OfiTE 170-182-1 Manuel utilisateur

HTHP Filter Press for Cement Testing
with 175-mL, Double-Capped Test Cell and N2Pressuring Manifold
#170-182: (115 V)
#170-182-1: (230 V)
Instruction Manual
Updated 6/20/2017
Ver. 3.0
OFI Testing Equipment, Inc.
11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
Tele: 832.320.7300 · Fax: 713.880.9886 · www.ote.com
©Copyright OFITE 2015

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 1
Intro..................................................................................................2
Specications .................................................................................3
Components ...................................................................................4
Safety...............................................................................................8
Quick Start ....................................................................................12
Operation.......................................................................................15
Data................................................................................................20
Cell Cap Assembly .......................................................................21
Maintenance..................................................................................22
Diagrams .......................................................................................24
Test Cell ....................................................................................25
Warranty and Return Policy ........................................................26
Table of
Contents

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 2
The OFI Testing Equipment (OFITE) High Temperature High Pressure
(HTHP) Filter Press is designed to evaluate the ltration characteristics of
drilling uids, cement slurries, fracturing uids, and completion uids under
elevated temperatures and pressures. Evaluating uids under HTHP condi-
tions similar to the downhole environment is of paramount importance. Fluid
properties must be monitored while under high temperatures and pressures
as ltration behavior and wall cake building characteristics of permeable
formations change with changing environments. These characteristics are
affected by the shape, type, and quantities of solids present in the uid and
their physical and electro-chemical interactions, all of which are affected by
changing temperatures and pressures.
OFITE manufactures and provides HTHP ltration units in two basic sizes,
175 mL and 500 mL capacities. Both are used extensively throughout the
world and in all environments, but in general the 175 mL units are designed
for eld portability, while the larger 500 mL units are designed for laboratory
usage at higher temperatures and pressures. All OFITE Filtration devices fully
conform to American Petroleum Institute (API) specications.
A complete HTHP Filter Press consists of a controlled pressure source, usu-
ally Nitrogen pressurization or Carbon Dioxide bulbs for the 175 mL units.
Top and bottom pressure manifolds are provided to simulate the differential
pressures found in a down-hole environment, and to prevent evaporation
of the base uid if exceeding the boiling point of that uid. The test cells are
provided in a variety of assemblies, depending upon the type of uid tested,
the lter media, and the temperatures and pressures desired. The test cells
are encased inside a heating jacket, which is adjustable.
A variety of lter media are available, the most common being standard API
lter paper, cement screens, and ceramic lters. The ceramic lters may be
obtained to match the pore throat or permeability of the formation. Natural for-
mation lters or cores may also be used of differing pore throat / permeability
sizes. Slotted disks of varying sizes are frequently used for lost circulation
materials studies.
Both the 175 mL and the 500 mL heating jackets are capable of reach-
ing 400°F (204°C), but lower uid volumes due to uid expansion at higher
temperatures, limit the 175 mL units to a useful working temperature of 300°F
(149°C). Anyone running tests above 350°F (177°C) must substitute a com-
plete set of o-rings after each and every test.
Intro

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 3
Size: 7.5" × 11" × 23.5" (19.1 × 27.9 × 59.7 cm)
Weight: 27 lbs. (12.3 kg)
Shipping Size: 17" × 23.5" × 12.5" (43.2 × 59.7 × 31.8 cm)
Shipping Weight: 38 lbs. (17.2 kg)
Maximum Temperature
(Heating Jacket): 400°F (204°C)
Maximum Temperature (Cell): 500°F (260°C)
Maximum Pressure (Cell): 5,000 PSI
Maximum Pressure (Receiver): 750 PSI (5.1 MPa)
Pressure Source: Nitrogen (at least 1,500 PSI / 10,343 kPa) -
Not Included
Test Cell Capacity: 175 mL
Receiver Volume: 15 mL
Heater: 400 Watt
Power Requirement: 115 VAC, 5 Amps, 50/60 Hz
230 VAC, 3 Amps, 50/60 Hz
Specications

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 4
Components #153-14 Graduated Cylinder, 50 mL × 1 mL
#154-10 Dual-Scale Thermometer with Dial, 5" Stem, 50° - 500°F (0° -
260°C)
#165-44-2 Anti Seize Compound, Silver, 7g Pouch
#170-00-1 Heating Jacket (115V) -OR-
#170-01-1 Heating Jacket (230V):
#130-10-52 Jam Nut; ⅜"-24; Stainless Steel; Qty: 2
#164-32 Male Connector for Power Cable (For 170-01-1 230V ONLY)
#170-05 Thermostat
#170-09 Insulation Board
#170-10 Thermostat Pilot Lot
#170-11 Heating Element; 115V; 200W; Qty: 2
#170-15 Base
#170-21 Stand Support Rod; Qty: 2
#170-25 Aluminum Well
#170-30 Stainless Steel Thermostat Cover
#170-44 ½" Rubber Foot; Qty: 4
#171-32 Midget Knob
#171-82 8' Power Cord with Male Plug 8; 16/3 SJ; Round (For 170-
00-1 115V ONLY)
#170-06-1 Back Pressure Receiver; 15-mL Stainless Steel Tube for N2
#144-11 ⅛" 90 Street Ell
#144-15 Plate Brass Bushing; ¼" NPT Male to ⅛" NPT Female
#170-07 O-ring
#170-28 Receiver Body
#170-32 ⅛" × ⅛" NPT Male Needle Valve
#171-23-1 Safety Pin with Lanyard
#170-19 Filter Paper; 2½" (6.35 cm); Specially Hardened for Filter Presses
#170-35 6" Adjustable Wrench
#170-182-S Test Cell, Stainless Steel (For Cement Testing)
#120-910-028 O-ring for Rupture Disk, Viton 75D, Qty: 1
#130-81-040 Retaining Ring, Qty: 2
#165-44-2 Anti Seize Compound, Silver, 7g Pouch, Qty: 4
#170-13-3 O-ring for Cell, Viton 75D, Qty: 4
#170-16 Valve Stem, Qty: 2
#170-17 O-ring for Valve Stem, Viton 75D, Qty: 4
#170-18 Cement Screen, Qty: 2
#170-180-020-S Cell Body, 175 mL, Qty: 1
#171-190-023 Locking Ring, Qty: 2
#171-190-027 Rupture Disk, Qty: 1
#171-190-029 Cap Wrench, Qty: 1
#171-190-032-S Cell Cap, Outlet, Cement, Qty: 1
#171-190-033-S Cell Cap, Inlet, Cement, Qty: 1
#171-190-057 O-ring for Valve Stem, Viton 90D, Qty: 4
#171-190-058 O-ring for Rupture Disk, Viton 90D, Qty: 1
#171-190-060 O-ring for Cell, Viton 90D, Qty: 4

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 5
#171-24 1350 / 750 PSI (9,308 / 5,171 kPa) Nitrogen Manifold
#170-08 Regulator; Qty: 2
#170-20 Manifold Block
#170-32 ⅛" × ⅛" NPT Male Needle Valve
#171-22 Retainer Pin
#171-24-1 Chrome Nut; R.H.; Reg Inlet CGA-580
#171-24-2 Nipple with Filter
#171-24-3 ¼" Flare × ⅛" FNPT Female
#171-24-4 ¼" NPT Pipe Plug; 316 Stainless Steel
#171-24-5 ¼" NPT Street Tee; 316 Stainless Steel; Qty: 2
#171-25-1 Relief Valve set at 750 PSI (5171 kPa)
#171-25-2 Relief Valve set at 1350 PSI (9308 kPa)
#171-26 5000# Hose; " × 3'; Qty: 2
#171-28 Dual Manifold Body
#171-38 1,000 PSI Gauge; 2½"; ¼" NPT Bottom
#171-40 1,500 PSI Gauge; 2½"; ¼" NPT Bottom
#171-42 3,000 PSI Gauge; 2½"; ¼" NPT Bottom
#171-90-06 ¼" MNPT × ⅛" FNPT Reducing Bushing; 316 Stainless
Optional:
#170-13 O-ring for Test Cell, Buna, For tests below 300°F
#170-37 Nitrogen Cylinder; 21" × 7"; Right-hand Thread
Part Number Description Qty.
#143-00-1 Diaphragm for Airco Regulator 1
#143-07 Repair Kit for Regulator (#143-00) 1
#153-14 Graduated Cylinder, 50 mL × 1 mL 2
#154-10 Dial Thermometer, 5" Stem, 50° - 500°F and 0 - 250°C 1
#170-07 O-ring for Back Pressure Receiver, Nitrile 3
#170-13-3 O-ring for Cell, Viton 75D 50
#170-16 Valve Stem 4
#170-17 Valve Stem O-ring 100
#170-18 Cement Screen 10
#171-190-057 O-ring for Valve Stem, Viton 90D 100
#171-190-060 O-ring for Test Cell, Viton®50
#171-23-1 Safety Pin with Lanyard 1
#170-182-SP Spare Parts Kit:

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 6
Optional Items for HTHP Filtration Testing:
The items listed below are not included in the HTHP Filter Press, but they are
items that will enable the technician to perform a more uniform and reproduc-
ible test while maintaining a high degree of safety. As optional items, the us-
age is not compulsory, but consideration should be given to these items when
running tests at elevated temperatures and pressures.
Interval Timer, 60 minute
(#155-20)
Cell Cap Removal Tool
(#170-33)
(Set Screw Cell Assemblies Only)
HTHP Pressure Relief Tool
(#170-91)
(To release trapped pressure)
Safety Shield (#171-06)
Cell Carrying Tool (#170-40)

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 7
Thermocouple Assembly
(#171-45-1)
(Direct temperature measurement
Of the uid Inside the Cell)
Stand for HTHP Cell Assembly
(#171-190-028)

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 8
Safety Pressure and Temperature Considerations
Do not use or recongure this equipment in a manner not specied in this
manual.
Pressure
There are two reasons for operating at elevated pressures when performing a
ltration analysis.
1. To test uids at temperatures above the boiling point, the vessel must
be pressurized, which in turn elevates the vapor pressure (boiling
point) so that the uid remains liquid and does not turn to steam.
2. If testing for drilling applications, pressurization will more approximate
down-hole conditions, enabling the technician to match both bottom
hole temperatures and pressures.
HTHP Filter Presses are pressurized either with Carbon Dioxide or with
Nitrogen gas. The OFITE 175 mL HTHP Filter Press, 170-00 Series, is small
enough to be portable, so it is usually pressurized with small (65 mm length)
Carbon Dioxide (CO2) bulbs, which contain 10 cm3 of CO2 gas and weigh 8
grams. These bulbs are pressurized to approximately 1,000 PSI and they
contain plenty of carrier gas to run a complete 30 minute ltration test, if run-
ning the standard API 500 PSI differential test, which is usually 600 PSI on
the top manifold and 100 PSI on the bottom or back pressure.
For temperatures less than 200°F (93°C) a Back Pressure Receiver is not
required as the ltrate will not reach the boiling point. However when operat-
ing above the boiling point of water, a suitable Back Pressure Receiver is
required, otherwise the test uid will turn to steam and the test is invalidated.
The standard receiver tube supplied with the 175 mL HTHP Filter Press holds
15 mL of ltrate, so it is very important that the receiver outlet ball valve is
opened after approximately ten seconds of ltrate collection, as a uid with
poor ltration qualities can easily ll the 15 mL receiver tube almost instan-
taneously. If this happens the ltrate hopefully will all be ejected from the
safety bleeder valve, but if there is too much uid volume, the liquid may end
up inside the regulator rendering it useless which will require servicing by a
knowledgeable technician.
Temperature
Normally when one experiences a loss of pressure it is not due to a failure
of the metal alloy in the cell, but rather is a failure of the o-ring or elastomer,
which provides the seal. These o-ring may deform or melt under elevated
temperatures causing a pressurization failure of the cell, which is often sud-
den and catastrophic. For example, if the valve stem o-ring suddenly fails,
then steam at 400°F and under great pressure may shoot outward horizon-
tally in one or several directions. A safety shield should always be used when
operating any HTHP Filter Press and especially when one is going to extreme
temperatures and pressures. Above 350°F (176°C) All o-rings must be re-
placed after each and every test.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 9
Heating Jacket
Turn the thermostat knob on the front of the metal plate to change the tem-
perature of the heating jacket. The thermostat cover is marked from 1 to 10
and each whole number represents a separate temperature. Each lter press
will heat up a little differently so it is a good idea to mark with a pen directly
on the face-plate of the thermostat cover the temperature which corresponds
with each number. The actual temperature of the heating jacket is measured
with a stem thermometer and then the thermometer is placed into the cell
body.
The heating jacket will get very hot during the test so be careful not to touch
the outside of the jacket at any time during the heat-up period or the test. It is
especially easy to get burned when installing or removing the Back Pressure
Receiver as the technician is working in a very conned space. The power
cord is supplied for either 115 Volts or 230 Volts and due to the various types
of plug connectors around the world it may be necessary to change the plug
on the end of the power cord. The power cable is 8 feet in length and the
heating jacket should be placed no farther than 8 feet from the appropriate
electrical outlet. The heat up time will vary from one instrument to the next
and the pilot light will turn on when the set temperature has been reached.
The temperature of the uid inside the cell however will not be at the set tem-
perature, so always allow one hour of heating time for the uid, after the cell
has been fully inserted into the heating jacket.