© 2022, Delta-T Devices Ltd Page 5
Configuration
Press [SET] to configure the meter or sensor. Scroll up or down to the option that you
want to configure, and press [SET] again to select. There are 5 menu options: Soil
type, Soil setup, Units, Address, and Contrast.
The WET150 measures permittivity, bulk conductivity and temperature directly. From
these measurements the Meter calculates pore water conductivity (ECp) and water
content. These calculations require user-configurable parameters which are stored on
the Meter handheld unit. By default, the Meter comes with generic parameters that
give comparable performance to the generic parameters in Delta-T’s well-regarded
WET2 sensor. We strongly recommend you review these settings. Best results will be
obtained with customised parameters for your specific medium; Delta-T also supplies
a list of generic parameters which give good performance for several common soils
and substrates. See the WET150 User Manual (available online) for more details.
Soil type
Select the soil type that will be used when calculating water content and ECp.
Definitions can be found in the WET150 User Manual.
Soil Setup
The ”Soil Setup” menu enables you to adjust the following parameters for each soil
type (except RAW):
▪Moisture a0
▪Moisture a1
▪Soil Parameter –this adjusts how the Meter calculates ECp
▪ECp dry limit –at low water contents, calculating ECp from permittivity and
ECb becomes unreliable. This option defines the “too dry” permittivity level
below which the Meter will not output ECp due to declining accuracy.
The “Back” option will return you to the list of available soil types.
The “Restore defaults” option will reset the selected soil type to its factory settings.
IMPORTANT: The configuration menu only adjusts the settings within the Meter
handheld unit. The only setting stored on the sensor which can be adjusted via the
Meter handheld unit is the SDI-12 address.
Units
This menu enables you to select which units are used to display EC and temperature
measurements.
These parameters adjust how the Meter calculates water
content