3
LabCon Map + Printer User Manual
The Alarm / Process Status space on the screen may show one or more of the several messages listed in Table 2.2 depending
upon the existence of active alarms or process events. In case of co-existence of multiple active alarms or process events, the
messages are scrolled one after the other with an approximate time interval of 3 seconds. Also if there is no active alarm or
process event, the message space is blank.
Mapping Status Screen
This is the default Run Mode screen that shows easured Temperature Value, et Temperature Value, Calendar Date, Clock M S
Time and Alarm / Process Status as illustrated in Figure 2.2 (a) above.
If Soak Timer function is enabled, the un ode screen also shows the elapsed time as R M illustrated in Figure 2.2 (b) above.
In case of measured value errors, the messages listed in Table 2.1 flash in place of process value as illustrated in Figure 2.3.
Message Error Type Cause
Sensor Open
Under-range
Over-range
Sensor (RTD Pt100) Broken / Open
Temperature above Max. Specified Range
Temperature below Min. Specified Range
OPN
Table 2.1
OVR
UNR
Figure 2.3
°C
6/01/16 10:40:06
OPN
25.0
Temp High SP
Figure 2.4
The Mapping Status Screen is shown upon pressing (Scroll) key from
Main Screen.
The Controller provides up to 4 mapping channels. The Mapping Status
Screen is available only if the controller is configured for at least one
mapping channel. The mapping status screen for 4 mapping channels is
shown in the figure 2.4. In case lesser mapping channels are configured,
the skipped channels are not shown on the screen.
Message String Alarm / Process Status
The Controller is Self Tuning.
The measured temperature at Mapping Channel
‘x’ (x = 1, 2, 3 or 4) is in Process High Alarm State.
Temp High
Temp Low
Map-x High
Map-x Low
Tuning
Mains Fail
The measured temperature at Control Channel is
in Process High Alarm State.
The measured temperature at Control Channel is
in Process Low Alarm State.
The measured temperature at Mapping Channel
‘x’ (x = 1, 2, 3 or 4) is in Process Low Alarm State.
The Controller is operating on Standby Power
Source like battery or inverter.
Table 2.2
Door Open The Equipment / Chamber door is open for a time
longer than the set ‘DOOR ALRM DLY’ time.
MAPPING STATUS
MAP PV °C ALARM
125.1 Normal
224.4 Low
325.0 Normal
425.4 High