7
5.8. If the operation time is set then the unit will stop after the set time interval elapses.
Flashing indication STOP (fig. 3/8) is shown on the display and a repeating sound
signal for the end of operation is heard. Press the Run Stop key to stop the signal.
5.9. Press the Run Stop key to repeat the preset program.
5.10. The unit can be stopped at any time during operation before the set time elapses if
necessary by pressing the Run Stop key. Pressing the Run Stop key again will start
the program from the beginning (timer will be restarted).
5.11. After finishing the operation, turn off the power switch.
5.12. Unplug the external power supply from the mains.
6. Program setting
6.1. The program consists of cycles. Each cycle includes motions of three different types
(rotational, reciprocating and vibro) set one after another with the duration from 0 to
250 seconds for rotational and reciprocal motion types and from 0 to 5 seconds for
vibro motion.
6.2. Changing parameters. Press the Select key (fig. 3/9) to choose the parameter to
change, each consecutive press activates the next parameter in cycle. Active
parameter is flashing. Use the ▲ and ▼ keys (fig. 3/10) to set the necessary value.
Pressing the key for more than 2 s increases value changing speed.
6.3. All parameters except overall timer can also be changed during the operation. Unit
automatically saves and applies changes to the current work program.
6.4. Following parameters are available: rotation and reciprocation speed, reciprocation
and vibro turning angle, time for each motion type and overall operation timer.
6.5. Skip a motion. If the time for a type of motion is set to zero (indication OFF), this
type of motion will be skipped in the cycle.
6.6. Pause. If the turning angle for reciprocating or vibro motions is set to zero (indication
0), unit will interpret this as a pause. The stirring element will not move in this mode
during the operation and the overall timer will be counted down.
6.7. The overall timer (fig. 3/7) is used to control the operation time. The timer can be set
for the period from 1 min to 96 hours (increment 1 min). If the timer is set to 00:00,
unit operates non-stop, see 5.7.
6.8. Table 1 shows possible motion combinations in the cycle.
Table 1. Motion type combinations
6.9. Further examples illustrate program setting for both separate motions and their
combinations in the cycle.