DUT Connection
If you elect to use test leads, follow these instructions to safely connect a DUT.
Hook the black return lead (P/N 99-10238-01) into the RETURN terminal located
on the front panel. Connect the other end of the return lead to the dead metal
on the chassis of the DUT.
Hook the high current lead (P/N 99-10239-01) into the front panel CURRENT
output terminal. You may need to unscrew the terminal slightly to make the
connection. Connect the other end to the ground pin of the DUT line cord.
Plug the high voltage lead (102-055-913) into the H.V. terminal located on
the front panel. Connect the clip end of the high voltage lead to the current
carrying conductors of the DUT’s circuitry.
Exposed Metal Chassis Setup
If your chassis does not have any exposed metal, you can wrap the enclosure of the
DUT in foil and then connect the return lead to the foil.
Non Exposed Metal Chassis Setup
1
2
3
1
2
3
3
1
2
Conducting a Test
WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.