9
USER MANUAL
Inteli-Mixer RM-2
Example 1. Racks rotates clockwise 180 deg. then
counter clockwise 270 deg. Store program in C1.
1.1 Select C1 on F display.
1.2 Hold button for 4.5 seconds. Rack will auto-
matically nd 00 initial point.
1.3 Pressing clockwise rotation button set 50
value on RPM indicator, that corresponds to
180 degrees (see Fig. 3).
1.4 Press enter button
1.5 Pressing counter clockwise rotation button
turn the rack until 75 value on RPM indica-
tor, that corresponds to 270 degrees (see Fig.
3).
1.6 Press enter button
1.7 Pressing clockwise rotation button or coun-
ter clockwise rotation button se 00 value
on RPM indicator, Findicator will show P0.
1.8 Press enter button
Program is successfully recorded.
Example 2. (Features entering position correction).
Rack rotates clockwise 180 deg., then counter clock-
wise 180 deg., entering position correction in 00
deg. point then rotates 90 deg. counter clockwise.
Store program in C2.
2.1 Select C2 on F display.
2.2 Hold button for 4.5 seconds. Rack will auto-
matically nd 00 initial point.
2.3 Pressing clockwise rotation button set 50
value on RPM indicator, that corresponds to
180 degrees (see Fig. 3).
2.4 Press enter button
2.5 Pressing counter clockwise rotation button
turn the rack until 00 value on RPM indica-
tor.
Caution ! reaching 00 point automatically is
oered P0 function that sets end of program. If
you would like to continue programming do not
press button on this step.
2.6 Choose SP function on Fdisplay.
2.7 Press enter button (position correction en-
tered)
2.8 Pressing counter clockwise rotation button
turn the rack until 75 value on RPM indica-
tor, that corresponds to 270 degrees (see Fig.
3).
2.9 Press enter button
2.10 Pressing clockwise rotation button or coun-
ter clockwise rotation button set 00 value on
RPM indicator, Findicator will show P0.
2.11 Press enter button
Program is successfully recorded.(actions 1.5, 1.6,
1.7 – position correction )
Example 3. Perform small amplitude shaking ‘u’
during 3 sec., then rotate counter clockwise 90 deg.,
then hold 1.5 sec., then shake 1.5 sec. with big am-
plitude ‘U’ & 1.5 sec. with small amplitude ‘u’, then
rotate clockwise 180 deg., then rotate counter clock-
wise to the 00 deg. point and hold 1.5 sec. Store the
program in C3.
3.1 Select C1 on Fdisplay.
3.2 Hold button for 4.5 seconds. Rack will au-
tomatically nd 00 initial point.
3.3. Select Pu function on Fdisplay.
3.4. Press enter button . 1.5 sec shaking is en-
tered.
3.5. Press enter button . Extra 1.5 sec shaking is
entered.
3.6. Pressing counter clockwise rotation button
turn the rack until 75 value on RPM indicator,
that corresponds to 90 degrees (see Fig. 3).
3.7. Press enter button .
3.8. Select PP function on F display 1.5 sec. pause.
3.9. Press enter button .
3.10. Select PU function on Fdisplay 1.5 sec. big
amplitude shaking.
3.11. Press enter button .
3.12. Select Pu function on Fdisplay 1.5 sec. small
amplitude shaking.
3.13. Pressing clockwise rotation button turn the
rack until 25 value on RPM indicator, that cor-
responds to 180 degrees (see Fig. 3).
3.14. Pressing counter clockwise rotation button
turn the rack until 00 value on RPM indicator.
Caution ! reaching 00 point automatically is
oered P0 function that sets end of program. If
you would like to continue programming do not
press button on this step.
3.15. Select PP function on Fdisplay 1.5 sec. pause.
3.16. Press enter button .
3.17. Select P0 function on Fdisplay.
3.18. Press enter button .
Program is successfully recorded.