31"diameter,mountingholes.Overalldimensionsandmountingholecentersareshownin
Figure1.
Figure 1. MS3500E Dimensions

4
B.BatteryInstallation
TheMS3500E/3510Einstrument uses six 1.5V,AAsize batteries,connectedinseries,asaprimary
powersource. While anyAAsize battery canbeused,Duracell or ProcellTypeMN1500 or PC1500
batteriesarerequiredtomeettheintrinsicsafetycertificationrequirements.SincetheMS3500E/3510E
operatesinalowpowerconsumptionmodebetweenreadings,batterylifewillbeafunctionofthe
readingintervalandmeasurementtimecyclesetbytheuser.Atthemostfrequentreadinginterval(1per
hour),aminimumbatterylifeoffourmonthsistobeexpected.Morepracticalintervals,suchasevery
12or24hours,willyieldabatterylifeinexcessoftwelvemonths.Generally, to avoid unexpected
powerdrains,batteriesshouldbe replaced once everysixmonths.Intheinstanceofalowreading
interval(1per4hoursorless),threemonthlybatteryreplacementsisadvisable.
Batteriesaremountedin the back of theMS3500E/MS3510E,behindtheinstrumentscreenand
keyboardpanel. Thebatteries are accessed bylooseningthetwothumbscrewsonthe right sideofthe
panel,thenswingingthehingedpanelopentotheleft. Careshouldbeusedwhenopeningandclosing
thepaneltoensure that the internal wiringisnotpinchedordamaged. Oncethepanelisopened,
batteriescanbeinserted or removed fromthespring-loadedretainerasshownbelow.
Figure2. MS3500E BatteryInstallation
+
-
+
-
+
-
+
-
+
-
+
-

5
C.4-20mALoopInstallation
TheMS3510Eisprovidedwitha4-20mAcommunication outletthatallowscontinuous,real-time
integrationofERcorrosiondatawithotherprocessparametersinthemainplantcomputer.The
4-20mAoutlet port isa2-pinmilitaryconnector (MIL 14S-9SF)onthe base of theunit,marked
4-20mA.Thematingconnector for this outlet (ET0612)issuppliedaspartoftheaccessoriesfor the
MS3510Eunit.Thepositiveleadofthecommunicationloopshouldbeconnectedtothe“A” terminal of
theET0612mating connector,and the negative leadtothe“B”.Theloopcanthen be connected tothe
4-20mAinstrumentoutlet:
Figure 3. MS3510ECurrentLoopTermination
Theloop should bepoweredby 10-24 VDC. Forhazardousapplications,an appropriate Zener
barriershouldbeinstalledintheloop,orpowershouldbesuppliedfromanintrinsicallysaferepeater
powersupply. Typicalhazardousinstallationisshownonpage49.
Corrosion Monitoring Systems
Metal Samples
PIN 'A' (+ CURRENT LOOP)
PIN 'B' (- CURRENT LOOP)
VIEW A-A
2-PIN MALE CONNECTOR
(VIEWED FROM SOLDER-SIDE)
AB
AA
MS3510E Connections
PROBE 4-20
mA
OPTI-LINK
DATA PORT

6
D.Probe-InstrumentConnection
TheMS3500E/3510E canbemounted directly ontoanER probe, oronto an HA700110-5 probe
adaptorinthecase of a retrievable probeusedinatwo-inch,high-pressure,accessfitting(see Figure
12,p.41).However,inmanyinstancesitiseitherinconvenientorundesirable to mount theinstrument
directlyontothe probe. Forthesecircumstancesa546810,ten-foot, extension cable isprovided.The
extensioncablehasintegralconnectors withaguidingkeywaytoensurecorrectpinalignment when
attached.Thefemaleconnector (MS3106E145-6S) mateswiththe probe, orHA700110-5probe
adaptor,andthe male connector mateswiththecentralterminalonthebase of the MS3500E/3510E
unit.
The length of the extension cable has been carefully determined so as to avoid excessive
signalnoise. Longerextension cablesshouldonly beemployed afterconsultationwith Metal
Samples.
TheextensioncableprovidedwillmatedirectlywithanyMetalSamples’ERprobeorthoseofany
othermajor manufacturer,orwithanHA700110-5 probe adaptor inthecaseofretrievable probes.