Snap-On EEDM575D Manuel utilisateur

EEDM575D
Clamp Meter
Operating
Instructions

TABLE OF CONTENTS
A. INTRODUCTION
1. Congratulations ..........................3
2. Product Description....................3
3. Declaration of Conformity ..........4
B. SAFETY CONSIDERATIONS ..............5
C. TECHNICAL DATA
1. Features and Benefits..................6
2. Product Applications ..................7
3. Specifications..............................8
D. MEASUREMENT TECHNIQUES
1. Controls and Functions ............12
2. Step by Step Procedures: ........14
E. OTHER FEATURES ..........................25
F. APPLICATION NOTES ......................30
G. TROUBLE SHOOTING GUIDE ..........35
H. MAINTENANCE ................................36
I. ACCESSORIES ................................37
2

A. INTRODUCTION
1. Congratulations!!
Thank you for purchasing Snap-on products. The
EEDM575D is easy to use and is built to last. t is
backed by a 5 year limited warranty. Please
remember to complete and return your product
warranty registration card.
2. Product Description
The slim design EEDM575D is a hand-held,
autoranging clamp-on DMM. Extra large numer-
als, min/max, capacitance, DC microamps and
duty cycle are just a few of the features of the
EEDM575D. An affordable choice, the EEDM575D
offers measurements in all basic electrical func-
tions with additional advanced features like non-
contact voltage.
The EEDM575D comes complete with the follow-
ing accessories:
Carrying Pouch
Test Lead Set
Temperature Probe
Instruction Manual
Battery & Fuse
3

3. EC Declaration of Conformity
This is to certify that Snap-on Model EEDM575D con-
forms to the protection requirements of the council direc-
tive 89/336/EEC, in the approximation of laws of the
member states relating to Electromagnetic compatibility
and 73/23/EEC. The Low Voltage Directive by application
of the following standards:
EN 61326 1997+A1+A2:2001
EN 50082-1 1992 Immunity Standard
EN 61010-1 2001 Safety Standard
EN 61010-2-32 2001 Safety Standard
EN 61010-2-032 1995 Safety Standard
To ensure conformity with these standard, this
instrument must be operated in accordance
with the instructions and specifications given in
this manual.
CAU ION: Even though this instrument complies
with the immunity standards, it’ s accuracy can be
affected by strong radio emissions not covered in
the above standards. Sources such as hand-held
radio transceivers, radio and V transmitters, vehi -
cle radios and cellular phones generate electro -
magnetic radiation that could be induced into the
test leads of this instrument. Care should be taken
to avoid such situations or alternatively, check to
make sure that the instrument is not being influ -
ence by these emissions.
CAU ION: Please follow manufacturers test pro -
cedures whenever possible. Do not attempt to
measure unknown voltages or components until a
complete understanding of the circuit is obtained.
4

B. SAFETY CONSIDERATIONS
WARNING:Please follow manufacturers test pro-
cedures whenever possible. Do not attempt to mea-
sure unknown voltages or components until a com-
plete understanding of the circuit is obtained.
Read instructions before operating:
Be sure these instructions accompany the tool when
passed from one user to a new or inexperienced user.
Equipment is used in a manner not specified by the
manufacturer, the protection provided by the equip-
ment may be impaired.
GENERAL GUIDELINES
ALWAYS
• Test the EEDM575D before using it to make sure it is
operating properly.
• nspect the test leads before using to make sure there are
no breaks or shorts.
• Double check all connections before testing.
• Have someone check on you periodically if
working alone.
• Have a complete understanding of the circuit
being measured.
• Disconnect power to circuit, then connect test leads to
the EEDM575D, then to circuit being measured.
NEVER
• Attempt to measure unknown high voltages.
• Attempt to measure DC microamps with the meter in
parallel to the circuit.
• Connect the test leads to a live circuit before
setting up the instrument.
• Touch any exposed metal part of the
test lead assembly.
5

INTERNATIONAL SYMBOLS
CAUT ON: R SK OF ELECTR C SHOCK
AC (Alternation Current)
DC (Direct Current)
REFER TO NSTRUCT ON MANUAL
GROUND
DOUBLE NSULAT ON
E THER DC OR AC
Application around and removal from HAZ
ARDOUS L VE conductors is permitted.
C. TECHNICAL DATA
1. Features and Benefits
Agency Meets CE and EC 61010-1.
Analog Bar Displays rapidly changing input
Graph signals.
NCV Non-contact voltage detection. Detect
the presence of voltage without con-
tacting the circuit.
P-H Captures min / max peak.
MAX/MIN Records Min/Max readings for all
functions and ranges.
REL Measurements made are displayed
relative to a stored value.
Auto Off Automatically powers off after
30 minutes of inactivity.
6

2. Product Applications
Perform the following tests and/or measurements with
the EEDM575D and the appropriate function:
HVAC/R
DCmV • Thermocouples in furnaces.
ACA • Heat anticipator current in thermostats.
ACV • Line voltage.
ACV or DCV • Control circuit voltage.
DCmmA•Flame safegaurd control current.
OHMS • Compressor winding resistance.
OHMS • Continuity of wiring.
CAP • Motor start and run capacitors.
ACA • Motor and compressor start up current.
Hz • Frequency on controls and line voltage.
TEMP • Air duct temperatures.
ALL • Bar graph to indicate rapid fluctuations.
ELECTRICAL
ACV • Measure line voltage.
ACA • Measure line current.
OHMS • Continuity of circuit breakers.
DCV • Voltage of direct drive DC motors.
ACA • Start up current of motors, relays,
contactors and transformers.
ELECTRONIC
ACV • Measure power supply voltage.
ACA • Measure power supply current.
OHMS • Continuity of circuit breakers and fuses.
7

3. Specifications
IEC 61010-1 Over Voltage:
CAT III 600V UL61010-1
Pollution D gr 2 CAT III 600V
Temperature for gauranteed accuracy: 23˚C +5˚C
DC VOLTS
Range Res. Accuracy
110.00mV 0.01mV mpedance:
1.1000V 0.0001V +(0.5% of reading + 2 digits) 10MW
11.000V 0.001V Overload Protection:
110.00V 0.01V 600VDC or AC RMS
600.0V 0.1V
(45Hz to 100Hz Frequency Response AC 400mV Range)
AC VOLTS (45Hz to 450Hz Frequency Response All Other Ranges)
Range Res. Accuracy
110.00mV 0.01mV +(1.2% of reading + 5 digits) mpedance:
1.1000V 0.0001V +(1.2% of reading + 3 digits) 10MW
11.000V 0.001V Overload Protection:
110.00V 0.01V +(1.5% of reading + 3 digits) 600VDC or AC RMS
600.0V 0.1V
8
(45Hz to 100Hz Frequency Response 200A to 400A)
AC AMPS (45Hz to 450Hz Frequency Response Below 200A)
Range Res. Accuracy
110A 0.01A +(3.0% of reading + 10 digits)
400A 0.1A +(3.0% of reading + 5 digits)

DC AMPS (Manual)
Range Res. Accuracy
110.00A 0.01A +(3.0% of reading + 10 digits)
400.0A 0.1A +(3.0% of reading + 5 digits)
DC Microamps
Range Res. Accuracy
110.00mA 0.01mA +(0.8% of reading + 2 digits) 0.5 A / 600V Fusee
1100.0mA 0.1mAe
* DC microamps are measured with the test leads in series with the
circuit under test.
OHM (Resistance, WW)
Range Res. Accuracy
110.00W0.01W+(1% of reading +5 digits)
1.1000kW0.0001kW
11.000kW0.001kW+(1% of reading +2 digits) 500VDC or AC RMS
110.00kW0.01kW
1.1000kW0.001MW+(1% of reading +5 digits)
11.000MW0.001MW+(1% of reading +10 digits)
110.0MW0.1MW+(1% of reading +20 digits)
Frequency (Hz)
Range Res. Accuracy
111.1Hz 0.1Hz*f1)
1110Hz 1Hz*f1)
11.000kHz 0.001kHz +(1% of reading +2 digits) 500VDC or AC RMS
110.00MHz 0.01MHz
1.1000MHz 0.0001MHz
11.000MHz 0.001MHz Unspecified
110.00MHz 0.01
*f1) If input frequency is less then 11.1Hz will diplay 0.0Hz
9

Diode Test
Test Current Over Load Protection
1.5mA MAX 500 V DC or AC RMS
Continuity Buzzer
Test Voltage Threshold Over Load Protection
3V <35W500 V DC or AC RMS
Non-Contact Voltage (NCV)
Detection Range: 24VAC and above
Temperature ( -Type thermocouple)
Range Res. Accuracy
-30˚F to 1000˚F 0.1˚F +(1˚F) 32.0˚F to 120˚F
(˚F Version) +(1%+1.5˚F) -30˚F to 750.0˚F
+(2%+4˚F) -30˚F to -4˚F
+(2%+4˚F) 750˚F to 1000˚F
-35˚C to 600˚C 0.1˚C +(1˚C) 0˚C to 48.9˚C
(˚C Version) +(1%+1˚C) -19.9˚C to -0.1˚C
+(1%+1˚C) 49.0˚C to 399.9˚C
+(2%+3˚C) -35˚C to -20˚C
10
Warning: Use only correct type and overvoltage
category rated test leads. Remove the
temperature probe when using the test
leads or amp clamp.
Table des matières
Autres manuels Snap-On Instrument de mesure
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur