BGB 6 Manuel utilisateur

BGB
Electronic Crimping and Decapping Tools
Model 6
PN
297040, 297041, 297042, 297043, 297044, 297045, 297046, 297047,
297048
Operation Guide
© BGB Analytik AG
Manual Part Number: DR-BGB-1908
July 2019

I
Maintenance and Cleaning
General maintenance
The electronic crimper tools do not contain user serviceable parts except for the battery pack. When
cleaning or replacing the battery make sure to keep fingers away from the jaws!
Support and Repair
If the crimping tool is still in the warranty period, contact BGB Analytik office for support. If the
warranty period has expired, please contact BGB Analytik. Find contact details on
www.bgb-shop.com.
Cleaning
The crimping tool may not be immersed in water or solvent. The outside of the case may be cleaned
with an ordinary detergent and wiped off with a damp rag. Care should be taken not to get the
electronics, the battery, or the battery connections wet.
Avoid permitting metal parts of the crimping tool to come into contact with corrosive material during
use. If they do, try to wipe them clean with a suitable mild neutralizing solution.
Storage and Shipping
To prevent accidental cycling when storing or shipping the tool, remove power and place the
protective cap over the jaws.
Safety Notices
A CAUTION notice warns of a hazard. It
calls attention to an operating procedure,
practice, or the like that, if not correctly
performed or adhered to, could result in
damage to the product or loss of important
data. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully
understood and met.
Manual Part Number
DR-BGB-1908
Edition
First edition, July 2019
www.bgb-shop.com
A WARNING notice warns of a hazard. It
calls attention to an operating procedure,
practice, or the like that, if not correctly
performed or adhered to, could result in
personal injury or death. Do not proceed
beyond a WARNING notice until the
indicated conditions are fully understood
and met.
Recycling
For recycling contact
your local BGB distributor.

II
Battery Replacement
Use only the specified replacement battery pack, part number 295081.
1. Remove screw holding the
battery cover in place.
2. Remove battery cover.
3. Pull battery out of case,
leaving wire connected.
4. Push down on connector
latch and pull battery loose
from the board.
5. Connect new battery to board,
making sure the latch is secure.
Push battery into case, bending
the wires if necessary.
6. Slide the cover back into the
case and replace the retaining
screw.

III
Table of Contents
Safety Notices I
Maintenance and Cleaning I
Support and Repair I
Storage and Shipping I
Battery Replacement II
Product Content 4
Safety and Regulatory Certifications 5
Symbols 5
Instrument Specification 6
Sound pressure 6
Intended use 6
Limits 6
Warnings 7
Warnings 7
Special battery warnings 7
Disposal of battery 7
Description and Setup 8
Description 8
Crimping tool setup 8
Operation 9
Charging the Battery 9
Compatible Vials, Caps and Seals 9
Settings 10
Language selection 10
Statistics 10
Log of recent crimping results 10
Using the Tool - Crimping Version 11
Using the Tool - Decapping Version 12
Reset 12
Error Conditions and Display Codes 13
Troubleshooting 14

4
Product Content
Part No. Description
297040 8 mm Crimper Model 6
297041 11 mm Crimper Model 6
297042 13 mm Crimper Model 6
297043 13 mm Crimper Model 6 for Flip Off Caps
297044 20 mm Crimper Model 6
297045 20 mm Crimper Model 6 for Flip Off Caps
297046 11 mm Decapper Model 6
297047 13 mm Decapper Model 6
297048 20 mm Decapper Model 6

5
Safety and Regulatory Certifications
The BGB Electronic Crimping Tools are designed and manufactured under a quality system registered
to ISO 9001.
Symbols
Warnings in the manual or on the instrument must be observed during all phases of operation, service
and repair of this instrument. Failure to comply with these precautions violates safety standards of
design and the intended use of the instrument. BGB assumes no liability for the customer‘s failure to
comply with these requirements.
See accompanying instructions for more information.
Remember to wear safety glasses when crimping or decapping.
The crimper or decapper jaws can pinch severely.
Never insert fingers into the crimping or decapping tool jaws.
You must not discard this electrical/electronic product
in domestic household waste.

6
Instrument Specification
Sound pressure
Sound pressure LpA = 70.1 dB(A).
Intended use
Electronic Crimping Tools are intended for use in a laboratory environment, all other uses are
prohibited.
Limits
Temperature 15 - 35 °C
Maximum humidity 75%
Pressure 0.75 - 1 bar

7
Warnings
Warnings
Wear safety glasses when crimping or decapping.
The crimper or decapper jaws can pinch severely.
Never insert fingers into the crimper or decapper.
Use only the 7.5 volt DC Power Supply supplied with the crimper for battery charging.
Use only the specified replacement battery pack, part number ????.
Special battery warnings
Risk of burns; battery may explode or catch fire if mishandled.
Do not disassemble or dispose of in fire.
Use only the manufacturer-supplied 7.5 volt DC power supply and charge the battery only in the crim-
ping tool.
Use of other batteries may cause fire during charging or use.
Do not heat above 60 °C.
Do not crush of modify.
Disposal of battery
Do not throw the battery away. Recycle it in accordance with local regulations.

8
Description and Setup
Description
The Electronic Crimping Tool can be used to crimp and decap standard crimp caps on laboratory sam-
ple vials.
Crimping tool setup
Please read through this entire guide to familiarize yourself with the operation of the tool before pro-
ceeding. Use the same degree of care as you would with any precision instrument.
If there is any visible damage, contact BGB Analytik immediately.
②Power connector
Go button ①

9
Operation
Charging the Battery
The battery must be charged before the crimper or decapper can be used.
An indicator appears when recharging is necessary.
Compatible Vials, Caps and Seals
BGB electronic crimping tools may not be used with Steel Caps or Bi-Metal Caps. Aluminum caps and
seals of standard size and thickness are appropriate.
Autres manuels pour 6
1
Ce manuel convient aux modèles suivants
9
Table des matières
Autres manuels BGB Outils électriques