1. SPECIFICATIONS
SPECIFICATIONS - NAMES OF PARTS AND FUNCTIONS
2. NAMES OF PARTS AND FUNCTIONS
●LED indicators
Blinking LED( ):Inclination exceeding 0.08mm/1m
Continuous LED( ):Inclination of 0.08mm/1m or less
Continuous LED( ):Inclination of 0.06mm/1m or less
Continuous LED( ):Inclination of 0.04mm or less
Continuous LED( ):Inclination of 0.02mm/1m or less
Continuous LED( ):Inclination of 0.01mm/1m or less
mode
HIGH
mode
Blinking LED( ):Inclination exceeding 0.8mm/1m
Continuous LED( ): Inclination of 0.8mm/1m or less
Continuous LED( ): Inclination of 0.6mm/1m or less
Continuous LED( ):Inclination of 0.4mm/1m or less
Continuous LED( ):Inclination of 0.2mm/1m or less
Continuous LED( ):Inclination of 0.1mm/1m or less
LOW
Specifications
Model
Minimum readable value
Power supply
Auto-power OFF
Operating temperature
Battery life
Dimensions
Weight
LVM-01
0.01mm/1m
Alkaline dry battery: LR03 (AAA) x4
The power turns OFF automatically after 30 minutes.
0°C
-
40°C (Recommended temperature 20°C ±5°C)
50 continuous hours
ø109 x H56
995g
Included items Alkaline dry battery: LR03 (AAA) x4,
Operation manual and inspection report card
①POWER switch ON OFF
Push the POWER switch to turn ON the power
and the LED level indicators. It is now possible
to execute the levelling operation.
②MODE switch HIGH LOW
Push the MODE switch to change the precision
level. The "HIGH" mode is selected when turning
ON the device.
③SET switch
This switch is used to execute the zeroadjustment.
⑥LED level indicator (Central position)
The LED (blue) turns on when the level status
reaches the central position.
The inclination is 0.1mm/1m or less in case of the
LOW mode.
The inclination is 0.01mm/1m or less in case of the
HIGH mode.
⑦Mode LED
The LED (red) turns on when the LOW mode
is selected.
⑧
Mode LED
The LED (green) turns on when the
HIGH
mode
is selected.
⑨Battery alarm LED
If the voltage of the batteries decreases to 4V or
less, the LED(orange) blinks notifying the necessity
of replacing the batteries.
⑩Buzzer
The buzzer emits a sound and the LED (blue)
described on point ⑥turns on when the level
status reaches the central position.
⑬Line markers
They are used for determining the direction of the
X and Y axes.
⑮Batteries box
It is the location where the batteries are installed
(LR03: AAA x4).
⑪Base
⑫Body case
④LED level indicators (X-axis)
They indicate the level status.
The inclination is between 0.8mm/1m and
0.1mm/1m in case of the LOW mode.
The inclination is between 0.08mm/1m and
0.01mm/1m in case of the HIGH mode.
⑤LED level indicators (Y-axis)
They indicate the level status.
The inclination is between 0.8mm/1m and
0.1mm/1m in case of the LOW mode.
The inclination is between 0.08mm/1m
and 0.01mm/1m in case of the HIGH mode.
3. HOW TO INSTALL THE BATTERIES
②Insert the 4 batteries included (LR03: AAA) in the base of the
batteries box as indicated on the figure on the right.
③After have installed the batteries, insert the projecting part of the
cover into the recess of the batteries box. Then, secure the cover
by tightening the screw.
①Loose the screw of the cover of the batteries box.
Batteries box
Batteries
LR03 (AAA) x4
Cover
Projecting part of the cover Screw
Always check the polarity of the batteries. If the batteries
are inserted with the polarity inverted, the device may not
work and the internal electric circuit may get damaged
making the device not utilizable.
Caution
⑭Buzzer sound ON OFF switch
This switch is used for turning on/off the buzzer.
①②
③
④⑤
⑥⑦ ⑧
⑨
⑩
⑪
⑫
⑬⑬
⑭ ⑮
11 12
HOW TO INSTALL THE BATTERIES
This product complies with the EMC standard EN 61326-1.