pyromation 28-02700-J Manuel utilisateur

Thermocouple Thermometer Models
28-02700-J (Type J)
28-02700-K (Type K)
28-02700-T (Type T)
5211Industrial Road
FortWayne,IndianaU.S.A.46825
(260)484-2580
(260)482-6805(Fax)
A-1299-0680
Edition05
OPERATING MANUAL

CERTIFICATE OF
CONFORMANCE
This thermometer was calibrated using
equipment traceable to the National Institute
ofStandardsandTechnology(NIST).
Thisinstrument conforms to
NISTmonograph175 revisedto
ITS-90.
The accuracy of the thermometer at the time
ofcalibrationwas within specificationsstated
intheoperatingmanual.
ModelNo.:_________________________
SerialNumber:______________________
Dateplacedinservice:________________
Topurchase an NIST certificate of traceability
withtestdata andtestdateformeterandprobe,
pleasecontact yourdealer or:
PYROMATION
5211Industrial Road
FortWayne,IndianaU.S.A.46825
260-484-2580

1
INTRODUCTION
Thisversatilehand-heldinstrument provideshighly
accuratetemperature measurements. The instru-
ment is designed for easy operation and includes
thefollowingfeatures:
• Operator selectionof Celsiusor Fahrenheit scale
• Resolution of 0.1° from–99.9 to 299.9°
• Four-digit LCD
• Hold feature for temporarily retaining areading
• Field calibrationcapability
• Low battery warning
• Built-in tilt stand for easy hands-freeoperation
• Two blade female ANSI mini-connectorinput
• Operates with a wide selection of probes

2
SAFETY PRECAUTIONS
THIS INSTRUMENT IS
DESIGNED TO ACCEPT
LOW LEVEL SIGNALS SUPPLIED BY STAN-
DARD THERMOCOUPLES. UNDER NO CIR-
CUMSTANCES SHOULDTHE INPUTVOLTAGE
EXCEEDTHE SPECIFIED 50V RMS.
DO NOT USE OR STORE
THIS INSTRUMENT IN
MICROWAVE OVENS OR ANY ABNORMALLY
HOT OR COLD AREAS.
WEAKBATTERIES SHOULD
NOT BE LEFT IN THE IN-
STRUMENT. DEADBATTERIES CAN LEAKAND
CAUSE DAMAGETO UNIT.
VOLTAGES PRESENT AT
THETHERMOCOUPLEMAY
ALSO BE PRESENT AT THE BATTERY TERMI-
NALS. ALWAYS DISCONNECT THE THERMO-
COUPLEWHEN CHANGING BATTERIES.
TO PREVENT IGNITION OF
A HAZARDOUS ATMOS-
PHERE,BATTERIESMUSTONLYBECHANGED
INAN AREA KNOWNTOBENON-HAZARDOUS.
AFIN DE PREVENIR
L'INFLAMMATION
D'ATMOSPHERES DANGEREUSES, NE
CHANGER LES BATTERIES QUE DANS DES
EMPLACEMENTS DESIGNES NON DAN-
GEREUX.
TOPREVENT IGNITION OF
A HAZARDOUS ATMOS-
PHERE BY ELECTROSTATIC DISCHARGE,
CLEANWITH DAMP CLOTH.
WARNING
WARNING
CAUTION
CAUTION
WARNING
DANGER
AVERTISSEMENT

3
SPECIFICATIONS
THERMOCOUPLE PROBES
Type Temperature Range
Type J: –200°C to 1000°C (–328°F to 1832°F )
Type K: –250°C to 1372°C(–418°F to 2501°F)
TypeT: –250°Cto 400°C(–418°F to 752°F)
Out of range display: - - - -
Resolution: 0.1°/1° autoranging, 0.1° from
–99.9°to299.9°, 1° outside this range.
Accuracy
> –99.9°: +(0.2%ofreading +0.5°C)
+(0.2% of reading+0.9°F)
< –99.9°: +(0.25%ofreading, +1°C)
+(0.25%ofreading +2°F)
Display:4-digitLCD with 0.5 in high numerals.
Display update rate: 0.6 sec per update.
Input: Onethermocouple withANSIconnector.
Input Protection: 50V rms
Battery: Two AA, 1.5V alkaline ANSI-L40,
IEC-LR6.
Battery Life: 750hours continuous,typical.
Low battery indication
Battery Symbol on when 8 to 10 hours of
batteryliferemains.
Batterysymbolblinking:replace battery
Operating Conditions
Stated accuracy: 18°C to 28°C (64°F to 82°F)
Useful range: 0°C to 40°C (32°F to 104°F)
Storage: –40°C to 65°C (–40°F to 149°F)
Humidity(non-condensing):10%to90%

4
Dimensions
3 cm D x 8.4 cm W x 15.8 cm H
(1.2 in x 3.3 in x 6.2 in)
Weight with batteries: 227 grams (8 ounces)
Ingress protection: MeetsIEC-529 IP-54 fordust
andwater-resistantenclosures.
Intrinsic safety
Thisproductisenergy limitedforintrinsicallysafe
operationin hydrogenatmospheres per Class I,
Division 1, Groups A, B, C and D hazardous
(classified)locations forUL per UL913andCSA
per C22.2 No. 0-M1982 and No. 157-M1987.
Maximumsurfacetemperature:135°C(T4);UL
fileNo.E182612 (1997).
Compliance: (For CE Mark)
EN61326-1/A1:1998(EU EMC Directive)

5
BATTERY INSTALLATION AND
REPLACEMENT
WEAKBATTERIES SHOULD
NOT BE LEFT IN THE
INSTRUMENT. DEAD BATTERIES CAN LEAK
AND CAUSE DAMAGETO UNIT.
VOLTAGES PRESENT AT
THETHERMOCOUPLEMAY
ALSO BE PRESENT AT THE BATTERY TERMI-
NALS. ALWAYS DISCONNECT THE THERMO-
COUPLEWHEN CHANGING BATTERIES.
TO PREVENT IGNITION OF
A HAZARDOUS ATMOS-
PHERE,BATTERIESMUSTONLYBECHANGED
INAN AREA KNOWNTOBENON-HAZARDOUS.
AFIN DE PREVENIR
L'INFLAMMATION
D'ATMOSPHERES DANGEREUSES, NE
CHANGER LES BATTERIES QUE DANS DES
EMPLACEMENTS DESIGNES NON DAN-
GEREUX.
CAUTION
DANGER
WARNING
AVERTISSEMENT

6
If battery indicator turns on, battery life is approxi-
mately8 to 10 hours.Thebatteryindicator will start
blinkingwith less than 1 hour of liferemaining.
ATTHISPOINT,BATTERYMUST BE CHANGED.
IF BATTERY VOLTAGE GOES TOO LOW, DIS-
PLAY WILL GO BLANK.
SeeSPECIFICATIONS forbattery type.
1. Beforechanging battery, turninstrument offand
disconnectthermocouple.
2. Loosen screw and lift battery cover off back of
case.
3. Removethe two AA batteries.
4. Insert twonewbatteries observing polarity.
5. Install cover and tighten screw.

7
CONNECTING A THERMOCOUPLE
Usethecorrect thermocoupletypeforyourinstru-
ment. Using an incorrect thermocouple type will
result in erroneous readings. Thermocouples are
colorcodedbytypeusing theNorthAmerican ANSI
ColorCode as follows:
TYPE COLOR
J Black
K Yellow
T Blue
Thermocoupleconnectorshaveonewidebladeand
onenarrowblade.
Donot forceconnector in back-
wards.
Connectthermocouple to receptacle at top
ofinstrument as showninthefollowingillustration.
Thermocouple wiring polarity must be correct. If
readings decrease as the temperature increases,
thethermocouple wires maybereversed. The red
wire is negative for thermocouple wires manufac-
turedin NorthAmerica.

8