Specifications of Measurement Data Management Software:
Supported OS:Japanese version of Windows® 10 64-bit OS
English version of Windows® 10 64-bit OS (Note 2, Note 3)
Monitor screen display: Chart display 0~5,000ppm (Max. 65,535ppm),
CO2 concentrations (0~65,535ppm) (Note 1)
Data file format: CSV file
Data recording interval: 2 seconds, 30 seconds, 60 seconds, 5 minutes, 10 minutes,
30 minutes, or 60 minutes
Data recording capacity: Measurement data for about 24 hours max. in a single file.
Note 1)For specifications of the CDM7162, see the CDM7162 Product Information.
Note 2)Windows and Excel are registered trademarks of Microsoft Corporation.
Note 3)Although the operation of the measurement data management software was tested on the
above operating systems, proper operation of this software is not guaranteed.
3. Setting the Method of Serial Communication
The I2C bus or UART interface can be chosen for serial communication between the CDM7162 CO2
sensor and this evaluation module.
Set the serial communication method by using the communication mode selection switches on the
EM7162 module before connecting the USB cable or the power supply.
Caution‼ Before changing the serial communication setting, be sure to turn off the power
switch (SW1) by setting it to the lower position as shown in Fig. 2.
3-1) Switch settings for I2C bus mode
Set switches SW2, SW3, and SW4 (1 to 4) as follows:
SW2, SW3: Set to I2C position (lower position as seen in Figure 2).
SW4-1, SW4-2: Set to ON position (upper position as seen in Figure 2).
SW4-3, SW4-4: Set to OFF position (lower position as seen in Figure 2).
Figure 2 - I2C Communication Setting
Caution‼ Do not change the preset positions of SW2, SW3, and SW4. Changing switch
positions may cause damage to the CDM7162 sensor.