User Menu Parameters
50
-10
0,8
-2.0
Hold T + H together for 3 seconds to enter menu.
-Use T /H buttons to choose parameter
-Use U/D buttons to change parameter values
-Hold T + H together for 3 seconds to save changes
hhi: high humidity alarm.
Started : set value +10
Delay time: 20 minutes
Notification: Alr 5
Voice alarm : continuous
hlo: low humidity alarm
Started : set value -10
Delay time: 20 minutes
Notification: Alr 6
Voice alarm : intermittently
(f)thi: high temperature alarm
Started : set value + 0,8
Delay time: 2 minutes
Notification: Alr 1/3
Voice alarm : continuous
(f)tlo: low temperature alarm
Started : set value - 2 0
Delay time: 30 minutes
Notification: Alr 2/4
Voice alarm : intermittently
tco: temperature calibration value
CAUTION! tCo is a critical parameter . Changing this can
results in chick looses. Plsease use qualified and
calibrated glass or electronic thermometers for
calibration purpose. Check temperature values regularly.
hco: humidity calibration value.
CAUTION! hCo is a critical parameter. Changing this can
results in chick looses. Please use qualified and
calibrated glass or electronic thermometers for
calibration purpose. Check humidity values regularly.
th
18
28
10
off
0.0
0.0
US2: USB connected device
no : no connected device
th : second sensor connected ( room/calibration)
ıot: iot module connected
Notification: Alr 15 (usb connection error)
below parameters active if Us2 is selected “th”
(f)rtl: second sensor low temperature alarm
Started : below 18 C / 65 F
Delay time: no
Notification: Alr 10
Voice alarm : intermittently
(f)rth: second sensor high temperature alarm
Started : over 28 C / 83 F
Delay time: no
Notification: Alr 9
Voice alarm : intermittently
rhl: second sensor low humidity alarm
Started : below 10 RH
Delay time: no
Notification: Alr 14
Voice alarm : intermittently
rhh: second sensor high humidity alarm
Started : over 85 RH
Delay time: no
Notification: Alr 13
Voice alarm : intermittently
rAt: second sensor temperature alarm activation
On: alarms activated
Off: alarms deactivated
rAt: second sensor humidity alarm activation
On: alarms activated
Off: alarms deactivated