Flowserve SEM-100SS Manuel du propriétaire

Experience In Motion
SEM-100SS
Electronic Flow Meter
PUIOM000649-00 (ENGLISH, REV 07-2022)
INSTALLATION OPERATION MAINTENANCE

Flowserve.com 2PUIOM000649-00-UIM (EN/AQ) August 2022
SEM-100SS User Instructions
Contents
1. Introduction and safety .................3
1.1 General .....................................3
1.2 Disclaimer .................................4
1.3 Copyright ..................................4
1.4 Duty conditions.........................5
1.5 Safety ........................................5
2. Transport and storage ...................6
2.1 Consignment receipt and
unpacking..................................6
2.2 Storage .....................................6
2.3 Recycling at end of
product life ................................6
3. Description.....................................7
3.1 General product description ...7
3.2 Operating parameters ..............8
3.3 Available equipment and
accessories ...............................8
3.4 Materials of construction ..........9
4. Installation and operation.............9
4.1 Unpacking the meter .................9
4.2 Installation .................................9
4.3 Calibration and operation ........10
5. Maintenance................................. 10
5.1 Battery removal and
replacement ...........................10
5.2 Meter chamber removal...........10
5.3 Meter assembly procedure......10
6. Parts list and parts diagram....... 11

Flowserve.com 3PUIOM000649-00-UIM (EN/AQ) August 2022
SEM-100SS User Instructions
1. Introduction and safety
1.1 General
These instructions must always be kept close to the
product’s operating location or directly with the product.
Flowserve products are designed, developed and manufactured with
state-of-the-art technologies in modern facilities. The unit is produced
with great care and commitment to continuous quality control, utilizing
sophisticated quality techniques and safety requirements.
Flowserve is committed to continuous quality improvement and
being at service for any further information about the product in its
installation and operation or about its support products, repair and
diagnostic services.
These instructions are intended to facilitate familiarization with the
product and its permitted use. Operating the product in compliance
with these instructions is important to help ensure reliability in service
and avoid risks. The instructions may not take into account local
regulations; ensure such regulations are observed by all, including
those installing the product. Always coordinate repair activity with
operations personnel, and follow all plant safety requirements and
applicable safety and health laws and regulations.
These instructions must be read prior to installing, operating, using
and maintaining the equipment in any region worldwide. The equipment
must not be put into service until all the conditions relating to safety
noted in the instructions have been met.

Flowserve.com 4PUIOM000649-00-UIM (EN/AQ) August 2022
SEM-100SS User Instructions
1.2 Disclaimer
Information in these User Instructions is believed to be reliable. In spite of
all the efforts of Flowserve to provide sound and all necessary information,
the content of this manual may appear insufficient and is not guaranteed by
Flowserve as to its completeness or accuracy.
Flowserve manufactures products to exacting International Quality
Management System Standards as certied and audited by external
Quality Assurance organizations. Genuine parts and accessories have
been designed, tested and incorporated into the products to help ensure
their continued product quality and performance in use. As Flowserve
cannot test parts and accessories sourced from other vendors, the
incorrect incorporation of such parts and accessories may adversely
affect the performance and safety features of the products. The failure to
properly select, install or use authorized Flowserve parts and accessories
is considered to be misuse. Damage or failure caused by misuse is
not covered by the Flowserve warranty. In addition, any modication of
Flowserve products or removal of original components may impair the
safety of these products in their use.
1.3 Copyright
All rights reserved. No part of these instructions may be reproduced, stored
in a retrieval system, or transmitted in any form or by any means without
prior permission of Flowserve.

Flowserve.com 5PUIOM000649-00-UIM (EN/AQ) August 2022
SEM-100SS User Instructions
1.4 Duty conditions
This product has been selected to meet the specications of your purchase
order. The acknowledgement of these conditions has been sent separately
to the Purchaser. A copy should be kept with these instructions.
The product must not be operated beyond the parameters specified for
the application. If there is any doubt as to the suitability of the product for the
application intended, contact Flowserve for advice, quoting the serial number.
1.5 Safety
1.5.1 Summary of safety markings: These User Instructions contain
specic safety markings where non-observance of an instruction would
cause hazards. The specic safety markings are:
This symbol indicates safety instructions where non-compliance would
affect personal safety and could result in loss of life.
1.5.2 Personnel qualication and training: All personnel involved in
the operation, installation, inspection and maintenance of the unit must be
qualied to carry out the work involved. If the personnel in question do not
already possess the necessary knowledge and skill, appropriate training and
instruction must be provided. If required, the operator may commission the
manufacturer/supplier to provide applicable training.
Always coordinate repair activity with operations and health and safety
personnel, and follow all plant safety requirements and applicable safety and
health laws and regulations.
PUMP IS NOT TO BE USED IN POTENTIALLY EXPLOSIVE ATMOSPHERES.

Flowserve.com 6PUIOM000649-00-UIM (EN/AQ) August 2022
SEM-100SS User Instructions
2. Transport and storage
2.1 Consignment receipt and unpacking
Immediately after receipt of the equipment, it must be veried for contents
and inspected for damages in transportation. Any shortage and/or damage
must be reported immediately to distributor/Flowserve Scienco and must
be received within 10 days of receipt of the equipment. Later claims cannot
be accepted.
Check any boxes or wrappings for any accessories or spare parts that
may be packed separately with the equipment.
Each product has a unique serial number. Check that this number corresponds
with that advised, and always quote this number in correspondence as well as
when ordering spare parts or further accessories.
2.2 Storage
Normal packaging is designed to protect the pump and parts during
shipment and storing.
2.3 Recycling at end of product life
At the end of the service life of the product, the relevant materials and parts
should be recycled or disposed of using an environmentally acceptable
method and in accordance with local regulations.
Make sure that hazardous substances are disposed of safely
and that the correct personal protective equipment is used. The
safety specifications must be in accordance with the current local
regulations at all times.

Flowserve.com 7PUIOM000649-00-UIM (EN/AQ) August 2022
SEM-100SS User Instructions
3. Description
3.1 Configurations
The SEM-100SS electronic ow meter employs positive displacement
nutating disk technology to meter liquids of various viscosities with great
accuracy and nominal pressure loss. Stainless steel and plastic construction
make the SEM-100SS an extremely versatile meter, suitable for use with
many chemicals and petroleum products.
The SEM-100SS ow meter features a large LCD display, an anti-glare
screen with back light, large positive touch buttons, and simple one-touch
operation, making the meters easy to read and easy to use in both high
light and low light settings.
The SEM-100SS ow meter is a positive displacement, nutating disk, uid
metering unit. As uid ows through the unit, it causes the nutating disk
to move which in turn causes the rotation of a hermetically sealed magnet
disk. The meter electronic assembly counts the number of revolutions of the
magnet and mathematically determines, based on the calibration number,
the quantity of liquid that has passed through the meter. The self-contained
electronic assembly features a large Liquid Crystal Display (LCD), and
simple four-button operation.

Flowserve.com 8PUIOM000649-00-UIM (EN/AQ) August 2022
SEM-100SS User Instructions
3.2 Operating parameters
• Flows from 7.5 lpm to 110 lpm (2 gpm to 30 gpm)
• Temperatures up to 85°C (185°F)
• Pressures up to 6.9 bar (100 psi)
3.3 Available equipment and accessories
The SEM-100 and SEM-100FT electronic ow meters are available individually
or in packages with the CT6 chemical transfer pump in multiple congurations,
each with varying equipment and accessories, including:
• Standard and special hoses
• Recirculation piping
• Various ttings and adapter
• Brackets
• Dispensing nozzles
• Dip-tubes

Flowserve.com 9PUIOM000649-00-UIM (EN/AQ) August 2022
SEM-100SS User Instructions
3.4 Materials of construction
The SEM-100SS electronic ow meters are resistant to corrosion from
typical ag-chem and light industrial chemicals. Wetted materials include:
• Fluid Housing/Form Cover ------- 316 Stainless Steel
• Meter Chamber ------- Polypropylene Sulde (PPS),
Stainless Steel
• O-rings ------- Viton
• Electronic Housing ------- Nylon
• Screws ------- Stainless Steel
Note: Ensure chemical compatibility between liquid metered and the
meter’s wetted parts before using
Note: Meter should not be used to transfer ammable petroleum products.
DO NOT use with uids that have a ashpoint lower than 37°C (100°F).
Refer to the Scienco Product Reference Manual FPD-1607 to learn
more about the available configurations.
® Viton is a registered trademark of the DuPont Company.

Flowserve.com 10PUIOM000649-00-UIM (EN/AQ) August 2022
SEM-100SS User Instructions
4. Installation and operation
4.1 Unpacking the meter
Unpack the meter, and ensure that meter conguration and materials of
construction are correct for your application.
4.2 Installation
Flow can be metered in either direction through the SEM-100SS ow meter.
Thread the meter inlet into the female discharge port of the pump or line.
Using an elbow or coupling, screw the discharge hose onto the other
port of the meter. Use thread sealant tape or pipe dope to seal thread
connections as necessary.
DO NOT over tighten threaded connections. Breakage can
occur resulting in fluid exposure and leakage.
DO NOT mount meter in a load bearing manner. Excessive
loads can cause breakage, resulting in fluid exposure and leakage.
Table des matières
Autres manuels Flowserve Instrument de mesure
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur