NKE PIEZOMETER Manuel utilisateur

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 1/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 1/41

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 2/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 2/41
Sommaire
SOMMAIRE
1.
INTRODUCTION 4
2.
OVERVIEW AND DESIGNATION 6
2.1
« D
ATING
&
S
TORAGE
M
ODULE
»
(MDS) 6
2.2
B
ASE PLATE
7
3.
LANCE ASSEMBLY PROCEDURE 8
3.1
L
ANCE ELEMENT
8
3.2
S
PECIAL TOOLS
9
4.
ASSEMBLY PROCEDURE 10
4.1
B
OTTOM END ASSEMBLY WITH BUS CONNECTION
10
4.2
S
TEP
-1 :
B
OTTOM SLEEVE ASSEMBLY WITH STANDARD TUBE
11
4.3
S
TEP
-2 :
W
INPIEZO CHECK
11
4.4
S
TEP
-3 :
N
EXT SLEEVE ASSEMBLY WITH BUS CONNECTION
11
4.5
MP&T
BUS
–
MDS
CABLE ASSEMBLY WITH BULKHEAD AND FIRST SLEEVE
12
4.6
C
HECK THE OPERATION
12
4.7
O
IL FILLING OF THE TUBE
12
4.8
F
INALIZE WITH A LANCE TUBE
13
5.
SOFTWARE WINPIEZO 14
5.1
P
RESENTATION AND
U
SE
15
5.2
C
OMMUNICATION MODE
15
5.2.1
Vitesse de communication du bus et configuration des Switchs de la carte MDS 15
5.3
I
NSTALLATION
16
5.4
I
NSTALL THE DRIVERS
16
5.4.1
Procedure : 16
5.4.2
Determine the number of the serial communication port . 16
5.5
D
ESIGNATION OF SOFTWARE COM
17
5.6
O
VERVIEW
18
5.7
"MDS
P
ROGRAMMING
"
FUNCTION
18
5.8
S
EQUENCE OF OPERATING MODES FOR THE
MDS
MODULE
19
5.8.1
BOOT mode 19
5.8.2
SYNCHRONIZATION mode 20
5.8.3
BUS mode 20
5.8.4
Le mode double cadences 21
5.8.5
ACQUISITION Mode 22
5.8.6
WAITING mode 22
5.8.7
RECORDING Mode 22
5.8.8
RECORDED Mode 22
5.9
D
ISPLAY THE DATA FILES
23
5.10
CALIBRATION
FUNCTION 24
5.11
D
ESCRIPTION OF COMMANDS
24
5.12
P
ROCEDURE
24
5.13
E
XPORTING THE GRAPH MEASUREMENTS
. 25
6.
COMMUNICATION INTERFACE 26
6.1
D
ESIGNATION OF THE EQUIPMENT
26

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 3/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 3/41
6.2
D
ESCRIPTION OF THE INTERFACE
26
6.3
U
SE WITH
"MDS
MODULE
" 26
6.4
U
SE WITH
"MP&T" 26
1. SHIP CARACTERISTICS 28
1.1.
DECK
(A
FRAME DEPLOYMENT
) 28
2. LANCE INSTALLATION (FROM 6 TO 15M) 29
2.1.
A
FRAME LAUNCHING
–
F
ILLING IN STOP
29
2.2.
P
AY OUT
30
2.3.
C
ALIBRATION STOP
31
2.4.
L
ANCE PENETRATION
32
2.5.
V
ERTICALITY CONTROL
33
2.6.
D
EAD WEIGHT RECOVERY
34
3. LANCE RETRIEVAL (FROM 6 TO 15M) 35
3.1.
R
ETRIEVAL OF PIEZOMETER
35
3.2.
H
OOK UP
36
3.3.
R
ETRIEVAL
37
3.4.
P
IEZOMETER RECOVERY
38
7.
1 MODE YOYO 39
7.1
G
REEMENT
: 39
7.2
P
ROCEDURE DE MISE A L
'
EAU
: 39
8.
SUPPORTING DOCUMENTS 40
8.1
A
SSEMBLY SHEET
41

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 4/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 4/41
1. INTRODUCTION
The piezometer is an instrument that consists of an expendable instrumented lance driven into
the sediment by gravity, based on the principle of a Kullenberg corer, and a retrievable module
that includes data storage and power supply. The sensors in the lance (torque, differential
pressure and temperature) are spaced at predetermined intervals in the lance.
The differential pressure sensors measure the difference between the hydrostatic pressure at the
water/sediment interface and the pressure in the sediment at the sensor via a porous stone filter.
The temperature sensor located as close to the porous stone as possible measures the
temperature of the sediment.
The measurements provided by the pressure/temperature sensors are sampled at regular
intervals.
The data is stored in digital format in a container that also houses the power supply.
After triggering an acoustic release, this container rises to the surface using buoyancy. Surface
tracking is provided by a system that comprises a flag, a beacon and a VHF radio.
The piezometer can be used in "yo-yo" mode, whereby it remains connected to the coring cable,
with a measurement time of a few hours, while the ship remains directly above the station. Once
the measurement is made, the piezometer and its tube are retrieved by heaving the coring cable.
In "yo-yo" mode, the acoustic release system, surface tracking system and trip gear are not
used.
In both modes, a query sent to an acoustic release may provide the tilting angle of the system
after it has sunk into the sediment.
The optional "MDS module" can manage up to two measuring systems at the same pressure
and provides data storage. It also includes an interface for configuration and calibration of the
instrument, as well as an interface to connect the instrument to a potential ocean bottom
observatory, which will provide real time data transmission to the user.
The instrumented lance has "connections" spaced at predetermined intervals, which
accommodate two types of sensors:
- A temperature sensor (CTN) with a measurement range of 0°C to +50°C
- A differential pressure sensor +/- 2 bar
Inclinometers integrated to the module indicate if the system is poorly sunk (e.g. buckling)..

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 5/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 5/41
Vue d'ensemble du piézomètre complet

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 6/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 6/41
2. OVERVIEW AND DESIGNATION
2.1 « Dating & Storage Module » (MDS)
The « Dating & Storage Module »integrates a batteries Pack 3S7P LS26500 SG; 10,8V 54Ah
and a « SD Card » which contains different files according:
- File format measure « xml »
- Configuration File « cfg »
- The logger « log »
The position of the container is given by the two inclinometer mounted on the board electronic
board.

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 7/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 7/41
2.2 Base plate
The « Base plate » is equipped with « Cônes »' at the corners and two « mobile tab » that are
used to receive the« Ballast».
In its upper part is integrated « skirt ».mounted on three « silent blocks » where the « MDS »
enclosure via the two postures clamps is fixed.
The «
Bulkhead
» receiving the "arrow" is assembled from beneath using the four bolts M10x30

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 8/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 8/41
3. LANCE ASSEMBLY PROCEDURE
3.1 Lance element
The lance comprises the elements described below:
Bulkhead : Allows the lance to be mechanically
connected to the ballast part of the piezometer. It will
receive the first tube of the lance.
First connecting tube:
Provides the connection between
the bulkhead and the first sensor sleeve.
The tube's length is 75 cm
MP&T Bus – MDS cable
: This cable allows the bus from
the firs
t sensor sleeve to be connected to the MDS
container.
Connection to the container is made via the 8-pin
Subconn connector. Connection to the bus is made via
the 6-pin “Micro-Fit” connector.
Sensor sleeve:
Allows two tubes from the lance to be
connect
ed. It contains the MPT card and the differential
pressure and temperature sensor.
The first sleeve is unique in that is has a 4-
pin connector
and a 6-
pin connector. All others are reversible (identical
4-pin connectors)
Contains 60 ml of oil after oil filling of the system.

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 9/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 9/41
Bus connection
: reversible cable used to connect the
transmission bus. The sizing of these cables depends on
the length of the tubes used to form the lance. A cannula
with a 4-pin Micro-Fit connector is fitted at each end.
Contains 20 ml of oil per metre.
Connecting tube: They are reversible. Currently
available in three lengths (0.75, 1.5 and 3 m)
Equipressure bladder:
Allows a potential lack of oil to
be compensated.
100 ml capacity.
Lance tube : Final element of the piezometer lance.
3.2 Special tools
Vacuum pump: Allows oil filling of the
communication bus.
Beaker: Used to fill the equipressure bladder.
Preferably use a model with a hole at the bottom, a
connecting pipe and a valve..

USER GUIDE PIEZOMETRE
33-30-077-
Piezomètre_UTI UK
Rev : 3 Page 10/41
nke Tel +33 (0)2-97-36-10-12 Fax +33 (0)2-97-36-55-17 http://www.nke-nstrumentation.com PAGE 10/41
4. ASSEMBLY PROCEDURE
To complete the assembly of a piezometer lance, we start from the bottom end of the lance and
finish with the bulkhead, which weighs around 20 kg.
To facilitate the work, arrange the sleeve numbers in the ascending order, starting from the top
end of the lance. The first sleeve with a 6-pin connector will not necessarily bear the lowest
number.
Description of the important steps of the assembly:
4.1 Bottom end assembly with bus connection
•Install the 6 joints on the sleeve using a minimum of grease
•Install the joint on the cannulas of the bus connection using a
minimum of grease
•Connect the card side of the sleeve to the bus connection
•Stow the connector and the wires inside the cannula
•Stow the wires inside the sleeve by rotating it
•Grease and fit the cannula onto the sleeve (two M5 screws)
Table des matières
Autres manuels NKE Instrument de mesure
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur