Teilch TCH Pro 202 Manuel utilisateur

Operation Manual TCH Pro 202 & 203
Handheld Particle Analyzer
Models TCH Pro 202
& TCH Pro 203
Page of1 33
TCH Pro v1.2.1

Operation Manual TCH Pro 202 & 203
Safety Information
This section gives instructions for promoting safe and proper handling of the Particle
Analyzer.
Laser Safety
The Model TCH Pro 20x handheld Particle Analyzer is a Class I laser-based instrument.
•During normal operation, you will not be exposed to laser radiation.
•WARNING: Class IV visible or invisible laser inside. Exposure to this light causes
blindness.
•DO NOT remove any cover or disassemble the device.
Important - Take these precautions
•DO NOT remove any parts from the particle analyzer unless you are specifically told to do
so in this manual.
•DO NOT remove the housing or covers. There are no user serviceable components inside
the housing. All repair and maintenance needs should be done by a qualified technician.
•Take precautions not to get any particles over 50 μm inside the unit. Failure due to
foreign objects found in the unit are not covered by the Warranty.
•Precautions for power use:
•AC Adaptor. The AC adapter accommodates voltage of AC 100~240V and frequency of
50/60 Hz.
•Batteries: 2 units of Li-ion, 3.7V, size 18650.
Page of2 33
TCH Pro v1.2.1

Operation Manual TCH Pro 202 & 203
Table of Contents
CHAPTER 1: PRODUCT OVERVIEW
Introducing TCH PRO MODELS"4
Specifications"6
CHAPTER 2: UNPACKING
Box Content"7
CHAPTER 3: THEORY OF OPERATION
II. Particle Detection"8
III. Calibration"11
CHAPTER 4: OPERATION
I. Getting Started"15
II. Screen Symbols"16
III. Menu Map"17
IV. Functionalities"18
V. Data Management"2 8
VI. Remote Operation"28
CHAPTER 4: TROUBLESHOOTING
CHAPTER 5: MAINTENANCE
I. Recommended Maintenance Schedule"30
II. TCH Pro Firmware Upgrade"30
III. Touchscreen Calibration"31
CHAPTER 6: WARRANTY
Warranty"32
Contact Information"33
Page of3 33
TCH Pro v1.2.1

Operation Manual TCH Pro 202 & 203
Chapter 1: Product Overview
INTRODUCING TCH PRO MODELS
The Handheld Particle Analyzer TCH Pro is a lightweight, handheld particle counter with a
3.5” Color LCD display, a touchscreen and a single button. It is simple to use and maintain.
It operates on battery or AC power. This model series has a 0.1 CFM with internal flow
meter (1% FS) and active flow control counts particle sizes of 0.2μm (50%) to 10μm full and
continuous particle size spectrum by turning the concept of channels obsolete. Over 1,000
data sets can be stored in removable SD card and downloaded for analysis and reporting
using the utility included with the device.
Page of4 33
TCH Pro v1.2.1
Key features:
•TCH PRO is the smallest 0.2μm particle
counter in the market. A truly portable device
•Not bound by channels. Full continuous
particle spectrum
•High power variable light source allows
unprecedented range of 0.2μm to 10μm
•State of the art analog and digital signal
processing
•Advanced particle analysis directly at device
•PM2.5 and PM10 display
•Differential pressure (1 Pa resolution) display
•Embedded flow meter (1% FS)
•Humidity and temperature sensors

Operation Manual TCH Pro 202 & 203
!
Page of5 33
TCH Pro v1.2.1
Back View Bottom-up View
Side Views
Note that the External Sensor Probe
only applies to Teilch Pro 203

Operation Manual TCH Pro 202 & 203
SPECIFICATIONS
!
Main Uses
Cleanroom certification. Air quality and filter
monitoring in household, office, industrial and
healthcare environments
Particle Size Range
0.2μm (50%) to 10μm, continuous
Channels
Unlimited number of channels. Capture size
information for each particle with 10nm resolution
Supported Standards
ISO 21501-4 (Instrument), US FED STD 209E and
ISO 14644-1 (Cleanroom Classification)
Sensor
High sensitivity, ultra-low noise and fully
compensated
Flow
0.1cfm with internal flow meter (1% FS) and active
flow control
Light Source
High power, >10,000hr laser diode with active
control. Full temperature compensation and
stabilization
Pressure
115kPa max, 1Pa noise-free resolution
Humidity
5% to 95%, +-2%
Temperature
0-50C, +-0.1C Internal and external (203) sensors
Display
3.5” Color LCD
Operating Method
Touchscreen and single button
Battery
User replaceable Li-Ion, 5 hr continuous counting,
10hr intermittent counting. Size 18650
Storage
External SD Card
Connection
USB, SD Card
Dimensions
82mm(W)x137mm(H)x42mm(D)
Weight
830gr (w/o accessories)
Page of6 33
TCH Pro v1.2.1

Operation Manual TCH Pro 202 & 203
Chapter 2: Unpacking
Carefully unpack Teilch Handheld Particle Analyzer from its case and verify that all the items
shown in the photos below and listed in the following tables are present. It is recommended
to unpack the particle counter within a cleanroom-type environment. Doing so will minimize
the counter’s exposure to dirt and moisture.
BOX CONTENT
Find below an inventory of the different components of the case . Contact us immediately if
any items are missing or broken.
•TCH Pro 202 Handheld Particle Analyzer
•Power Supply
•Isokinetic Probe
•Barbed Probe
•Inlet Filter
•Hand Strap (for 203 model only)
•External Sensor (for 203 model only)
•SD Memory Card (2GB) - Installed
•USB Cable
•USB Flash Drive (Software, documentation, and
Operation Manual)
•Inlet Caps (x2)
•Case Protector
•Carrying Case
•Inlet O-ring Seal Replacements (x4)
•Calibration Certificate and Quick Start Guide
Customer support in the US:"Customer support in Asia:
265 Cambridge Ave, #60250 "7B Mountain Drive, Light Industry & Science Park II
Palo Alto, CA 94306, USA" " Calamba City, Laguna 4027, Philippines
Ph. (+1) 650 319 3018" " Ph. (+63) 49 545 6957
Page of7 33
TCH Pro v1.2.1

Operation Manual TCH Pro 202 & 203
Chapter 3: Theory of Operation
A clean, contamination-free work environment is critical to maximize yields as certain
particles are likely to cause defects, irregularities, inaccuracies and even failures. Analyzing
particle contaminants is an important step in having control over operating environments
and reducing particle deposition in critical components.
Isolating, characterizing and understanding the types of particles found in products or
processes can usually yield a likely source for the particles. Once the source is known then
elimination of the particulate contamination becomes substantially easier.
II. PARTICLE DETECTION
PARTICLE COUNTING PRINCIPLE
A particle counter is an instrument that detects and counts particles. A particle counter
works on the principle of either light scattering, light blocking, or condensation for detecting
and measuring particle size or size distribution.
When a particle is illuminated by a light beam, light is reemitted (if light scattering is used) or
absorbed (obstruction). The scattered or absorbed light (depending on method used) is
detected by a photo detector, measured, and the particle is counted and tabulated into
standardized counting bins. If direct imaging is used, light illuminates particles from the back
Page of8 33
TCH Pro v1.2.1

Operation Manual TCH Pro 202 & 203
within a cell while a high definition, high magnification camera records passing particles.
Recorded video is then analyzed by computer software to measure particle attributes.
Notes:
•The light blocking optical particle counter method is typically useful for detecting
and sizing particles greater than 1 micrometer in size. This type of technique allows high
resolution and reliable measurement.
•The light scattering method is capable of detecting smaller sizing particles. Typical
detection sensitivity of the light scattering method ranges between 0.1µm to 25µm.
Scattered light is significantly smaller for small particle sizes. A 0.2µm particle emits
1*10-10 of the intensity emitted by a 10µm particle.
•The condensation method is used to detect particles over 5nm, but it cannot
differentiate among different particle sizes.
Page of9 33
TCH Pro v1.2.1
1E-10%
1E-09%
1E-08%
0.0000001%
0.000001%
0.00001%
0.0001%
0.001%
0.01%
0.1%
1%
10%
0.1% 1% 10%
Light%Intensity%
Par6cle%Size%(um)%

Operation Manual TCH Pro 202 & 203
ANATOMY OF A PARTICLE COUNTER’S SENSOR
A particle counter is made up of the following components:
•Light Source (bright laser light)
•Reflector
•Light Sensor
•Electronic Signal Processing Circuit
•Pump
•Flow Sensor
•Filter"
Page of10 33
TCH Pro v1.2.1
Diagram: Anatomy of a particle counter
Ce manuel convient aux modèles suivants
1
Table des matières
Autres manuels Teilch Instrument de mesure
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur