INTRODUCTION:
In dealing with liquid hazardous waste companies have had to pay a costly rate for pick-
up
and repurchase of new chemicals for their process. This recycler reduces these costs.
The means of accomplishing the above is through a process known as distillation. Each
group of chemicals has a vapor temperature, which we utilize to separate from
contaminants. This unit is explosion proof, which means it can be used for flammable and
non
-
flammable solvents ranging from acetone to water.
SAFETY FEATURES:
The unit comes equipped with the following patent pending safety features:
A)
Automated temperature and time setting. The unit will automatically set its own time
and temperature in accordance with the requirements of the particular solvent or
solvents poured into its distillation chamber for recycling. The result being the
operator only needs to pour the solvent in and press START .
B)
Oil immersion
element.
C)
Element control and safety probes.
D)
Oil maintenance timer and indicator. The unit will automatically indicate when oil
requires to be changed. If oil is not changed within a safe time period, the unit will
automatically shut down indicating the word OIL on its display panel until oil is
changed. This ensures proper and safe operation of the unit for years to come.
E)
Intrinsically safe controls.
F)
Self
-
cooling condensation drum.
G)
A superior vapor management system. The vapor management system limits VO
C
emissions during distillation, resulting in a safe healthy working environment with an
ultimate recovery of solvent.
H)
OFF button for manual shut off at any time. If the unit is in cycle and the stop
button is pressed the unit will shut off and automatically go into cool down mode
with yellow light flashing and digital displaying temperature until the unit reaches
134.6
Fahrenheit (57° Celsius). The yellow light and display will turn off and green
light will come on indicating unit is safe to open and re
-
start cycle.
I)
Automatic shut off when cycle is complete.
J)
Unit operates at zero pressure. Cover will release at 0,15 Bar (2 PSI).
L)
Temperature Probes check each other 256 times per second to ensure the other is
working within 50 degrees Fahrenheit (10 deg
rees Celsius) of itself.
M)
Temperature Probes are linked to an Analog circuit as well as the microprocessor.
In the event the processor malfunctions, the Analog circuit will turn the unit off at
395 degrees Fahrenheit (202 degrees Celsius) until the unit
cools.
N)
Fuse Protection.
O)
Overfill protection. Float sensor automatically shuts elements down in the event
that the collection drum becomes overfull and displays Er10 .