GE Sensing RPT 301 Manuel utilisateur

RPT 301
Digital Output Pressure Transducer
User Manual K177
GE
Sensing

© General Electric Company. All rights reserved.

K177 Issue No. 3
RPT 301
Digital Output Pressure Transducer
User Manual
K177
© Druck Limited 1998
This document is the property of Druck Limited and may not, either in part or whole, be copied or
otherwise reproduced, communicated in any way to third parties, nor stored in any Data Processing
System, without the express written authority of Druck Limited.

K177 Issue No. 3

RPT 301 User Manual i
K177 Issue No. 3
Safety
The manufacturer has designed this instrument to be safe when operated
using the procedures detailed in this manual. Do not use this instrument for
any other purpose than that stated. Do not apply values greater than the
maximum value stated.
This manual contains operating and safety instructions that must be
followed for safe operation and to maintain the instrument in a safe
condition. The safety instructions are either warnings or cautions issued to
protect the user and the equipment from injury or damage. Use qualified*
technicians and good engineering practice for all procedures in this manual.
Pressure
Do not apply pressure greater the maximum safe working pressure to the
instrument.
Electrical Safety
This instrument is designed to be safe when using options and accessories
supplied by the manufacturer.
Toxic Materials
There are no known toxic materials used in this instrument.
Maintenance
The instrument must be maintained using the manufacturer’s procedures
and should be carried out by authorised service agents or the manufacturer’s
service departments.
Technical Advice
For technical advice contact the manufacturer or subsidiary refer to section
5.5.
*A qualified technician must have the necessary technical knowledge,
documentation, special test equipment and tools to carry out the required
work on this instrument.

K177 Issue No. 3
ii RPT 301 User Manual
Abbreviations and Symbols
The following abbreviations are used in this manual.
Note:
Abbreviations are the same in the singular and plural.
EMC electromagnetic compatibility
FS full-scale
IEC International Electrical Council
inHg inch of mercury
kg/cm2kilogram per centimetre squared
kPa kilo Pascal
mA milli Ampere
mm milli metre
NL, H & R non-linearity, hysteresis and repeatability
PCB printed circuit board
psi pound per square inch
RPT resonant pressure transducer
V Volt
VA Volt amp
Hazard, a warning highlighting a danger that would result in
injury. Also a marking on the instrument referring the user to
this publication.
This product meets the essential protection requirements of
the relevant EEC directives. Further details of applied stand-
ards may be found in the product specification.
CE
The following symbols are used in this manual
This symbol, on the instrument or in the manual, indicates a
static sensitive component.

RPT 301 User Manual iii
K177 Issue No. 3
Table of Contents
1 Introduction ................................... 1
1.1 Description ................................................ 1
1.2 Specification .............................................. 1
2 Installation ..................................... 4
2.1 Mounting the Transducer ......................... 4
2.2 Pressure Connection ................................ 4
2.3 Electrical Connection ................................ 6
3 Operation ....................................... 11
3.1 System Operation ..................................... 11
3.2 Command Summary ................................. 11
3.3 General Command Format ....................... 13
3.4 General Reply Format ............................... 15
3.5 Error Message .......................................... 15
3.6 Detailed Command Definitions ................ 17
4 Calibration ..................................... 23
5 Maintenance .................................. 26
5.1 General ...................................................... 26
5.2 LED Indicator ............................................ 26
5.3 Diagnostic Mode ....................................... 26
5.4 Cleaning .................................................... 27
5.5 Approved Service Agents ......................... 28

K177 Issue No. 3
iv RPT 301 User Manual

K177 Issue No. 3
RPT 301 User Manual 1
1 Introduction
1.1 Description
The RPT 301 Digital Output Pressure Transducer measures air pressure
applied to the connection port and produces an electrical output. It
comprises a resonant sensor connected to a printed circuit board to
produce a serial digital output and is contained in a metal enclosure to
IP65 standard. The instrument is designed to communicate with
compatible equipment using a serial communications system which
can be selected by the user for either RS232 or RS485 standard
communications interface. The power supply for the instrument is a
wide tolerance DC voltage with a low power consumption A remote
start-up facility enables the instrument to be remotely switched on and
off to save power.
1.2 Specification
Accuracy
Non-linearity, hysteresis and repeatability:
Standard ................................................................. ± 0.02% FS
Option A .................................................................. ±0.01% FS
Temperature affects: ............... ±0.02% FS over operating temperature range
Long term stability:
Standard ............................................. 150 ppm per 12 months
Option A ............................................. 100 ppm per 12 months
Ranges
.................................................................. 750 to 1150 mbar a
.................................................................... 35 to 1300 mbar a
.................................................................... 35 to 2620 mbar a
.................................................................... 35 to 3500 mbar a
Dimensions
Weight: ...................................................................................... 0.65 kg
Size: ...................................................... 125 mm x 80 mm x 57 mm
Environment
Operating temperature range ..................................................... -20 to + 60C
Storage temperature range ......................................................... -40 to + 80C
Sealing: ........................................................................................... IP65

K177 Issue No. 3
2 RPT 301 User Manual
Power supply
Electrical Safety: .......................................... EN61010 (IEC 1010) as applicable
Voltage .............................................................. 4.5 to 32 V DC
Power .......................... 0.6 Watts at 5 V to 0.75 Watts at 32 V
Electromagnetic compatibility
............................................................ EN50081-1 (emissions)
............................................................. EN50082-2 (immunity)
Switch on
Settling Time: ... Less than 5 seconds (with filtering disabled)
Pressure measurement
Measurement cycle: ........................................ Fixed, nominally 0.5 seconds
Reading Filter:
Single pole low pass filter, variable time constant and step response
Reading Resolution: ........................ Programmable number of decimal places
Data Communications:
Serial RS232 or RS485 (selectable)
Software or hardware handshaking on RS232
software handshaking only on RS485.
Power Control
(Remote Start-Up) A floating ‘volt free’ two-wire control input is provided
to enable the instrument to be remotely switched on
or off.
Pressure connection
.............................................................................. G1/8 female
overpressure ................................... 1.25 x calibrated full-scale
pressure containment ................................................... 5 bar a
Pressure medium compatibility: .......................
Dry gases compatible with:
Glass, Silicon, Titanium, Stainless Steel and Epoxy adhesive
Note: Continuing development sometimes necessitates specification changes
without notice.
Table des matières
Autres manuels GE Transducteur

GE
GE CModel C-RS Manuel utilisateur

GE
GE BAKER HUGHES DPS 5000 Series Manuel utilisateur

GE
GE C-PT Manuel utilisateur

GE
GE T17 Manuel utilisateur

GE
GE CF-LP Manuel utilisateur

GE
GE Baker Hughes Masoneilan 4411 Manuel utilisateur

GE
GE Bently Nevada 7200 Series Manuel utilisateur

GE
GE ApexPro T9 Manuel utilisateur
Manuels Transducteur populaires d'autres marques

Mianyang Weibo Electronic
Mianyang Weibo Electronic WB Series Manuel utilisateur

ProMinent
ProMinent Dulcometer DMT Manuel utilisateur

MKS
MKS MicroPirani 925 Series Comment utiliser

WIKA
WIKA WU-20 Manuel utilisateur

Alcatel Vacuum Technology
Alcatel Vacuum Technology BARATRON 622A Manuel utilisateur

Camille Bauer
Camille Bauer SIRAX CH-5610 Manuel utilisateur