Luminar 5030 Manuel utilisateur

LUMINAR
TM
AOTF NIR SPECTROMETER
Luminar Model 5030
Installation Manual
Hand Held
Miniature NIR Spectrometer
Version .0, Revision 7
Copyright 2018 Brimrose Corporation of America

5030 Miniature NIR Spectrometer User Gen Install Manual Page 2 of 29
DO : SUMan-5030GenInstallManual-v3-rev7-20180314
opyright © 2018 Brimrose orporation of America. All rights reserved. No part of this manual may be
reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording,
storing in an information retrieval system, or translating, in whole or in part, without the prior written
consent of Brimrose orporation of America.
Trademarks
Unscrambler is a trademark of amo AS.
Windows™ is a trademark of Microsoft orporation.
Pentium™ is a trademark of Intel orporation.
Luminar™ is a trademark of Brimrose orporation of America.
Snap!™ is a trademark of Brimrose orporation of America.
Snap32!™ is a trademark of Brimrose orporation of America.
Document Number: SUMan-5030GenInstallManual-v3-rev7-20180314
Release Date: March 14, 2018
Supersedes: SUMan-5030GenInstallManual-v3-rev6-20150210
Version Number
3.0
Revisions
7
Brimrose orporation of America
19 Loveton ircle, Sparks, MD 21152-9201 USA
Phone: (410) 472-7070
Fax: (410) 472-7960
Email: [email protected]
Website: www.brimrose.com

5030 Miniature NIR Spectrometer User Gen Install Manual Page 3 of 29
DO : SUMan-5030GenInstallManual-v3-rev7-20180314
Table of Contents Page
1.0 Electrical Installation ....................................................................................................................... 5
1.1 Power Requirements ....................................................................................................................... 5
1.2 Connecting the Spectrometer .......................................................................................................... 5
1.2.1 A Mains to the Power Supply ......................................................................................... 5
1.2.2 UPS onnections (Optional) .............................................................................................. 5
1.2.3 External Power Supply connection to the Spectrometer .............................................. 5
1.2.4 Ethernet onnection ............................................................................................................ 5
1.3 Battery Power to the Spectrometer (Optional ........................................................................... 6
1.3.1 Battery installation ............................................................................................................ 6
1.3.2 Battery Removal ................................................................................................................... 6
1.3.3 Battery charging ................................................................................................................... 6
2.0 Communications ............................................................................................................................... 7
2.1 Ethernet Cabling ............................................................................................................................. 7
2.2 Wireless LAN (Optional ................................................................................................................ 8
2.2.1 hanging the communication protocol ............................................................................. 8
2.2.1.1 Accessing your spectrometer IP Address ......................................................................... 8
2.2.1.2 Using the PUTGET program to change IP Address ....................................................... 10
3.0 Software Installation ...................................................................................................................... 12
3.1 Snap32! Installation...................................................................................................................... 12
3.2 Ethernet Adapter Installation ....................................................................................................... 12
3.3 TCP/IP Protocol Installation ........................................................................................................ 12
4.0 Pre-aligned amp ........................................................................................................................... 13
4.1 Pre-aligned Lamp Replacement ................................................................................................... 13
5.0 Gain Adjustments ........................................................................................................................... 13
5.1 Hardware Gain Adjust ........................................................................................................... 13
6.0 Probes (Optional) ............................................................................................................................ 13
6.1 Installing the Probe ...................................................................................................................... 13
Appendix 1- Basic Cabling Diagram .................................................................................................. 15

5030 Miniature NIR Spectrometer User Gen Install Manual Page 4 of 29
DO : SUMan-5030GenInstallManual-v3-rev7-20180314
Appendix 2 - WLAN Installation Diagram ......................................................................................... 16
7.0 Appendix 3 - Pre-Aligned amp Replacement ............................................................................ 17
8.0 Appendix 4 - Exchanging the Cap with the Probe ....................................................................... 22
8.1 Unit with cap in place ............................................................................................................ 2
8.2 Unit with ap Removed ........................................................................................................ 24
8.3 Unit with probe prior to installation ..................................................................................... 25
8.4 Unit with optional probe installed ........................................................................................ 26
9.0 Appendix 5 – Block Diagrams and Information for Miniature Spectrometers
(4030/5030/7030) ................................................................................................................................... 27

5030 Miniature NIR Spectrometer User Gen Install Manual Page 5 of 29
DO : SUMan-5030GenInstallManual-v3-rev7-20180314
1.0 Electrical nstallation
1.1 Power Requirements
The Hand Held Miniature Spectrometers use D power accomplished by a dedicated external power
supply. The A line provides power to the external power supply. Battery power is available as an
option for portable or mobile applications. (see section 1.3 below.)
1.2 Connecting the Spectrometer
CAUT ON: Before doing anything, be sure the spectrometer power switch is OFF.
See Appendix 1: Basic abling Diagram
1.2.1 AC Mains to the Power Supply
onnect the male plug of the IE 320 line cord to a grounded 100-120/220-240V A , 60/50 Hz.
mains outlet. Plug the female connector of the IE 320 line cord into the A receptacle of the
power supply.
1.2.2 UPS Connections (Optional)
onnect the male plug of the IE 320 line cord from the UPS unit to a grounded 100-120/220-
240V A , 60/50 Hz. Mains outlet. Plug the female connector of the IE 320 line cord into the UPS
input power receptacle. Plug the power cord from the external power supply into the UPS output
receptacle.
1.2.3 External Power Supply connection to the Spectro eter
Plug the external power supply onnector into the spectrometer’s power receptacle.
This is the smaller socket (labeled “POWER”).
1.2.4 Ethernet Connection
Plug the computer’s Ethernet connector into the spectrometer’s communication receptacle.
(labeled “ OMM”).
Switch the spectrometer ON. The RED status light, next to the power switch, should come on.
After the spectrometer boots the YELLOW status light should come on. (This will take
approximately 30 seconds.)
The spectrometer is ready.

5030 Miniature NIR Spectrometer User Gen Install Manual Page 6 of 29
DO : SUMan-5030GenInstallManual-v3-rev7-20180314
1.3 Battery Power to the Spectrometer (Optional)
The spectrometer is fitted with an external battery mount (referred to as a mount.), that allows
Rechargeable Ni ad batteries to be easily installed and removed. A fully automatic battery charger
provides support.
1.3.1 Battery installation
With the spectrometer power switch turned OFF remove the external power supply connector
from the spectrometer. Place a charged battery into the mount connector and slide it into
position so that it latches. Turn the spectrometer ON and observe that the RED status light comes
on.
NOTE: Do not apply power to the spectrometer with both the external power supply and battery
connected.
1.3.2 Battery Re oval
Press the latch at the top right side of the battery, slide the battery up and pull it out of the
-mount.
1.3.3 Battery charging
Place the battery into the -mount on the charger and slide it into place so that the latch
engages. harging is fully automatic. There is no need to fully discharge the battery prior to
charging and it may be left in place after it is fully charged. Refer to the owner’s manual for
information about the use of the charger.

5030 Miniature NIR Spectrometer User Gen Install Manual Page 7 of 29
DO : SUMan-5030GenInstallManual-v3-rev7-20180314
2.0 Communications
2.1 Ethernet Cabling
Remote control of the spectrometer through the use of a client or remote P is provided through an
Ethernet connection utilizing Twisted-Pair (10Base-T) cabling. The twisted-pair cabling for the Hand
Held Unit uses an RJ-45 modular connector at the client end; the connection at the spectrometer end
uses a latching cylindrical connector that is more robust than the RJ45 jack.
Two Ethernet 10Base-T devices can communicate only if the transmitter on one device is connected
to the receiver of the other device. Thus when connecting two identical twisted - pair ports a crossover
must be implemented in the wiring as in the following diagram. This crossover is configured in the 3
meter long Ethernet UTP cabling ( at.5) supplied with the unit.
Spectrometer onnector RJ45
3
1
4
2
(Rx+)
(Rx-)
(Tx+)
(Tx-)
1
2
3
6
(Rx+)
(Rx-)
(Tx+)
(Tx-)
If the 3 meter cable provided with the unit needs to be extended by use of a modular connector to an
additional cable then the two pairs of wires must be straight through.
RJ45 RJ45
1
3
2
6
(Rx+)
(Rx-)
(Tx+)
(Tx-)
1
2
3
6
(Rx+)
(Rx-)
(Tx+)
(Tx-)

5030 Miniature NIR Spectrometer User Gen Install Manual Page 8 of 29
DO : SUMan-5030GenInstallManual-v3-rev7-20180314
2.2 Wireless LAN (Optional)
To change the wireless network IP Address contact manufacture.
See Appendix 2 - WLAN Installation Diagram
2.2.1 Changing the co unication protocol
2.2.1.1 Accessing your spectro eter IP Address
1) Begin by removing the spectrometer cover to expose the serial keyboard and VGA
monitor ports.
2) Hook up a monitor and keyboard to your spectrometer.
3) After hooking up the monitor and keyboard, you should see “Luminar T P/IP onsole
Manager” as shown in Figure 1:
Figure 1

5030 Miniature NIR Spectrometer User Gen Install Manual Page 9 of 29
DO : SUMan-5030GenInstallManual-v3-rev7-20180314
4) From within the “Luminar T P/IP onsole Manager”, type “quit” to close the onsole
Manager and Exit to the DOS prompt.
5) You should now be in the “ :\FLASH>” directory. If not, type
“ D \FLASH” to navigate to the proper location.
Example: CD \FLASH
6) Type “edit pctcp.ini” to open the file as shown in Figure 2:
Example: C:\FLASH>edit pctcp.ini
Figure 2
7) Scroll down until you see the [pctcp ifcust 0] section and within this section you will
find your spectrometer IP Address. Write your IP Address below.
IP ADDRESS:
SUBMASK:

5030 Miniature NIR Spectrometer User Gen Install Manual Page 10 of 29
DO : SUMan-5030GenInstallManual-v3-rev7-20180314
8) When finished making your changes, press F7 to exit the editor. If prompted to save
the file, simply click the “enter” key.
9) Reboot your spectrometer by recycling the power off and back on.
2.2.1.2 Using the PUTGET progra to change IP Address
1) Execute the PUTGET program and then click the List Directory button.
2) hange to the “ :\FLASH” directory and then highlight "pctcp.ini" file.
Figure 3
Ce manuel convient aux modèles suivants
2
Table des matières
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur