Page 3of 4
Setting the Output Options >OPT<
1. At the Main Menu use the 4 and 6 keys, scroll to OPT (Output Options) then press ENTER.
2. Choose Channel A (or B) then press ENTER
3. The 6000 SERIES will prompt you for the Physical Quantities for your desired output. Use
the 2 and 8 keys to scroll up and down through the available list. Press ENTER.
4. Using the 2 and 8 keys, choose your Velocity In units from the available list then press
ENTER.
5. Answer NO to the prompt for Temperature T1 then press ENTER
6. Answer NO to the prompt for Temperature T2 then press ENTER
7. Choose a Damping (averaging) time in seconds then press ENTER
8. Answer NO to Store Measured Data then press ENTER or YES if required.
9. Answer NO to Serial Output then press ENTER or YES if required.
10. Enter YES to the Current Loop I1: prompt.
11. Choose either Absolute or Signed at the Measured Values prompt to indicate if you
would like only positive (Absolute) or both positive and negative (Signed) flow.
12. At the Zero-Scale Value prompt enter a value that will represent the 4 mA output signal.
13. At the Full-Scale Value prompt enter a value that will represent the 20 mA output signal.
Once entered you will return to the Main Menu.
Enter the Measure Mode >MEA<
1) Using the 4 and 6 keys, scroll back to MEA (Measuring) then press ENTER
2) Below each channel will appear a minus sign (-), a check mark (√)or a dot (∙). Using the 4
and 6 keys, select Channel A (or B), then ensure there is a check mark below that channel by
pressing the 2 or 8 keys. Then press ENTER
3) The next prompt to appear may be different, depending upon how the 6000 SERIES has been
initialized. If Reflection Mode appears, continue with step 4 below. If a prompt appears
asking for a Sound Path, continue with step 5 below.
4) If Reflection Mode prompt appears, choose NO and the 6000 SERIES will measure in the
Direct Mode, where the transducers are mounted on opposite sides of the pipe. If you
choose YES, the 6000 SERIES will measure in the Reflection Mode where the transducers
are mounted on the same side of the pipe. It is recommended you answer YES and mount the
transducers on the same side of the pipe. However, if problems are experienced getting good
solid flow data, it is recommended you go through the MEA routine again, this time answering