
8
Cannonball3’s zero to 100 percent LEL detection range for
units without a dilution pump, or zero to approximately
200 percent detection range for units with a dilution pump.
The safest course of action is to immediately leave the
affected area, and to return only after further testing
determines that the area is once again safe for entry.
The combustible gas alarm is activated when the percent LEL
(Lower Explosive Limit) gas concentration exceeds the pre-set
alarm level.
Two oxygen alarm set points have been provided; one for low
concentrations associated with oxygen deficiencies and one for
high concentrations associated with oxygen enrichments.
Two additional oxygen alarm set points have been provided for
Cannonball3 instruments equipped with the dilution pump for
use in inert or nearly inert atmospheres.
Three alarm set points have been provided for each toxic gas
monitored: TWA (Time Weighted Average), STEL (Short Term
Exposure Limit), and Ceiling.
Appendix A discusses alarm levels.
The procedure for adjusting alarm settings is covered in
section 3.5.2.
1.5.3 Sensor overrange alarms.
The Cannonball3 will go into alarm if a sensor is exposed to a
concentration of gas that exceeds its established range. If the
peak alarm is enabled it will go off before an overrange alarm
provided a STEL or TWA alarm is not activated first. If the
peak alarm is disabled and a toxic sensor goes into overrange
alarm a “SENSOR OUT OF RANGE” message will appear at
the bottom of the display while the audible and visible are
activated. The maximum range value will be displayed for the
sensor in alarm. If the LEL sensor goes into overrange alarm,
the message “LEL OVERRANGE” will appear at the bottom of
the screen, both the audible alarm and the flashing LED alarms
will be activated and an “X” will appear on the LCD in the place
of the numeric reading for the LEL sensor.
In the event of an alarm condition it is
important to follow established procedures. The safest
course of action is to immediately leave the affected area,
and return only after further testing determines that the
area is once again safe for entry.
In the event of an LEL overrange
alarm the Cannonball3 must be turned off, brought to an
area that is known to be safe and then turned on again to
reset the alarm.
Note: The Cannonball3 features automatic warning against
LEL sensor failure due to lack of oxygen. In the case of
oxygen levels below the safe limit for the LEL sensor to
operate, the Cannonball3 will display a message indicating
that O2is too low for LEL to operate.
1.5.4 Low battery alarms
Note: The voltage figures given below are for Cannonball3
instruments with firmware version 1.55 and may be
slightly different for units with other versions of
instrument firmware.
The Cannonball3 includes low battery alarms that are activated
whenever battery voltage approaches a level that will soon
lead to instrument shut down. When the battery voltage is
reduced to approximately 5.15 volts in alkaline units, or 4.82
volts in NiMH rechargeable units, an audible alarm will sound,
and the display screen will indicate that a low battery condition
exists. At this stage, the low battery alarms may be silenced
for up tofifteen minutes by pressing the MODE button. After
the first low battery alarm, the alarm will sound again every
fifteen minutes until the voltage drops to the “Very Low Battery”
level.
The “Very Low Battery” level occurs when the battery voltage
drops to 5.06 volts in alkaline units, or 4.75 volts in NiMH
rechargeable units. Due to the risk of imminent shut down,
when the battery voltage reaches the “Very Low Battery” level
it is no longer possible to silence the low battery alarms. At
this point, it is necessary to either immediately leave the
hazardous area in which the instrument is being used or to
immediately install a new battery pack with a charge sufficient
to avoid the low battery alarms.
When the voltage drops to 5.00 volts in alkaline units, or 4.73
volts in NiMH rechargeable units. the Cannonball3 will display
a "Dead Battery" message to warn the user of imminent shut
down. The instrument will then automatically turn itself off.
After any low battery alarm the batteries should be replaced if
the Cannonball3 is equipped with alkaline batteries or the
battery should be recharged if the Cannonball3 is equipped
with a NiMH rechargeable battery.
Use only Energizer E95 or EN95,
Duracell MN1300, or Duracell PC1300, 1.5V D cell Alkaline
batteries in the Cannonball3. Substitution of batteries may
impair intrinsic safety.
1.5.5 Calibration reminder
The Cannonball3 has a calibration reminder to let the user
know that the instrument is due for calibration. The reminder
can be set to any interval between 1 and 180 days, or can be
disabled entirely. Additional controls allow the user to
determine if the reminder will be shown once per operating
session, or every 15 minutes. Successful calibration resets the
reminder.
See section 4.7.4 for further details about the calibration
reminder.
1.5.6 Other alarms and special microprocessor features
Cannonball3 software includes a number of additional alarms
designed to ensure the proper use of the instrument. When
the Cannonball3 detects that an electronic fault or failure
condition has occurred, the proper audible and visible alarms
are activated and an explanatory message is displayed.
The Cannonball3 is designed to detect
potentially life threatening atmospheric conditions. Any
alarm condition should be taken seriously. The safest
course of action is to immediately leave the affected area,
and return only after further testing determines that the
area is once again safe for entry.
1.6 Other electronic safeguards
Several automatic programs prevent tampering and misuse of
the Cannonball3 by unauthorized persons. Each time the
detector is turned on, the Cannonball3 automatically tests the
LED alarm lights and audible alarm. The battery is monitored
continuously for proper voltage. The Cannonball3 also
monitors the connection of sensors that are currently installed.
The detection of any electronic faults causes the activation of
the audible and visible alarms and causes the display of the
appropriate explanatory message.