© Kurth Electronic GmbH, All Rights Reserved
4
Operation
1. ON / OFF
Press the ON button for 2 seconds to turn on the light source. The display
will show the wavelength, battery level and modulation frequency. By default,
the device is set to the wavelengths 1310 nm (KE8200), 850 nm (KE8100)
and DC light (0 Hz). Press the OFF button for 2 seconds to turn o the unit.
The symbol in the lower left corner of the display indicates that the auto-o
function is active and that the unit switches o 10 minutes after the last key
was pressed. Briey pressing the OFF key switches o the auto-o function,
pressing the OFF key again reactivates the auto-o function.
2. Wavelength selection
Pressing the 𝛌button allows you to select the wavelength. The selected wa-
velength is shown in the centre of the LCD display.
3. Modulation output
Pressing the Mode button allows the selection of the dierent laser light mo-
dulations. The following options can be selected:
0 Hz ( DC ), 270 Hz, 330 Hz, 1 kHz and 2 kHz. The selected modulation
frequency is automatically detected by the KE8000 Optical Power Meter and
the measured value is displayed correctly. The selected modulation is shown
in the upper right corner of the display.
4. Backlight on / o
Pressing the 🔆 button turns the LCD backlight on or o. However, if the
backlight is turned o, it will be turned on for 8 seconds each time the key is
pressed.
5. Battery charge status
When the inserted batteries reach the lower state of charge, the battery in-
dicator changes to . If the battery voltage falls below the protective voltage,
the device switches o automatically.
6. Measurement
The measurement is usually carried out in combination with the KE8000 Op-
tical Power Meter. Please compare the explanations in the corresponding
operating instructions or on the following pages.