ETCR ETCR6200 Manuel utilisateur

ETCR6200 AC/DC
Leakage Current Clamp Meter
User Manual

CONTENT
Warning....................................................................................................... 1
I. Introduction..............................................................................................2
II. Range and Accuracy.............................................................................3
III. Electrical Signs..................................................................................... 3
IV. Technical Specifications......................................................................4
V. Structure of the Leaker.........................................................................6
VI. LCD Display..........................................................................................7
VII. Operating Method...............................................................................7
1. Startup, Shutdown............................................................................ 7
2. AC / DC switch.................................................................................7
3. Reset................................................................................................. 8
4. Current Measurement.......................................................................8
5. Peak Holding.................................................................................. 10
6. Hold, Storage and Access Reading................................................ 10

7.Data upload...................................................................................11
VIII.Application.......................................................................................11
IX. How to Change Battery.....................................................................12
X. Accessories..........................................................................................13

Warning
Thank you for purchasing our company's ETCR6200 AC/DC Leakage
Current Clamp Meter, for better use of this product, please:
——Read carefully the user's manual.
——Follow strictly safety rules and notes listed in this manual.
Under any circumstances, please pay special attention to your safety in
the course of using this leaker.
Give heed to label texts and symbols on panel and back plate of this
leaker.
Please be more careful if the line voltage is above 60VDC or 30VAC.
When test AC current, the wire should be in the jaw center, otherwise the
error will increase.
Please don't place and store this leaker in hot and humid condition,
locations with moisture condensation and under direct sunlight for a long
time.
In case voltage of battery was low, please replace batteries.
If you expect not to use the leaker for a long time, please take out
batteries.
When changing batteries, please pay attention to the polarity of battery.
Use, disassembly and maintenance of this leaker shall be operated by
authorized personnel.

In case dangers would occur if continue to use a faulty leaker, please stop
to use it and seal it for safekeeping immediately; and then, send it for
disposal of authorized agency.
Users shall carry out operation based on danger signs “ ” on leaker
and manual.
Users shall carry out safety operation based on instructions listed in this
manual, e.g. “ ” and danger signs on this manual.
I. Introduction
ETCR6200 AC/DC Leakage Current Clamp Meter also named: Car Leakage
Current Clamp Meter, Car Standby Current Clamp Meter, Car Inrush Current
Tester, Car Charge Current Tester. It is specially designed for online
measurement of DC, AC leakage current, current below 600V by adopting
up-to-date CT technology and digital integration technology. It is specially
suitable for measure car standby current, leakage current, inrush current. It is
a product with the characteristic of relatively small size, high accuracy, strong
anti-interference capability and automatic shifting. It is apply to car leakage
rapid detection, consumed current detection, quickly find the fault of the wire
broke leakage, equipment fled electric, line contact with iron, measuring the
motor inrush current and alternator charging current etc. It could be widely
use in those fields as car repair, car leave factory test, car parts test, the
current detection in telecommunication room, electric power overhauling,
meteorology, oil detection, etc.
It has those functions as peak holding, data holding and data storage. It

obtains one RS232 interface, communication cable and software, through
the computer, it can online real-time monitoring, historical query, dynamic
display, MIN-AVG-MAX indication, alarm setting and indicating, data reading,
saving and printing.
II. Range and Accuracy
Function
Measure Range
Accuracy
Resolution
DC current
0.0mA~60.00A DC
±2%rdg±5dgt
1mA
AC current
0.0mA~60.00A DC
±2%rdg±5dgt
1mA
(Condition: 23°C±5°C, below 75%rh, wire in the center of jaw)
III. Electrical Signs
Extremely dangerous! Operators shall strictly keep to
safety rules; otherwise there would be dangers of electric
shock to cause personal injuries or casualties.
Dangerous!Operators shall strictly keep to safety rules;
otherwise there would be dangers of electric shock to
cause personal injuries or casualties.
Warning !Operators shall strictly keep to safety rules;
otherwise personal injuries or equipment damages might
be caused.

Double insulation
AC
DC
IV. Technical Specifications
Model
ETCR6200
Function
Measurement of AC/DC current, leakage
current, repairing the car circuit.
Power Supply
Zn-Mn dry battery, 6F22, 9V, continuous use for 30h
Test Mode
Clip-on CT, noncontact measurement
Clamp Size
25mm×35mm
Measurement
Range
0mA~60.0A AC/DC
Resolution
1mA AC/DC
Accuracy
±2%rdg±5dgt (23°C±5°C, below 75%rh)
Display Mode
Four digits LCD display
Dimension
LWH: 168mm×65mm×34mm
LCD
Dimension
35mm×21.5mm; display domain: 32mm×15mm
Sampling Rate
2 times/s

Frequency
AC: 45Hz-400Hz
Polarity
Indication
DC current auto identified and display“—”
Test Position
Tested wire in the jaw center
Range Shift
Automatically
Line Voltage
Do not exceed AC 600V
RS232
Interface
Data stored in the memory of the meter via RS232
upload to PC
Com-Configure
Baud rate:9600, data bit:8, stop bit:1, check bit:
NONE
Data Memory
99units, FULL blinks when the memory is full
Reading Hold
DH indicating the reading is hold
Out of Range
OL indicating the current is out of range
Auto Power-off
5 minutes after power on, it will power off
automatically to lower the power consumption
Battery Voltage
Indicating the battery voltage is lower than
7.2V.Then the battery has to be changed.
Weight
180g (including the battery)
Working
Current
Max 20mA
Humiture
-10°C-50°C, below 80%rh
Limit
Temperature
-10°C-0°C and 40°C-50°C, within 2%rdg

Error
Humiture
Storage
-10°C-60°C,below 70%rh
Insulating
Strength
AC 3700V/rms. (between the alloy of the clamp and
the housing)
Applicable
safety rules
IEC1010-1, IEC1010-2-032, pollution:2, CAT III
(600V)
Accessories
Meter: 1pc; toolkit: 1pc; RS232 communication line:
1pc; disk: 1pc; case: 1pc.
V. Structure of the Leaker
1. Clamp.
2. Totoid opening lever.
3. POWER key.
4. LCD.
5. USB interface.
6. DC current positive input indication.
7. HOLD Key.
8. AC / DC switch key.
9. ZERO Key.
10. The screw of connect the upper and lower cover together.
11. The screw to fix the battery cover.

12. Battery cover.
VI. LCD Display
1. DC indication
2. AC indication
3. AC indication
4. Low battery symbol
5. Data hold indication
6. Stored data code
7. Unit of current indication 8. Value of current
VII. Operating Method
1. Startup, Shutdown
Press POWER key to start up, LCD will begin to display; Press the POWER key
again, the leaker will shut down. After starting up for 5 minutes, LCD will flick
notes that the leaker will shut down automatically; after flickering for 30s, it
will shut down formally to reduce battery consumption. In case you have
pressed POWER key when LCD was flickering, the leaker will continue to
work for 5 minutes. If LCD was very dark after starting up, it might be caused
by low-voltage battery, in this case, please change battery immediately.
2. AC / DC switch
After startup, instrument default into DC current testing. Press to
switch the AC and DC current. LCD display DC indicates DC current, display
Table des matières
Autres manuels ETCR Instrument de mesure

ETCR
ETCR ETCR 7000 Manuel utilisateur

ETCR
ETCR 6300D Manuel utilisateur

ETCR
ETCR 9000 Manuel utilisateur

ETCR
ETCR 6000 Manuel utilisateur

ETCR
ETCR 7100 Manuel utilisateur

ETCR
ETCR 6100 Manuel utilisateur

ETCR
ETCR 9200 Manuel utilisateur

ETCR
ETCR 2800B Manuel utilisateur

ETCR
ETCR ETCR 4000A Manuel utilisateur

ETCR
ETCR 6300 Manuel utilisateur
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur