-2-
3. Classifications
[ Classification under the provision of 93/42/EEC(MDD)] ClassⅠ
ECP-9000 is classified into ClassⅠsystem.
[Protection method against electric shock] ClassⅠ
ECP-9000 is Classified into ClassⅠinstrument.
ClassⅠinstrument in which protection against electric shock does not rely
on basic insulation only, but which includes an additional safety precaution
with the provision of the connection of the instrument to the protective
earth conductor in fixed wiring of the installation in such a way that any
accessible conductive parts cannot become live in the event of a failure
of the basic insulation.
[Degree of protection against ingress of liquids] IPX0
ECP-9000 is the ordinary instrument (enclosed instrument without
protection against ingress of liquids). Be careful not to splash water
on the instrument.
[Degree of protection against flammability]
ECP-9000 is classified as an equipment not suitable to be used in a
Potentially flammable anesthetic mixture with air or with oxygen
or nitrous oxide.
Do not use near flammable materials.
[Mode of operation]
Continuous operation