SENKO MGT Version II Manuel utilisateur

Table of Contents
2
MGT Docking Station version II
Warning …………………….…………………………………………………………
1. Overview …………………………………………………………………………..
1.1 Description ………………………………………………………...
1.2 Instructions and Parts ……………………………………………..
1.3 Key Features …………………………………………………….…
1.4 Specification …………………………………………………….…
2. Set up and Operation ……………………………………………………………..…
2.1 Power turn on/off …………………………………………………..
2.2 Bay LEDs and meanings ……………………………………......….
2.3 Charging battery …………………………………………………....
2.4 Exhaust Port & Fresh air Inlet ………………………………………
2.5 Installation of Gas Cylinder ……………………………………..…..
3. Testing …………………………………………………………………………….…
3.1 Calibration ………………………………………………………..…
3.2 Bump Test ……………………………………………………….…
4. Trouble Shooting ………………………………………………………………….…
5. Docking Station II Manager Software ……………………………………………..…
5.1 Software Installation ……………………………………………..…
6. Connecting to the MGT Docking Station II ……………………………………….….
6.1 Connecting with SENKO IR Link ……………………………………
6.2 Connecting with USB ……………………………………………….
7. Using Software …………………………………………………………………...….
7.1 Overview of Software ………………………………………………
7.1.1 Docking Station Tab …….………………………..
7.1.2 MGT Tab ……………………………………….…
7.1.3 Logs Tab …………………………………......…..
7.1.4 Certification Tab …………………………......…...
Limited Warranty …………………………………………………………………….…
3
4
4
4
5
5
6
6
6
6
7
7
7
7
8
8
9
9
11
12
12
13
13, 14
15
16
17
18

MGT Docking Station version II
WARNING
Any unauthorized attempt to repair or modify the product, or any
other cause of damage beyond the range of the intended use
including damage by fire, lightening, or other hazard, voids liability
of the manufacturer.
Do not use if the device appears to be damaged.
Use the device in a clean environment, which has no hazardous
gases.
Do not use the expired gas cylinders.
Ensure to use the certified gas cylinders.
Ensure that the gas types and the concentrations match with
manuals and devices.
Severe vibration or shock to the device may cause a sudden
reading change and cause the device malfunction.
Do not leave or use the device at the watery place or at where
there is any liquid.
Do not use a device which has been failed the test.
Read the operation manual before use
✓Before operation, read the manual carefully and follow all the
instructions.
✓IR link communication works only when it is placed at the bay 1on
the docking station.
✓The deice only operates when the power LED turns to green.
✓Before upgrading firmware, connect the plug to avoid any data loss
and malfunction by unexpected power off. And the file names are
fixed as below.
- Docking F/W File Name : DK_FW.bin,
- MGT F/W Name : MGT_FW.bin
3

1. Overview
A Bump test and the calibration are the key features to ensure the safety of
devices and users. MGT docking station version II provides bump testing, event
management and calibration from a simple to use multi-unit station and gives
fleet or safety managers the ability to update configurations on large fleets. It
also maintains all the important event logs that demonstrate users are working
within company requirements.
4
1.2 Instructions and Parts
1.1 Description
a
b
c
d
e
f
g
h
i
j
k
l
m
n
op
q
No Description
a Bay LEDs
b Bay 1
c Handle
d Thumbscrews
e Power LED
f Bump Test button
g Calibration button
h Bay 2, 3, 4
i Pressure gauge
j Cylinder 58L/116L
k Cylinder Straps
l Gas regulator
m
Cylinder compartment
n Exhaust port
o Fresh air Inlet
p LAN port
q Charging port
rUSB port
MGT Docking Station version II
Figure 1
r

5
✓Zero and Span Calibration, Bump Test Function
✓Rechargeable internal battery lasts up to 1000 bump tests on a single charge
✓Indicate the current status of operation by the Power LED and Unit LED color
✓Configure settings and automatic update firmware of up to four MGT devices
✓Gas Ventilation system to remove a remained gas
✓Computer not required for normal operation
✓Save the log events after calibration and bump test
✓Enable users to adjust the setting points of Docking station and MGT via IR
Link or USB
5
1.3 Key Features
1.4 Specifications
Size 52.5 x 43.7 x 21.3cm (20.66” x 17.20” x 8.40”)
Weight
(Without cylinder) 11.0kg
Temperature 41 to +104˚F (5 to +40˚C)
Warranty Full 2 years
Battery Type Rechargeable Lithium Ion
Battery Life Up to 1,000 Bump Tests
LEDs 6 Blue LEDs for each bay, 3-color LED for power,
Switch Back Light LED (Orange, Green)
Memory USB 8GB Standard
Log Capacity Approximately Up to 5 million tests (8GB, removable USB memory)
Tests Performed Bump Test and Calibration
Information
Stored
Bump/Calibration Logs, Individual Monitor Event Logs, Firmware and Unit
Configurations
Gases Available LEL (CH4), CO, H2S and O2
Compatibility Works with SP-MGT(N, P)
Calibration Gas Compatible with 58L & 116L cylinders
Interface Ethernet RJ-45(TCP/IP), 10Mb/s
Charging Adaptor DC12.0V, 3A/h
MGT Docking Station version II

6
MGT Docking Station version II
2. Set up and Operation
2.1 Power turn on/off
The MGT Docking Station version II operates using only 2 (two) buttons for
calibration and Bump Test.
To turn on the power, follow the below:
✓Press and hold either Calibration or Bump Test button for 2 seconds.
✓The Power LED turns on to orange and blinks during the warming up and
after while turns to green when it is ready. When the unit has a low battery,
the LED turns to red.
To turn off the power, follow the below:
✓Press and hold both Calibration or Bump Test buttons simultaneously for 2
seconds.
2.2 Bay LEDs and meanings
Each Bay has six blue LEDs (see Figure 1) which indicates the process status. Each
of them will turn on by order from left to right while testing. After testing is
finished, make sure all six bays are turned on otherwise, 3rd bay LED will be
shown if test is failed.
2.3 Charging battery
The unit`s Power LED turns to red when the battery is very low as shown above.
However, it comes with a DC adapter with 5plugs which can be used to charge
the unit using a 110V power supply. A full charge will take approximately 3 hours
and will last for approximately 1000 bump tests.
Figure 2
Note: Refer to Figure 1 for charging location.
Note: If unit is inactive for 300 seconds, it will automatically off (save power mode)
The button/unit only works when the power LED turns to green

77
MGT Docking Station version II
2.4 Exhaust Port & Fresh Air Inlet
There are two small ports on the left side of the charging port (see Figure 1). The
Fresh air inlet port will draw clean air in during an active test. The incoming air
automatically counts as a fresh air (O2, 20.9%vol).
The Exhaust port is the inverse of the Fresh air inlet port. During an active test,
the used test gas will exit the unit through the exhaust port. Optionally, tubing
can be attached to the exhaust port using the included twist connector to direct
the used test gas to a suitable location.
2.5 Installation of Gas Cylinder
Gas Cylinder which fits to Docking Station II must have the C10 fitting on it. The
size of 58L and 116L are both suitable in cylinder compartment. Before installing
the cylinder for testing MGT, ensure that Calibration concentration of gases are as
same as MGT.
To install a cylinder, follow the below:
✓Remove thumbscrews and lift the handle up.
✓Install the cylinder to the gas regulator.
✓Fasten the cylinder straps and close the cylinder compartment.
✓Check the pressure gauge on the front panel.
✓Re-install thumbscrews.
✓After installing a gas cylinder, ensure USB is attached.
3. Testing
3.1 Calibration
Calibration is the comparison of measurement values delivered by a device under
test with those of a calibration standard of known accuracy. It is very important
to calibrate a device periodically for better performance of the device and sensor
accuracy, especially, when it is safety equipment.
To perform a calibration, follow the below:
✓Insert the monitor(s) into the bay facing down.
✓Slightly press down on the monitor until it locks into place.
✓Press the calibration button for 2 seconds to execute Calibration test.
✓During the 90seconds, bay LEDs begin to light as the test progress.
✓When test is finished, all (six) blue Bay LEDs illumination indicates if the MGT
has passed the test (see Figure 3).
✓Otherwise, the third blue Bay LED illumination indicates if the MGT has failed
the test (see Figure 3).
Bump test and calibration are the key features to ensure the safety of devices and
users. The MGT Docking Station version II provides bump testing, event
management and calibration from a simple to use multi-unit station.

888
MGT Docking Station version II
3.2 Bump Test
Bump test is the process which checks and verifies the performance of the
monitors and ensures that sensors are correctly responding to the target gas.
To perform a Bump Test, follow the below:
✓Insert the monitor(s) into the bay facing down.
✓Slightly press down on the monitor until it locks into place.
✓Press the Bump Test button for 2 seconds to execute Bump test.
✓During the 20~30seconds, bay LEDs begin to light as the test progress.
✓When test is finished, all (six) blue Bay LEDs illumination indicates if the MGT
has passed the test (see Figure 3).
✓Otherwise, the third blue Bay LED illumination indicates if the MGT has failed
the test (see Figure 3).
Figure 3
MGT has passed
the test (six LEDs)
MGT has failed the
test (only 3rd LED)
Bay locking
mechanism
Note: Do not use the monitors that failed the calibration test.
To remove the MGT monitors from the bay, pull to the left side bay
locking mechanism (lower-left side of each bay).
4. Trouble Shooting
4.1 Bay Light does not turn on /test not performed
✓Clean Monitor IR window.
✓IR interference (bright lights/sunlight).
✓Monitor in gas alarm state.
4.2 All tests fail
✓Gas cylinder empty –check pressure gauge if it indicates at least 100psi/10bar.
✓Gas Expired –check gas expiration date located on the cylinder.
✓Wrong gas concentration(s) –gas cylinder and MGT concentrations not same.
✓Wrong gas type –check gas type.

4.3 Test results are not stored
✓USB not attached or full storage.
MGT Docking Station version II
9
4.4 Uploaded Settings are not saved
✓“Write” button has not clicked.
✓Connection has lost.
5. MGT Docking Station II Manager Software
The purpose of MGT Docking Station II Manager Software is to customize and
configure the Station`s settings and MGT devices` internal configurations. The
software can adjust and make changes to the Docking Station and the MGT units
using SENKO IR Link and/or partly USB Disk.
5.1 Software Installation
✓After downloading the file, navigate to the saved location.
✓Double click for running an installer.
✓Click “Read More” and “Run Away” if you receive a warning message.
✓The Docking Manager Setup window will open. Click “Next”.
✓The “Select Installation Folder” comes up.
✓To install to a different folder, click “Browse” or enter the link.
✓To install in default folder, click “Next”

MGT Docking Station version II
10
✓The Installation will be ready to begin.
✓Click “Install”.