Blackrock Microsystems Cerebus Manuel utilisateur

630 Komas Drive |Suite 200
Salt Lake City |UT 84108 |USA
P +1 801.582.5533 |F +1 801.582.1509
www.blackrockmicro.com
Revision 15.00 / LB-0028 –Cerebus Instructions for Use
Cerebus, CerePlex, CerePort, and NeuroPort are registered and unregistered trademarks of Blackrock
Microsystems
Cerebus
Instructions for Use

Revision 15.00 / LB-0028 –Cerebus Instructions for Use
2
Contents
Introduction................................................................................... 4
System Description ................................................................................... 4
Cerebus Models ........................................................................................ 4
Host PC Requirements .............................................................................. 4
Packing Contents....................................................................................... 5
Setup Overview......................................................................................... 5
Cerebus System Hardware ............................................................ 6
Front End Amplifier................................................................................... 7
Amplifier Input .......................................................................................... 7
Analog Headstages.................................................................................... 8
Amplifier Power Supply (APS) ................................................................... 8
Cables ....................................................................................................... 9
Digital Cerebus System Hardware ............................................... 10
Digital Hub 128........................................................................................ 10
Digital Hub Inputs ................................................................... 10
CerePlexTM Digital Headstages................................................................. 11
Neural Signal Processor (NSP) ..................................................... 12
NSP Front Panel ...................................................................................... 13
NSP Back Panel........................................................................................ 15
Quick Setup Guide ....................................................................... 17
Setting up the Cerebus System ............................................................... 17
Setting up the Digital Cerebus System..................................................... 17
Setting up the Ethernet Link.................................................................... 17
Troubleshooting...................................................................................... 18
The NSP is stuck in the initializing step. .................................. 18
How can I update the Cerebus firmware on the NSP to the
latest version? ........................................................................ 19

Revision 15.00 / LB-0028 –Cerebus Instructions for Use
3
The Front End Amplifier power supply red error LED is on..... 20
The analog headstage attached to the Front End Amplifier is
not receiving power................................................................ 20
The fiber optic LED indicator is illuminating red. .................... 21
The source status LED on Digital Hub is illuminating red. ....... 21
The Central application fails to start....................................... 22
Warranty ..................................................................................... 23
Support........................................................................................ 23
Appendix 1................................................................................... 24
Warnings ................................................................................ 24
Cautions.................................................................................. 24
Specifications.......................................................................... 24
Symbols of Contraindications, Warnings, Cautions ................ 25
Appendix 2................................................................................... 26
Electromagnetic Immunity ..................................................... 26

Revision 15.00 / LB-0028 –Cerebus Instructions for Use
4
Introduction
The Blackrock Cerebus Neural Processing System supports recording, processing and displaying bio-
potential signals from various types of electrodes. Bio-potential signals may include Electrocorticography
(ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG),
electrooculography (EOC), action potentials (AP), and evoked potentials (EP).
System Description
The Cerebus System records and processes neural signals from up to 128 surface or penetrating
electrodes in addition to auxiliary analog signals and digital experimental events. Multiple systems can be
synchronized to achieve higher input channel counts.
The system can perform real-time signal processing algorithms including noise cancellation, digital
filtering, simultaneous extraction of spike and field potentials, manual and automatic online spike sorting.
Functionality can be further increased with custom user-defined scripts.
Cerebus Models
The Cerebus System is offered in the following two versions.
1. Cerebus System:
The Cerebus System is packaged with a Front End Amplifier (FEA). The Front End Amplifier is
compatible with Blackrock’s analog headstages which provide an isolated low-impedance output
that ensures signal integrity and minimizes tether artifact.
2. Digital Cerebus System:
The Digital Cerebus System includes a Digital Hub. The Digital Hub allows the Cerebus System to
be used with CerePlex™ digital headstages which digitize and multiplex neural signals to reduce
lead wire size and noise acquisition.
Host PC Requirements
The recommended specifications for a Cerebus System host PC are mentioned below.
•Microsoft Windows 7 or 10 professional
•Intel Core i-series processor
•8 GB RAM
•250 GB SSD for operating system
•Separate SSD for data storage
•Discrete video/graphics card
•2 gigabit Ethernet Ports
A Host PC that is configured and tested by our engineers is available. Please contact
sales@blackrockmicro.com for more information.

Revision 15.00 / LB-0028 –Cerebus Instructions for Use
5
Packing Contents
The Cerebus System is shipped with components listed below. The exact type and quantity of the
components can vary depending on the Cerebus model, line voltage (110v or 220v), and total input
channel count.
•Neural Signal Processor (NSP) and rubber feet
•Rack mounting ears and screws
•For Cerebus System only: Front End Amplifier (FEA) and isolated Amplifier Power Supply
(APS).
•For Digital Cerebus System only: Digital Hub 128 and isolated power supply.
•Alligator clip to touch proof ground cable
•Power cables and connectors
•Crossover Ethernet cable
•Cerebus user’s manual & Central Suite software installation CD
Setup Overview
The following diagrams depict overall views of how the Cerebus and the Digital Cerebus Systems are
integrated within a typical laboratory environment. The experiment subject area is both electrically and
optically isolated from the experiment operation and control area. In the Cerebus System, digitization and
amplification takes place inside the Front End Amplifier. The amplifier receives analog signals either
directly from electrodes or via analog headstages then performs differential amplification and digitization.
In the Digital Cerebus System, the neural signals are digitized by the CerePlex digital headstages before
they enter the Digital Hub. Both the Digital Hub and Front End Amplifier are also responsible for
converting the digitized signals into fiber optic data and transferring them to the processor via a fiber
optic link.

Revision 15.00 / LB-0028 –Cerebus Instructions for Use
6
Figure 1 (lower panel) also depicts a simplified flowchart to describe the signal processing stages that take
place inside the Cerebus system. The signals on both the electrode and reference wire are initially
buffered using unity gain amplifiers. The buffered signals then enter a differential amplifier which
subtracts the neural signal from the reference to suppress common noise and achieve a bipolar recording.
The resultant signal then goes through the gain and filtering stage. The amplified signal is then digitized,
converted to fiber optic signal, and transferred to the processor for further analysis, display or storage.
Cerebus System Hardware
The Cerebus System components are depicted inside the following block diagram. The architecture
generally consists of analog headstages, Front End Amplifier, and the Neural Signal Processor which are
described in the following sections.
Figure 1- Setup schematics showing the Cerebus and Digital Cerebus Systems (top). Overview of signal
processing stages in Cerebus data acquisition system (bottom).

Revision 15.00 / LB-0028 –Cerebus Instructions for Use
7
Figure 2- Simplified block diagram of the Cerebus system
Front End Amplifier
The Front End Amplifier (FEA), depicted in Figure 3, receives analog signals either directly from the
electrodes or via headstages (unity-gain voltage followers) depending on the impedance of the
electrodes. The analog signals are amplified (unity gain), filtered (1st-order high- pass at 0.3 Hz and 3rd-
order low-pass at 7,500 Hz), and digitized (30 kHz, 16-bits at 250 nV resolution), converted into the optical
domain and then transmitted to the NSP via a fiber-optic link, which is immune to electromagnetic field
interference.
The Amplifier ground is not connected to the earth ground. The Amplifier ground is isolated and floating.
There are ESD shunt circuits that will conduct differences of 1000 V or more. The Amplifier can be
connected to earth ground via any one of the four ground pins located on the input side of the Amplifier.
Amplifier Input
Figure 3-Isometric views of the Front End Amplifier showing the front (left) and back side
(right).

Revision 15.00 / LB-0028 –Cerebus Instructions for Use
8
Figure 4- Front End Amplifier front view and pinout
The Amplifier has four 34-pin banks as shown in Figure 4. Each bank consists of 32 channels, a bank
reference pin, and a ground pin. The electrodes within each bank are differentially amplified with respect
to the reference input of the same bank. Connect one reference channel to all banks to use a single
reference for all measurement channels. Four 6-pin Amplifier Power Supply banks provide power (±5V) to
the analog headstages and can deliver up to 130 mA of combined current. The ground pins on every bank
are internally tied together and should be used as the primary experimental ground.
Analog Headstages
Blackrock analog headstages acquire neural signals from electrodes and provide an isolated low-
impedance output to the Front End Amplifier via direct cable connections for each channel. Blackrock
analog headstages ensure maximum common-mode noise rejection by the Front End Amplifier. The high
input impedance and low bias current of the Front End Amplifier inputs make it possible to connect
microelectrodes with 20 kΩ(at 1kHz) or lower impedance values directly to the Front End Amplifier
inputs without the need for a headstage. This configuration makes the application more susceptible to
environmental electromagnetic noise.
To minimize environmental noise, it is recommended to keep direct electrode connections shorter than
20 cm (8 inches). For longer connections, headstages from Blackrock Microsystems are recommended.
Headstages allow microelectrodes with impedances up to 5 MΩ(at 1kHz), and cables up to 6 feet or 3
meters in length.
Amplifier Power Supply (APS)
The APS consists of five analog and digital supply channels with monitoring, sequencing, and emergency
shutdown controls. The On/Off switch is located on the back of the unit. Once it is switched on, the APS

Revision 15.00 / LB-0028 –Cerebus Instructions for Use
9
will start a power-up sequence of analog and digital supplies ending with the green ON LED illuminating.
When the power switch is turned off, this sequence is reversed and the output of the APS is tied to
ground. In the event of an error in voltage or power delivery, the APS will shut down and a red error LED
will turn on. After checking all amplifier and patient cable connections for any inadvertent shorting, turn
off the APS and turn it back on to reset the error.
Cables
Included with the system is a power cable for the NSP (not shown), a power cable for the APS (not
shown), a fiber-optic link that connects the Front End Amplifier to the NSP, a crossover Ethernet cable
that connects the NSP to the Host PC, and a cable that connects the APS to the Front End Amplifier.
Note: The fiber-optic cable is very delicate. Do not bend the cable (with a smaller
bend radius than 5.0 cm) or crush it!
Figure 6- Fiber-optic Cable (Left) and Amplifier Power Supply Cable (Right).
Figure 5- Amplifier Power Supply

Revision 15.00 / LB-0028 –Cerebus Instructions for Use
10
Digital Cerebus System Hardware
The Digital Cerebus System is compatible with the whole family of the Blackrock CerePlex digital
headstages. Blackrock digital headstages enable low noise transmission in a minimally invasive package.
The output of the CerePlex digital headstages are multiplexed, meaning that only a reduced set of wires
are needed to transmit the already digitized data from all the input channels. The following block diagram
depicts the different components and their roles within a Digital Cerebus System.
Digital Hub 128
The Blackrock Digital Hub provides an interface between the NSP and CerePlex digital
headstages. The Digital Hub converts digital signal to an optic-digital format which is sent directly
to the NSP. This dramatically reduces the noise introduced to the signal during transmission. The
Digital Hub can handle many possible configurations when it is connected to multiple CerePlex
digital headstages provided that the total number of channels does not exceed maximum 128
channels.
For more information about different models of CerePlex family of products, please
refer to the Blackrock Microsystems website.
Digital Hub Inputs
Figure 7- Simplified block diagram showing different components of the Digital Cerebus and their
intended function within the system.
Autres manuels pour Cerebus
1
Table des matières
Manuels Matériel informatique populaires d'autres marques

EMC2
EMC2 VNX Series Manuel du propriétaire

Panasonic
Panasonic DV0PM20105 Manuel utilisateur

Mitsubishi Electric
Mitsubishi Electric Q81BD-J61BT11 Manuel utilisateur

Gigabyte
Gigabyte B660M DS3H AX DDR4 Manuel utilisateur

Raidon
Raidon iT2300 Manuel utilisateur

National Instruments
National Instruments PXI-8186 Manuel utilisateur