Explorer Group DLV40 Manuel utilisateur

Operator’s Manual:
DLV
40 Data logging Voltmeter
Rev: 02
Date: 5/28/2012
DLV40 Firmware: Rev 000.008 and above
Winmeter 5.0 Software: Rev 1.0.0 and above
Operator’s Manual:
40 Data logging Voltmeter
Operator’s Manual:
40 Data logging Voltmeter

Tab e of Contents:
1.0 Overview
1.1 User Interface
1.1.1 Graphica Back it LCD Disp ay (128x32)
1.1.2 Membrane Keypad
2.0 Menu System
2.1 Main Menu
2.2 Discharge Sub-Menu
2.3 “20 x 48” Sub-Menu
2.4 “8 x 256” Sub-Menu
3.0 Power Saving Features
3.2 Auto-Off
3.2 Low Battery Indication
3.3 Checking battery Vo tage
4.0 Taking a Measurement (ALL Modes)
4.1 Over Vo tage Indication (ALL Modes)
4.2 Recording Over a Previous Measurement (ALL Modes)
4.3 De eting a Previous Measurement (ALL Modes)
5.0 Discharge Test Procedure
5.1 Step1: Se ect DLV40 “Discharge Test” Mode & C ear Existing Test Data
5.1.1 Option 1: Using Winmeter 5.0 Software
5.1.2 Option 2: Manua y
5.2 Step 2: Measure String F oat Ce Vo tages
5.3 Step 3: Starting the Discharge Test Timer & Taking Discharge Readings
5.4 Step 4: Up oad Discharge Test Data to Winmeter 5.0
5.5 Step 5: Generate Ce Discharge Report(s)
5.6 Step 6: Save Ce Discharge Report(s)

6.0 F oat Test Procedure (“8x256” and “20x48” Modes)
6.1 Step1: Se ect DLV40 “8x256” /“20x48” Mode & C ear O d String Data
6.1.1 Option 1: Using Winmeter 5.0 Software
6.1.1 Option 2: Manua y
6.2 Step 2: Measure String F oat Ce Vo tages
6.3 Step 3: Up oad “8x256/20x48” Test Data to Winmeter 5.0
6.4 Step 4: Se ect F oat String to Save
6.5 Step 5: Generate Ce F oat Report(s)
6.5.1 Add/Edit Report Information and Parameters
6.5.2 View/Print Report(s)
6.5.2.1 Graphs
6.5.2.2 Tabu ar Report
6.5.2.3 Saving Report to PDF or Exce Format
6.5.3 Export Data to Exce (Optiona )
6.6 Step 6: Save Ce Test Report(s)
7.0 Connecting the DLV40 to PC/Winmeter 5.0
8.0 Upgrading DLV40 Firmware
For the most accurate results, please read and follow these instructions carefully.

1.0 Overview
The DLV40 data logging voltmeters have been designed specifically for the battery test industry. The
DLV40 is designed to measure and automatically record DC cell voltages between 0.1 VDC and 1 .
VDC (inclusive). It has a built in real time clock and calendar which allows for date and time stamping of
readings. The stored readings can then be transferred to a PC via a standard USB connection and the
companion Winmeter 5.0 Battery Analysis Software.
The DLV40 is capable of operating in (3) distinct measuring modes:
(i) “8 by 256” Mode: The DLV40 can measure and store up to 2048 readings of cell voltage.
These readings are stored in 8 separate data strings of 256 readings, denoted A through H.
The date and time of the last reading in each string is also recorded.
(ii) “20 by 48” Mode: The DLV40 can measure and store up to 60 readings of cell voltage.
These readings are stored in 20 separate data strings of 48 readings, denoted A through T.
The date and time of the last reading in each string is also recorded.
(iii) “Discharge Test” Mode: The DLV40 can measure and store up to 2048 readings of cell
voltage. These readings are stored in 8 separate data strings of 256 readings, denoted A
through H.
Readings in String A are reserved for the initial (f oat) voltage of the cell and are NOT time
stamped.
Readings in Strings “B” through “H” are reserved for cell voltages during the discharge (or
charge) test with each reading being time stamped to the nearest second from the start of
the discharge test. The date and time of the last reading is also recorded.
All stored readings can be downloaded via Winmeter 5.0 software to generate detailed test reports
including statistical and graphical analysis and then stored into a custom database. This software
communicates with the DLV40 via USB and allows the User to set the time/date, change modes, delete
data string(s) and upgrade the DLV40 firmware via PC interface.
All
DLV40 modes operate
independent
y
. Erasing the memory in one mode does
NOT
affect readings in the remaining (2) modes
.

1.1 User Interface
1.1.1 Graphica Back it LCD Disp ay (128x32)
1.1.2 Membrane Keypad
MENU
From measurement mode:
Push once to enter the
sub-menu.
Push twice to enter Main
menu.
ON/OFF:
“Short” push to turn ON
“Long” push to turn OFF
UP & DOWN
Main or sub menu mode:
Push to change the current
selection.
Measurement mode: Push
to change selected
readings.
ENTER
Push to accept the current
selection.
Dischg
0:00:03
+0.000
B:001
000.0
Sum of cell voltages
for the selected
string.
Shows 000.0 as no
discharge readings
have been taken.
Discharge Time
(Currently 3 seconds
into the test).
Selected String (“B”) and
the next reading in the
selected string (“001”).
(Float readings have been
taken but no discharge
readings have been
taken).
“Discharge Test”
Mode
Current measurement
mode.

2.0 Menu System
The DLV40 uses a simple two tier menu system. The “Main Menu” and (3) sub-menus, one for each
measurement mode (“8 by 256”, “20 by 48” and “Discharge Test”).
2.1 Main Menu
“MENU” Button:
When the DLV40 is in one of the measurement modes, push the MENU button once to enter the
appropriate sub-menu, push the MENU button a second time to enter the Main Menu.
When the DLV40 is in the Main Menu, push once to cancel the current selection.
“ENTER” Button:
Push the ENTER button to accept the current selection.
“UP” and “DOWN” Buttons:
In Main or sub-menu modes push to change the current selection.
In any of the (3) measurement modes push once to increment/decrement the cell pointer, push and
ho d to scroll through the readings.
*** Main Menu***
MODE : 8 by 256
*** Main Menu***
About DLV40
*** Main Menu***
Date & Time?
*** Main Menu***
De ete ALL Data
*** Main Menu***
MODE : DISCHARGE
*** Main Menu***
MODE : 20 by 48
DLV
40
Rev:
000.005
09-15-11 14:30:21
Battery 95% Ca : 06-15-11
Disp ays Date & Time
(Date/Time is set via USB
connection with Winmeter 5.0)
*** Main Menu***
De ete ALL Data?
*** Main Menu***
De ete Data 05%
Discharge Test Mode
(Please refer to Section 5.0 for the
Discharge Test Procedure )
20 by 48 Mode
(Please refer to Section 6.0 for the
Float Test Procedure(s) )
8 by 256 Mode
(Please refer to Section 6.0 for the
Float Test Procedure(s) )

2.2 “Discharge” Sub-Menu
*** Discharge Menu***
De ete B:003?
*** Discharge Menu***
Exit?
*** Discharge Menu***
Start DC Test?
*** Discharge Menu***
Change String?
*** Discharge Menu***
Print String?
Discharge Test Mode
(Please refer to Section 5.0 for the
Discharge Test Procedure )
*** Discharge Menu***
Print VDC String B?
*** Discharge Menu***
Printing Data…
*** Discharge Menu***
Printing Comp ete
Discharge Test Mode
(Please refer to Section 5.0 for the
Discharge Test Procedure
)
*** Discharge Menu***
Reading De eted
Discharge Test Mode
(Please refer to section X.X for
more information)
*** Discharge Menu***
De ete DC Test?
OR
Start/De ete Discharge Test
Please refer to section X.X (Start)
and section X.X (Delete) more
information.
Discharge Test
String A 024 Readings
UP/DOWN
to change strings
ENTER to select and return to
Discharge Test Mode
UP/DOWN
to change strings

2.3 “20 by 48” Sub Menu
*** 20 x 48 Menu ***
De ete J:025?
*** 20 x 48 Menu***
Exit?
*** 20 x 48 Menu***
Reading De eted
20 x 48 Test Mode
(Please refer to section X.X for
more information)
20 x 48 Test Mode
(Please refer to Section 6.0 for
the Float Test Procedure(s) )
*** 20 x 48 Menu***
De ete String J?
*** 20 x 48 Menu***
De eting String J…
20 x 48 Test Mode
(Please refer to section X.X for
more information)
*** 20 x 48 Menu***
Change String?
*** 20 x 48 Menu***
Print String?
*** 20 x 48 Menu***
Print VDC String J?
*** 20 x 48 Menu***
Printing Data…
*** 20 x 48 Menu***
Printing Comp ete
*** 20 x 48 Menu***
String J 024 Readings
UP/DOWN
to change strings
ENTER to select and return to
Discharge Test Mode
UP/DOWN
to change strings
20 x 48 Test Mode
(Please refer to Section 6.0 for
the Float Test Procedure(s) )

2.4 “8 by 256” Sub Menu
*** 8 x 256 Menu ***
De ete A:025?
*** 8 x 256 Menu***
Exit?
*** 8 x 256 Menu***
Reading De eted
8 x 256 Test Mode
(Please refer to section X.X for
more information)
8 x 256 Test Mode
(Please refer to Section 6.0 for
the Float Test Procedure(s) )
*** 8 x 256 Menu***
De ete String A?
*** 8 x 256 Menu***
De eting String A…
8 x 256 Test Mode
(Please refer to section X.X for
more information)
*** 8 x 256 Menu***
Change String?
*** 8 x 256 Menu***
Print String?
*** 8 x 256 Menu***
Print VDC String B?
*** 8 x 256 Menu***
Printing Data…
*** 8 x 256 Menu***
Printing Comp ete
*** 8 x 256 Menu***
String B 024 Readings
UP/DOWN
to change strings
ENTER to select and return to
Discharge Test Mode
UP/DOWN
to change strings
8 x 256 Test Mode
(Please refer to Section 6.0 to for
the Float Test Procedure(s) )

3.0 Taking a Ce Vo tage Measurement (ALL Modes)
Cell voltage measurements are taken in a similar fashion to a traditional voltmeter.
(i) Turn the DLV40 ON.
(ii) Select the required measurement mode (“20x48”, “8x256” or “Discharge”).
(Please refer to section X.X for Mode selection for more information)
(iii) Select required string.
For “20x48” or “8x256” mode:
- If old data is present, de ete string(s) data as required.
For “Discharge Test” mode:
If an old Discharge Test data is present, de ete the existing Discharge Test data.
The DLV40 is designed to measure absolute DC cell voltages between 0.1 VDC and 1 . VDC.
Attempts to measure higher DC voltages will result in an “OVERVOLTAGE!!!” warning.
8x256
+0.000
B:001
000.0
Sum of cell voltages
for the selected string
(shows 000.0 as no
discharge readings
have been taken).
Current measurement
mode.
Selected String (“B”) and
the next reading in the
selected string (“001”).
(No readings have been
taken).
When turned
ON
, the DLV40 will automatically resume in the previous mode and “point” to the next
reading location.
“20x48” or “8X256” Mode
Dischg 0:00:03
+0.000
B:001
000.0
Sum of cell voltages
for the selected string
(shows 000.0 as no
discharge readings
have been taken).
Discharge Time
(Currently 3 seconds
into the test)
.
Selected String (“B”) and
the next reading in the
selected string (“001”).
(Float readings have been
taken but no discharge
readings have been
taken).
“Discharge Test”
Mode
Current measurement
mode.
The DLV
40 USB cable
MUST
be
DISCONNECTED
whenever measuring voltages
!
Table des matières
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur