english
operation
• To take a temperature reading, point the unit at the object you wish to measure.
Press the SCAN button (A) and read the temperature on the unit’s LCD, “SCAN” will
appear on the display while the unit is actively taking a temperature, “HOLD” when
the unit is displaying the last recorded temperature. The unit will hold the last read-
ing for 10 seconds after the button is released.
• To change between Celsius and Fahrenheit, press button (A) then press the °C/°F
button (B) until the change occurs on the display.
• The Pyropen also measures Maximum (MAX), Minimum (MIN) and Average (AVG)
temperatures each time you take a reading. This data is stored and can be recalled
until new measurements are taken.
• To recall stored values. With the unit in HOLD, press the MODE button (C). Each
time you press the MODE button, you advance through the MODE cycle.
• To select MAX, MIN and AVG, press button (A), then press the MODE button (C)
until the appropriate code appears.
• If the object you wish to measure is highly reflective, you may need to lower the
emissivity setting of the Pyropen. NOTE: Most materials have high emissivities
(paper, plastics, rubber, painted or oxidised metals) and it will not be necessary to
alter the factory-set emissivity value of 0.95.
• To change the emissivity. With the unit in HOLD, press the MODE button (C) until
the symbol is displayed. Press button (A) ( symbol flashes) then use buttons (B)
and (C) to lower or raise the emissivity. When the emissivity is correct, press button
(A) to complete the operation.
DiSpLaY
a Main temperature reading
B Scan indicator
C Hold indicator
D Celcius or Fahrenheit indicator
e Low battery indicator
F Max/Min/Average temperature
G Laser Indicator
H Emissivity indicator
MoDe DiSpLaY CoDeS
SCan
a°C/°F (B)
MoDe (C)
27mm
163mm
SCan MaX Min aVG

3
›
2
1
Distance: Sensor to object (inches)
0
0.3
8
2.8
70
5.2
133
D:S Nominal 8:1
7.7
195
0
20
500
40
1000
60
1500
Distance: Sensor to object (mm)
Spot Dia.
(mm)
Spot Dia.
(Inches)
english
• To ensure accurate aiming of the Pyropen, laser sighting is fitted. When aiming at a
target, the red dot of the laser will be approximately 1cm to the right of the centre
of the spot being measured.
• To enable laser sighting. Press button (A), then press the MODE button (C) until the
laser symbol ( ) flashes. Press button (B) to enable or disable laser sighting (dis-
play reads ON or OFF accordingly). Press button (C) to complete the operation.
NOTE: For safety reasons the laser is activated ONLY when SCAN button (A) is
pressed.
WARNING: Laser radiation may be harmful to the human eye.
Avoid direct exposure of human eyes to laser light.
Avoid indirect exposure via reflective surfaces such as glass and mirrors. Never aim
the unit at another person. Laser will not operate in ambient temperatures above
40°C
• The Pyropen can be locked into the SCAN mode without the need to continually
press button (A). To lock the unit into the SCAN mode: With the unit in HOLD mode,
press buttons (B) and (C) simultaneously until “LOCK” appears in the LCD. To
switch the Pyropen off press button (A) twice.
• The Pyropen U can store up to 100 temperature values, which can be viewed on
the Pyropen LCD or transferred to a PC. To log data to the Pyropen, ensure that the
unit is in LOCK mode, then press buttons (B) and (C) simultaneously until “dLOG”
flashes on the LCD. When “L-00” appears on the display, press and release button
(A) to store a temperature value in “Location 0”. Use buttons (B) and (C) to select
the remaining 99 locations and button (A) to store temperature values in them. To
return to SCAN mode press buttons (B) and (C) simultaneously. To switch the
Pyropen off press button (A) twice.
FieLD oF VieW
DiStanCe & Spot SiZe

4
›
english
MeaSUreMent teCHniQUeS
Locating a Hot Spot: To find a hot spot, use a scanning technique. Aim the ther-
mometer outside the area of interest, then scan across with an up and down motion
until you locate the hot spot.
Field of View: The target area must be at least as large as the spot size. In the dia-
gram above, Object 1 is the correct distance from the unit. Note the target
is slightly larger than the spot size being measured. Object 2 is too far away.
Distance & Spot Size: To ensure accurate temperature readings, the distance from
the target must be correct. The further the distance from the target,
the larger the spot size as above.
Reminders
• Measure surface temperature only. To measure internal temperatures of liquids or
food, stir contents and point unit into open container.
• Avoid thermal shock. Allow 30 minutes for unit to adjust to large changes in ambi-
ent temperature. Do not store unit below freezing.
• Close in on your target. The smaller the target, the closer you should be to it.
Refer to the D:S diagram above.
• Do not subject unit to excessive heat or cooling by holding it too close to target.
• Measuring shiny surfaces. Not recommended for use in measuring shiny or pol-
ished metal surfaces (stainless steel, aluminium etc).
MaintenanCe
Lens cleaning: Blow off loose particles using clean compressed air. Gently brush
remaining debris away with a camel hair brush. Carefully wipe the surface with a
moist cotton swab. The swab may be moistened with water or a water-based glass
cleaner. NOTE: DO NOT use solvents to clean the plastic lens.
Case Cleaning: To clean the exterior housing, use soap and water or a mild commer-
cial cleaner. Wipe with a damp sponge or soft rag.
Ce CertiFiCaton
This instrument conforms to the folowing standards:
Electomagnetic Emmisions EN55022B,
Electomagnetic Susceptibility EN61326-1,

5
›
english
CaLeXSoFt SoFtWare
SYSTEM REQUIREMENTS
• CalexSoft software is designed to run under Windows. Suitable versions are Windows
2000, Windows XP and Windows Vista.
• The recommended minimum computer specification is:
Intel Pentium processor
VGA display with 640x480 resolution
USB 2.0 port
INSTALLATION
1. Insert the disk provided
2. Installer should run automatically (if not, select the appropriate drive and double click
on Setup.exe)
3. Follow the on-screen instructions
NOTE Do not connect the PyroPen U to the PC before the software is installed
CONNECTING THE PYROPEN U TO A PC
1. Connect the USB Mini B connector on the USB cable provided to the PyroPen U.
2. Connect the USB A connector on the USB cable provided to an available USB 2.0
port
3. Double click on the CalexSoft desktop icon
NOTE Do not disconnect the USB cable before closing down the CalexSoft
program
ConFiGUration
• Setting Temperature Units
Temperature units can be set on the main temperature panel by pressing °C or °F.
• Setting Target Emissivity
To set an Emissivity value for the PyroPen U, use the main program menu to select:
[Setup] → [Emissivity]
Emissivity can then be set to a value between 0.3 and 1.0, or selected from a pre-
set list of materials by selecting [Data].
SpeCiFiCation
Temperature Range -20°C to 500°C
Accuracy ±1% or ±1°C (±2°F) whichever is greater
Repeatability ±0.5% of reading or ±0.5°C (±1°F) whichever is greater
Response Time 500mSec 95% response
Spectral Response 8 to 14µm
Emissivity adjustable 0.3 to 1.0
Temperature display 0.1°C
Ambient operating range 0 to 50°C (32 to 120°F)
Relative Humidity 10-95% RH non-condensing
Weight 60g
Dimensions 163 x 27 x 16mm
Power 2 x AAA batteries
Laser Class 2

6
›
WarrantY
Calex warrants this product to be free from defects in material and workmanship
under normal use and service for a period of one year from date of purchase, except
as hereinafter provided. This warranty extends only to the original purchase (a pur-
chase from CALEX or CALEX’S licensed distributors is an original purchase). This
warranty shall not apply to fuses or batteries. Factory calibration is warranteed for a
period of one year. The warranty shall not apply to any product which has been sub-
ject to misuse, neglect, accident or abnormal conditions of operation or storage.
Should CALEX be unable to repair or replace the product within a reasonable amount
of time, purchaser’s exclusive remedy shall be a refund of the purchase price upon
return of the product.
In the event of failure of a product covered by this warranty, CALEX will repair the
instrument when it is returned by the purchaser, freight prepaid to an authorised
Service Facility within the applicable warranty period, provided CALEX’S examination
discloses to its satisfaction that the product was defective. CALEX may, as its option,
replace the product in lieu of repair. With regard to any instrument returned within the
applicable warranty period, repairs or replacement will be made without charge and
with return freight paid by CALEX, unless the failure was caused by misuse, neglect,
accident or abnormal conditions of operation or storage, in which case repairs will be
billed at a reasonable cost. In such cases, an estimate will be submitted before work
is started, if requested.
The foregoing warranty is in lieu of all other warranties, expressed or implied, includ-
ing but not limited to any implied warranty of merchantability, fitness, or adequacy for
any particular purpose or use. CALEX shall not be liable for any special, incidental, or
consequential damages, whether in contract, tort, or otherwise.
english
Data Logging MoDe
Ensure that the PyroPen U has stored temperature data, is connected to the PC and
switched on. To enter the data logging mode, use the main program menu to select:
[View] → [Logged Temperatures]
A window will open which enables you to view, save or retrieve logged data.
Data acquisition MoDe
Ensure that the PyroPen U is connected to the PC and switched on. To enter the data
acquisition mode, use the main program menu to select:
[Setup] → [Data Acquisition Setup]
A window will open which enables you to set the sampling interval, set the number of
samples and select a storefile. To begin data acquisition, click the [Data Acquisition Start]
button in the main window.

7
›
français
FonCtionneMent
• Pour relever une température, pointer l’unité vers l’objet dont la température doit
être mesurée. Appuyer sur le bouton ‘SCAN’ (A) et lire la température sur l’affichage
à cristaux liquides de l’unité, ‘SCAN’ apparaît sur l’écran pendant que l’unité prend
la température activement, ‘HOLD’ lorsque l’unité affiche la dernière température
enregistrée. L’unité affiche le dernier relevé pendant 10 secondes après que le
bouton a été relâché.
• Pour passer de Celsius en Fahrenheit et vice versa, appuyer sur le bouton (A), puis
appuyer sur le bouton °C/°F (B) jusqu’à ce que le changement apparaisse sur
l’affichage.
• Le Pyropen mesure également les températures Maximum (MAX), Minimum (MIN)
et Moyennes (AVG) à chaque fois qu’un relevé est effectué. Cette donnée peut être
mémorisée et rappelée jusqu’à ce que de nouvelles mesures soient prises.
• Pour rappeler des valeurs mémorisées. L’unité doit être en mode ‘HOLD’ et appuy-
er sur le bouton ‘MODE’ (C). Chaque fois que le bouton ‘MODE’ est appuyé, le
cycle de ‘MODE’ progresse.
• Pour sélectionner MAX, MIN, et AVG (moyenne), appuyer sur le bouton (A), puis
appuyer sur le bouton ‘MODE’ (C) jusqu’à ce que le code approprié apparaisse.
• Si l’objet, dont la température doit être mesurée, est très réflectif, vous aurez peut-
être besoin d’abaisser le réglage d’émissivité du Pyropen. NOTE : La plupart des
matériaux possèdent une émissivité très élevée (papier, plastique, caoutchouc,
métaux peints ou oxydés) et il ne sera pas nécessaire de changer la valeur de 0,95
réglée en usine.
aFFiCHaGe
a Indication de température principale
B Indicateur Scan
C Indicateur Hold
D Indicateur Celsius ou Fahrenheit
e Indicateur de charge de la pile
F Température Max/Min/Moyenne
G Indicateur Laser
H Indicateur d’émissivité
CoDeS D’aFFiCHaGe DeS MoDeS
SCan
a°C/°F (B)
MoDe (C)
27mm
163mm
SCan MaX Min aVG

8
›
français
• Pour changer l’émissivité. L’unité doit être en mode ‘HOLD’. Appuyer sur le bouton
‘MODE’ (C) jusqu’à ce que le symbole s’affiche. Appuyer sur le bouton (A) (le
symbole clignote), puis utiliser les boutons (B) et (C) pour faire diminuer ou élever
l’émissivité. Lorsque l’émissivité est correcte, appuyer sur le bouton (A) pour com-
pléter l’opération.
• Pour assurer une visée précise du Pyropen, une visée à laser a été installée.
Lorsqu’une cible est visée, le point rouge du laser se situe à environ 1cm à droite
du milieu du point qui est mesuré.
• Pour activer la visée à laser. Appuyer sur le bouton (A), puis appuyer sur le bouton
‘MODE’ (C) jusqu’à ce que le symbole du laser ( ) clignote. Appuyer sur le bou-
ton (B) pour activer ou désactiver la visée à laser (l’affichage montre ON ou OFF).
Appuyer sur le bouton (C) pour compléter l’opération. NOTE : Pour des raisons de
sécurité, le laser n’est activé QUE lorsque le bouton ‘SCAN’ (A) est appuyé.
ATTENTION: Les radiations de laser peuvent endommager les yeux. Eviter
d’exposer les yeux à la lumière du laser. Eviter les expositions indirectes sur des
surfaces réflectives telles le verre et les miroirs. Ne jamais viser l’appareil sur une
autre personne. Le laser ne fonctionne pas dans les températures ambiantes supé-
rieures à 40°C.
• Le Pyropen peut être verrouillé sur le mode ‘SCAN’ sans avoir besoin d’appuyer sur
le bouton (A) en continu. Pour verrouiller l’unité sur le mode ‘SCAN’ : L’unité doit
être en mode ‘HOLD’ et appuyer simultanément sur les boutons (B) et (C) jusqu’à
ce que ‘LOCK’ apparaisse sur l’affichage à cristaux liquides. Pour éteindre le
Pyropen, appuyer sur le bouton (A) deux fois.
• Le Pyropen U peut emmagasiner jusqu’à 100 valeurs de température, qui peut être
regardé sur le Pyropen LCD ou transféré à un PC. Pour noter des données au
Pyropen, assurer que l’unité est dans le mode de LOCK, alors appuyer sur boutons
(B) et (C) simultanément jusqu’ à “dLOG” les éclats sur le LCD. Quand “L- 00”
apparaît sur le Pyropen LCD, le bouton de presse et relâchement (A) emmagasiner
une valeur de température dans “Location 0”. Usage boutonne (B) et (C) choisir le
rester 99 emplacements et bouton (A) emmagasiner température dans les. Se pour
retourner SCAN le mode appuyer sur boutons (B) et (C) simultanément. Pour étein-
dre le Pyropen, appuyer sur le bouton (A) deux fois.
Distance: Sensor to object (inches)
0
0.3
8
2.8
70
5.2
133
D:S Nominal 8:1
7.7
195
0
20
500
40
1000
60
1500
Distance: Sensor to object (mm)
Spot Dia.
(mm)
Spot Dia.
(Inches)
DiStanCe et taiLLe DU point
2
1
CHaMp De ViSee
teCHniQUeS De MeSUreS
Localiser un point chaud : Pour trouver un point chaud, utiliser une technique de bal-
ayage. Effectuer un balayage vertical jusqu’à ce que le point chaud soit localisé, en
commençant en dehors de la zone d’intérêt.

9
›
français
Champ de visée : La zone cible doit être au moins aussi grande que la taille du
point. Dans le schéma ci-dessus, l’objet 1 se trouve à la bonne distance de l’unité. Il
faut remarquer que la cible est légèrement plus grande que la taille du point mesuré.
L’objet 2 est trop éloigné.
Distance & taille du point : Pour s’assurer que les lectures de température sont pré-
cises, la distance entre la cible et le thermomètre doit être correcte. Plus la cible est
éloignée, plus la taille du point est grande, comme ci-dessus.
Rappels
• Ne mesure que la température de surface. Pour mesurer la température interne
de liquides ou d’aliments, mélanger le contenu et pointer le thermomètre vers un
récipient ouvert.
• Eviter les chocs thermiques. Allouer 30 minutes au thermomètre, pour qu’il
s’adapte à d’importantes fluctuations de température ambiante. Ne pas stocker
l’unité au-dessous de zéro.
• Rapprocher le thermomètre de la cible. Plus la cible est petite, plus le thermomè-
tre doit en être près. Se référer au schéma D:S ci-dessus.
• Ne pas exposer à des températures excessives en le tenant trop près de la cible
• Mesurer des surfaces brillantes. Il n’est pas recommandé de mesurer des surfac-
es brillantes ou en métal poli (acier inoxydable, aluminium, etc.).
entretien
Nettoyage de la lentille : Eliminer toutes les particules libres avec de l’air comprimé
propre. A l’aide d’un pinceau en petit gris, brosser doucement tous les débris
restant. Nettoyer délicatement la surface avec un Coton-Tige humide. Le Coton-Tige
doit être humidifié avec de l’eau ou un nettoyant à base d’eau. NOTE : NE PAS uti-
liser de solvant pour nettoyer la lentille en plastique.
Nettoyage de l’étui : Pour nettoyer l’extérieur de l’étui, utiliser du savon et de l’eau ou
un nettoyant doux. Essuyer avec une éponge humide ou un chiffon doux.
CertiFiCation Ce
Cet instrument est conforme aux normes suivantes :
Emissions électromagnétiques EN55022B,
Sensibilité électromagnétique EN61326-1,

10
›
français
SpéCiFiCationS
Echelle de température -20°C à 500°C
Précision ± 1% ou ± 1°C (±2°F), en prenant le plus élevé
Fidélité ± 0,5% de la lecture ou ± 0,5°C (±1°F), en
prenant le plus élevé
Temps de réponse 500m Sec, 95% réponse
Réponse spectrale 8 à 14µm
Emissivité réglable de 0,3 à 1,0
Effichage des températures 0.1°C
Echelle de fonctionnement ambiante 0°C à 50°C (32 à 120°F)
Humidité relative 10-95% HR sans condensation
Poids 60g
Dimensions 163 x 27 x 16mm
Alimentation 2 piles AAA
Laser Classe 2
CaLeXSoFt SoFtWare
SPÉCIFICATIONS DU SYSTÈME.
• Le logiciel CalexSoft est conçu pour fonctionner sur Microsoft Windows. Les ver-
sions appropriées sont Windows 2000, Windows XP et Windows Vista
• Les spécifications informatiques minimales recommandées sont :
Processeur Intel Pentium
Ecran VGA avec une résolution 640x480
USB 2.0 port
INSTALLATION
1. Insérer le dique fourni
2. Le programme d’installation devrait se mettre en marche automatiquement. (Si ce
n’set pas le cas, sélectionner le lecteur approprié et cliquer sur Setup.exe)
3. Suivre les consignes affichées
NOTA Le PyroPen U ne doit pas être connecté au PC avant l’installation du logiciel
CONNECTION DU PYROPEN U À UN PC
1. Relier le connecteur USB MINI B sur le cable USB fourni au PyroPen U
2. Insérer le connecteur USB A sur le cable USB fourni dans un port USB 2.0 disponible.
3. Double-cliquer sur l’icône CalexSoft
NOTA Il ne faut pas déconnecter le cable USB avant de fermer le logiciel
CalexSoft
ConFiGUration
• Sélection de l’unité de temperature.
On sélectionne l’unité de temperature en appuyant sur le bouton °C ou °F sur le
console de temperature principal.
• Réglage de l’émissivité de la cible
Pour fixer une valeur d’émissivité pour le PyroPen U, aller au menu principal du
programme et sélectionner:
[Setup] → [Emissivity]
Il est alors possible de fixer une valeur d’émissivité entre 0,3 et 1,0, ou bien, une
valeur peur être sélectionnée d’une liste de materiel préfixée en sélectionnant [Data].