GMC-I Prosys CP 41 Manuel utilisateur

CP 41 / CP 410 /
CP 1010
AC / DC CURRENT
CLAMP METERS
Operating Instructions

2
Order Reference Order No.
CP 41 Current Clamp Meter P-12.230.0
CP 410 Current Clamp Meter P-12.480.0
CP 1010 Curent Clamp Meter P-12.601.0
Battery is included.
Thank you for buying this product. For safety reasons
and optimum use of this instrument read through the
operating instructions very carefully.
Table of Contents
1 SAFETY..........................................3
2
INTRODUCTION ............................4
3 SPECIFICATIONS..........................4
3.1 Electrical data.................................4
3.2 General data...................................4
4 OPERATING INSTRUCTIONS.......6
4.1 Switch On .......................................6
4.2 Auto Zero .......................................6
4.3 Current Measurement.....................6
4.4 Data Hold........................................6
4.5 Auto Power OFF.............................6
4.6 Battery Replacement......................6
5 SAFETY STANDARDS ..................7
6 WARRANTY...................................8
7 OTHER PRODUCTS......................8

3
1. SAFETY
The following symbols appear on the products:
Application around and removal from
HAZARDOUS LIVE conductors is
permissible
Complies with the relevant European
standards.
To avoid electric shock:
• Use caution during installation and use of this
product; high voltages and currents may be present
in circuit under test.
• This product must be used only by qualified
personnel practising applicable safety precautions.
• Do not use product if damaged.
• Always ensure the clamp meter is removed from
any live electrical circuit, before removing the
battery cover.
• Do not hold the clamp meter anywhere beyond the
tactile barrier see FIG 1.
Attention! Refer to Manual
Double/Reinforced Insulation
Do not dispose of this product as unsorted
municipal waste. Contact a qualified
recycler for disposal.
Read all instructions completely before
using this product.
UL Listing Mark
(CP 41 / CP 410)

4
2. INTRODUCTION
The CP 41, CP 410 and CP 1010 current clamp meters
have been designed for reliable and accurate, non-
intrusive measurement of AC, DC and complex
waveform currents.
Using advanced Hall Effect technology the CP 41/
CP 410 / CP 1010 can measure currents accurately
from 5mA to 40 Amps (CP 41) and from 100mA to
400A / 1000A (CP 410 / CP 1010).
Measurement features include:
True RMS reading
1mA / 10mA Resolution
Autoranging / Autozeroing / Auto Power Off
Data Hold
These features make it a powerful tool for use with
inverters, switch mode power supplies, industrial
controllers and other applications requiring current
measurement.
3. SPECIFICATION
3.1 Electrical data
(All accuracies stated at 23°C ± 1°C)
Measuring range (CP41) .......... 0 to 40A DC or AC
pk
(CP410)........ 0 to 400A DC or AC
pk
(CP1010)...... 0 to 1000A DC or AC
pk
Current Ranges (Autoranging).. 4A / 40A (CP41)
40A / 400A (CP410)
400A / 1000A
(CP1010)
Resoution (CP41)..................... 1mA / 10mA
(CP410) ..................10mA / 100mA
(CP1010)................. 100mA / 1A
Overload capacity..................... 200A / 600A / 1500A
.................................................(60s)
Overall DC accuracy................. ±1% of rdg ±5 digits
Temperature Coefficient........... ±0.1% of reading/°C
Frequency range (-0.5dB)......... DC and 15 – 400Hz
Working voltage ....................... 300V AC
RMS
or DC
3.2 General data
Operating temperature ............. 0°C to +50°C
Storage temperature with
battery removed ....................... -20°C to + 85°C

5
Power supply.........9 V Alkaline battery
..............................PP3, MN 1604
..............................or IEC6LR61
Battery life..............15 hours (CP41)
..............................50 hours (CP410 / CP1010)
Display...................4000 count
Conductor size.......25 mm diameter (CP41 / CP410)
..............................32 mm diameter (CP1010)
Weight...................235 g
12
3
4
7
9
54
6
8
FIG 1
1. Jaws
2. Conventional current direction
3. Tactile Barrier
4. Battery cover
5. Battery cover screw
6. Jaw trigger
7. ON/OFF and Range switch
8. Auto zero button
9. LCD Display

6
4. OPERATING INSTRUCTIONS
4.1 Switch On
Move the switch from OFF position to either DC or AC
to select the required mode of operation.
4.2 Auto Zero
When in DC mode the display zero may change due to
thermal shifts and other environmental conditions. To
null the display reading depress the Auto Zero button
>2s. Ensure that the clamp meter is away from the
current carrying conductor whilst the display reading is
being nulled.
4.3 Current Measurement
Select DC or AC option as required using the switch.
Zero the display reading using the Auto Zero button.
Clamp the jaws round the conductor ensuring a good
contact between the closing faces of the jaws.
Observe and take measurements as required. Positive
output indicates that the current flow is in the direction
shown by the arrow on the probe.
4.4 Data Hold
To activate the Data Hold press the Hold button once.
The last reading will be held on the display. Press the
hold button again to go back to the normal operation.
4.5 Auto Power OFF
In order to save battery life, the clamp meter will
automatically switch itself off after approximately 10
minutes. To disable the Auto power off function, Switch
Off the probe and Switch On whilst pressing the auto
zero button.
4.6 Battery Replacement
When the Low Battery symbol is illuminated in the
display the minimum battery voltage has been reached.
Refer to Fig 1 and use the following procedure to
replace the battery.
SAFETY WARNING
Before removing the battery cover,
make sure that the probe is remote
from any live electrical circuit.

7
Unclamp the clamp meter from the conductor, turn it off
using the On - Off switch.
Loosen the captive screw which secures the battery
cover. Lift the cover through 30°and pull it clea r of the
body as shown in Fig 1. Replace the battery and re-fit
the battery cover and fasten the screw .
Fit only Type 9 V PP3 Alkaline (MN 1604).
5. SAFETY STANDARDS
EN 61010-1:2001
EN 61010-2-032:2002
EN 61010-031:2002 +A1:2008
300V Cat III Pollution Degree 2
EMC Standards
EN 61326-2-2:2006
ROHS and WEEE compliant
This product is designed to be safe under the following
conditions:
- indoor use
- altitude up to 2000m
- temperature 0°C to +50°C
- maximum relative humidity 80% for
temperatures up to 31°C decreasing linearly
to 40% relative humidity at 50°C.
Use of the current clamp meter on uninsulated
conductors is limited to 300V AC
RMS
or d.c. and
frequencies below 1kHz.
Safety in its use is the responsibility of the operator
who must be a suitably qualified or authorised person.
Ensure that your fingers are behind the protective
barrier see FIG 1 when using the clamp meter. Always
inspect the clamp meter for damage before use.
To avoid electric shock, keep the clamp meter clean
and free of surface contamination.
Use Isopropyl alcohol to clean the clamp meter.

8
6. WARRANTY
Your GMC-I PROSyS clamp meter is guaranteed for
one year from the date of purchase against defective
material or workmanship. If the probe fails during the
warranty period, we shall at our discretion, repair or
replace it with a new or reconditioned unit provided we
are satisfied that the failure is due to defective material
or workmanship. To make a claim under warranty, the
probe should be returned to us, postage prepaid, with a
description of the defect. The use of a battery, other
than that specified invalidates this warranty.
Goods alleged by the buyer to be defective shall not
form the subject of any claim for injury, loss, damage,
or any expense howsoever incurred, whether arising
directly or indirectly from such alleged defects other
than death or personal injury resulting from the seller‘s
negligence.
No condition is made or to be implied, nor is any
warranty given or to be implied as to the life or wear of
goods supplied or that they will be suitable for any
particular purpose or for use under specific conditions,
notwithstanding that such purpose or conditions may be
made known to the seller.
7. OTHER PRODUCTS
GMC-I PROSyS offers a range of electrical test
products to ensure the safe and efficient operation of
electrical equipment and installations including:
Current probes: Flexible AC Probes and a wide
range of customised solutions
for AC only and AC / DC
current measurement
CP 41_CP 410English Rev 1.6

CP 41 / CP 410 /
CP 1010
AC / DC
Anzeigemessgeräte
Bedienungsanleitung

2
Bestellbezeichnung Bestell-Nr.
CP 41 AC/DC-Anzeigemessgerät P-12.230.0
CP 410 AC/DC-Anzeigemessgerät P-12.480.0
CP 1010 AC/DC-Anzeigemessgerät P-12.601.0
Batterien sind im Lieferumfang enthalten.
Vielen Dank für den Kauf dieses Produktes!
Aus Gründen der Sicherheit und der
bestmöglichen Verwendung dieses Gerätes bitten
wir Sie die Bedienungsanleitung
sorgfältig zu lesen.
Inhaltsverzeichnis
1 SICHERHEIT..................................3
2
EINLEITUNG..................................4
3 TECHNISCHE DATEN...................4
3.1 Elektrische Daten ...........................4
3.2 Allgemeine Daten ...........................4
4 BEDIENUNG ..................................6
4.1 Einschalten.....................................6
4.2 Nullabgleich....................................6
4.3 Strommessung ...............................6
4.4 Messwerte halten............................6
4.5 Automatisches Ausschalten ...........6
4.6 Austausch der Batterie ...................6
5 SICHERHEITSNORMEN................7
6 GARANTIE.....................................8
7 WEITERE PRODUKTE ..................8
Ce manuel convient aux modèles suivants
2
Table des matières
Langues :
Autres manuels GMC-I Prosys Instrument de mesure
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur