Introduction
Thank you for choosing one of our precision instruments. Your balance is
designed and manufactured to the most rigorous standards in order to give you
years of service. First, check the contents of the shipping carton. You should find
the following:
• Manual • Balance • AC Adapter • Cover Pan
Next, follow the instructions for installing your balance.
Now you’re ready to begin using your balance. To take advantage of its many
features, carefully read your operating manual. It contains step-by-step proce-
dures, examples, and other vital information.
Finally, remember to return your completed warranty card within ten days and
to record all purchase information.
Warning: Use of this product in a
manner not specified by the manufacturer may impair any safety protection
provided by the equipment!
Specifications
Model XP-300 XP-600 XP-1500 XP-3000
Range 300g 600g 1500g 3000g
Readability 0.01g 0.02g 0.05g 0.1g
Linearity ±0.01g ±0.02g ±0.05g ±0.1g
Pan 5” x 7” 5” x 7” 5” x 7” 5” x 7”
Dimensions (127 x 178mm) (127 x 178mm) (127 x 178mm) (127 x 178mm)
Permissible
Calibration 200g 500g 1000g 2000g
Weights
Temperature 15 - 40 °C
Humidity 80% for temperature to 31°C, decreasing linearly to 50% relative
humidity at 40°C
Altitude 3000m
Common Specifications: Unit dimensions (L x W x H) 8.2 x 8.3 x 2.7” (208 x 211 x 69mm);
Net Weight: 3.75 lbs. (1.7 kg). Power: 15 VDC, 100mA, AC adapter included. Mains supply
voltage fluctuations not to exceed ±10% of nominal supply voltage. Equipment id suitable
for continuous operation with AC adapter/charger. Pollution degree:2; Installation catego-
ry: II; Sound Pressure Level emitted by equipment does not exceed ambient noise.
Storage/Transport Temperature: -20° – +40°C.
NOTE: The battery is not a user serviceable part. If the battery does not
recharge with use of AC adapter, contact an authorized service technician.
Caution: Use AC adapter/charger supplied with unit only!
Consult Denver Instrument Company for replacement.
!
!
Maintenance and Troubleshooting
Display Shows: Cause: Remedy:
I/O key pressed Press the I/O key.
and display is off.
Power cord not Connect cord.
connected.
No power to outlet or Check power supply.
improper voltage. and voltage switch.
Temporary fault Disconnect and reconnect
power cord. (Wait at least five
seconds before reconnecting.)
Weight exceeds Reduce weight.
balance capacity.
Pan obstructed. Check pan placement.
Balance operating error. Re-calibrate balance.
Load cell damaged. The most common cause
is from objects being dropped
onto the weighing pan or
from the balance itself being
dropped. The balance must be
returned for service.
Air movement Move balance to stable,
around balance. draft-free location.
Usually detected when The balance must be placed in
the calibration operation a stable environment before
is being attempted. It is calibration occurs.
usually the result of the
balance being exposed
to vibrations, air drafts or
unstable environments.
Pan was not in place Turn unit off and place pan on
when unit was turned on. and turn unit on.
This is an indication that The unit must be returned
an electronic malfunction to the factory for evaluation.
has occurred.
This fault will appear on the The most common cause
display when the unit is is from objects being dropped
powered on. This occurs onto the weighing pan or
when the electronics are no from the balance itself being
longer within the factory set dropped. The balance must be
parameters. returned for service.
(Blank Screen)
(No stability indicator)
E-02
E-30
E-54