1) SAFETY
This manual contains information and warnings that must be followed for operating the
meter safely and maintaining the meter in a safe operating condition. If the meter is
used in a manner not specified by the manufacturer, the protection provided by the
meter may be impaired.
Observe proper safety precautions when working with voltages above 30 Vrms, 42.4
Vpeak or 60 VDC. These voltage levels pose a potential shock hazard to the user. Do
not expose this product to rain or moisture. The meter is intended only for indoor use.
Keep your hands/fingers behind the hand/finger barriers (of the meter and the test
probe assembly, where applicable) that indicate the limits of safe access of the
hand-held parts during measurements. Inspect lead wires, connectors, and probes for
damaged insulation or exposed metal periodically. If any defects are found, replace
them immediately. Only use the test probe assembly provided with the meter or a UL
Listed test probe assembly to the same meter ratings or better.
Optional offer premium test probe assembly using silicone lead wire insulation, at
agent’s discretion, is equipped with white inner insulation layers as wear indicators.
Replace them immediately if any of the white layers has become visible.
Disconnect the test leads from the test points before changing functions.
The meter meets IEC/EN/BSEN/CSA_C22.2_No./UL standards of 61010-1 Ed. 3.0,
61010-2-033 Ed. 1.0, 61010-2-033 Ed. 1.0 to Measurement Categories CAT III 1000V
and CAT IV 600V ac & dc.
The accompanied test probe assembly meets IEC/EN/BSEN/CSA_C22.2_No./UL
standards of 61010-031 Ed. 2.0 to the same meter ratings or better. The 61010-031
requires exposed conductive test probe tips to be ≤ 4mm for CAT III & CAT IV ratings.
Refer to the category markings on your probe assemblies as well as on the add-on
accessories (like detachable Caps or Alligator Clips), if any, for applicable rating
changes.

2
INTERNATIONAL SYMBOLS
Marking of Electrical and Electronic Equipment (EEE). Do not dispose of this
product as unsorted municipal waste. Contact a qualified recycler
Refer to the explanation in this Manual
Possibility of electric shock
Earth (Ground)
Meter protected throughout by Double Insulation or Reinforced insulation
Fuse
Direct Current (DC)
Alternating Current (AC)
3 Three-phase Alternating Current
Application around and removal from hazardous live conductors is permitted
BRIEF INFORMATION ON MEASUREMENT CATEGORIES
Measurement Category IV is applicable to test and measuring circuits connected at
the source of the building’s low-voltage MAINS installation. Examples are
measurements on devices installed before the main fuse or circuit breaker in the
building installation.
Measurement Category III is applicable to test and measuring circuits connected to
the distribution part of the building’s low-voltage MAINS installation. Examples are
measurements on distribution boards (including secondary meters), circuit-breakers,
cables, bus-bars, junction boxes, switches, socket-outlets, stationary motors in the
fixed installation, and equipment for industrial use.
Measurement Category II is applicable to test and measuring circuits connected
directly to utilization points (socket outlets and similar points) of the low-voltage MAINS
installation. Examples are measurements on MAINS CIRCUITS of household
appliances, portable tools and similar equipment.
2) EUROPEAN DIRECTIVES AND UK STATUTORY REQUIREMENTS
The instruments conform to EUROPEAN (CE) Low-Voltage Directive 2014/35/EU,
Electromagnetic Compatibility Directive 2014/30/EU, and RoHS 2 Directive 2011/65/EU
plus amendment Directive (EU) 2015/863. The instruments also conform to the UK
(UKCA) Electrical Equipment (Safety) Regulations 2016, Electromagnetic Compatibility
Regulations 2016, and The Restriction of the Use of Certain Hazardous Substances in
Electrical and Electronic Equipment Regulations 2012.

3
3) PRODUCT DESCRIPTION
Panel Illustration
4) OPERATION
NOTE
Before and after hazardous voltage measurements, test the voltage function on a
known source such as line voltage to determine proper meter functioning.
1) 5-4/5 digits 500000
counts LCD display
2) Push-buttons for special
functions & features
3) Selector to turn the
Power On or Off and
Select a function
4) Input Jack for 10A (+)
(20A for 30sec) current,
and for T2 (-) function
5) Input Jack (+) for all
functions EXCEPT current
(
A, mA, A) and T2
functions
6) Common (Ground
reference) Input Jack (-)
for all functions EXCEPT
T2 function
7) Input Jack (+) for
milli-amp, micro-amp, and
T2 (+) functions

4
AC Voltage, DC Voltage, DC+AC Voltage, & Hz Line Level Frequency
In AC Voltage, press SELECT button momentarily to toggle between AC and dBm. In
DC Voltage, press SELECT button momentarily to toggle between DC, and DC+AC. In
mV Voltage, press SELECT button momentarily to select DC, AC, or DC+AC. The new
settings will be saved automatically to the non-volatile memory as power up default. In
DCV and DCmV, press 500000 button momentarily to toggle between 4-4/5 digits and
5-4/5 digits readings. In voltage or current functions, press the Hz push button
momentarily to activate or to exit Line Level Frequency measuring function. Line Level
Frequency measuring function is designed especially for noisy electrical high voltage
signals.
Note: In dBm function, power up default reference impedance will be displayed for 1
second before displaying the dBm readings. Press dBm- (RANGE) button
momentary to select different reference impedance of 4, 8, 16, 32, 50, 75, 93, 110,
125, 135, 150, 200, 250, 300, 500, 600, 800, 900, 1000, up to 1200. The new
impedance value will be saved automatically to the non-volatile memory as power up
default.

5
Note: Line Level Frequency measuring function input sensitivity varies automatically
with voltage (or current) function range selected. The lower the measuring range the
higher the sensitivity. That is, mV function has the highest and the 1000V range has
the lowest as in voltage function ranges. It is recommended to first measure the signal
voltage (or current) level then activate the Hz function in that voltage (or current) range
to automatically get the most appropriate trigger level. When activated from voltage
function, you can also press the RANGE button momentarily to select another trigger
level range manually. The analog bargraph pointer will point at the selected trigger
level range scale 1, 2, 3, or 4. If the Hz reading is unstable, select lower sensitivity to
avoid electrical noise. If the reading shows zero, select higher sensitivity.
Hz Logic Level Frequency and % Duty Cycle functions
Press SELECT button momentarily to toggle between Hz and % (duty cycle) readings.
The new setting will be saved automatically to the non-volatile memory as power up
default. Press 500000 button momentarily to toggle between 5 full digits and 6 full digits
Hz readings.

6
Note: Unlike the Line Level Frequency measuring function as previously stated, this
Logic Level Frequency function is set only at the highest input sensitivity for measuring
digital type electronic signals.
T1-T2 Dual Channels Temperature function (Model 859s only)
Press SELECT button momentarily to toggle between C and F readings, and the
new setting will be saved automatically in the non-volatile memory as power up default.
Press T1-T2 (RANGE) button momentarily to select T1, T2, or T1-T2 readings.
Note: Insert the banana plug K-type temperature bead probe Bkp60 (standard
accessory x 1) with correct polarities. Dual channels T1-T2 readings require 2
probes. You can also use a plug adapter Bkb32 (Optional purchase) with banana pins
to K-type socket to adapt other standard K type mini plug temperature probes.

7
Resistance, Continuity functions
Press SELECT button momentarily to toggle between and Continuity functions.
The new setting will be saved automatically to the non-volatile memory as power up
default. Continuity function is convenient for checking wiring connections and operation
of switches. A continuous beep tone indicates a complete wire.

8
NOTE
Using resistance or continuity function in a live circuit will produce false results and
may damage the meter. In many cases the suspected component must be
disconnected from the circuit to obtain an accurate reading.
Capacitance, Diode test function
Press SELECT button momentarily to toggle between Capacitance and
Diode test functions. The new setting will be saved automatically to the non-volatile
memory as power up default.
NOTE
Discharge capacitors before making any measurement. Large value capacitors should
be discharged through an appropriate resistance load.
Note: Normal forward voltage drop (forward biased) for a good silicon diode is between
0.400V to 0.900V. A reading higher than that indicates a leaky diode (defective). A zero
reading indicates a shorted diode (defective). An OL indicates an open diode
(defective). Reverse the test leads connections (reverse biased) across the diode. The
digital display shows OL if the diode is good. Any other readings indicate the diode is
resistive or shorted (defective).

9
A, mA, A, and %4-20mA Current functions
Insert the red test lead into the correct A/mA or A input jack. Press SELECT button
momentarily to select DC, AC, or DC+AC. The new settings will be saved automatically
to the non-volatile memory as power up default. In DC mA function, neither in AC nor in
DC+AC, press and hold the %4-20mA ( Hz) button for 1 second or more to display
the current digital data in terms of loop current percentage (%) value. It is set at 4mA =
0% (zero) and 20mA = 100% (span) with 0.01% high resolution, which virtually extends
the meters’ capability to test and regulate the externally powered loop current in the
industrial process control applications. The analog bar-graph remains showing the mA
current value to alert the user.
NOTE: When measuring a 3-phase system, special attention should be taken to the
phase-to-phase voltage that is significantly higher than the phase-to-earth voltage. To
avoid exceeding the voltage rating of the protection fuse(s) accidentally, always
consider the phase-to-phase voltage as the working voltage for the protection fuse(s).
Ce manuel convient aux modèles suivants
1
Table des matières
Autres manuels Brymen Instrument de mesure
Manuels Instrument de mesure populaires d'autres marques

Endress+Hauser
Endress+Hauser Proline Promag 50 Caractéristiques techniques

Siemens
Siemens SITRANS F Coriolis FCT030 Manuel de la liste des pièces

KLINGER
KLINGER CMF V Series Manuel utilisateur

EXFO
EXFO FTB-2 Manuel d'exploitation et d'entretien

Keysight
Keysight M8290A Manuel utilisateur

ADTEK
ADTEK MW-5 Manuel utilisateur