Benchmark H5000-H Manuel utilisateur

MultiTherm Shaker
Operations Manual
Item: H5000-H and H5000-HC

- 1 -
Foreword
Thank you for purchasing a MultiTherm™ Shaker. This operations manual
contains instructions for the care of this Instrument. In order to use the
instrument properly, please read this manual carefully before operation.
Please keep it available for future use.
Prior to the initial operation:
Please check the Instrument and the accessories against the packing list
when you first open the shipping carton. If you find there is damage, a
discrepancy, or missing items, please contact the supplier immediately.

- 2 -
Safety Warnings and Guidelines
1. Important operation information:
Please read this manual carefully prior to any attempt to operate the Instrument.
2. Security:
The operation, maintenance and repair of the Instrument must comply with the
basic guidelines and warnings stated below. Failure to comply may jeopardize the
instrument life, warranty coverage and user safety.
Read the manual prior to power up of the instrument. Read the
guidelines and directions below and be familiar with all safety
warnin
g
s and
g
uidelines.
Use this product indoors. Do not operate in a damp or wet
environment.
Do not attempt to disassemble or repair the Instrument by
yourself. Disassembly will void the warranty. If there is a
malfunction contact your supplier for repair service.
The three pin plug supplied with the instrument is a safety
device that should be matched with a suitable grounded outlet.
Do not attempt to defeat the three pin plug or its grounded
function.
During normal operation, the temperature of metal block will
become very hot. Do not touch the block during or within 30
minutes of operation.
Close all test tube lids before inserting into the block. Do not
allow liquids to spill into the block or any part of the instrument.

- 3 -
3. Maintenance of the Instrument
The block wells should be cleaned by a cloth dampened with alcohol to ensure
good heat transmission between the block and the test tube. If there are marks, dirt
or stains on the instrument exterior, clean them with a cloth dampened with water
or a mild detergent solution.
Before connectivity to a power source, check that the voltage
listed on the serial number label (on the bottom or back side of
the unit) to ensure that it matches the voltage available in your
area.
Be sure to install the Instrument in a cool, dry location away
from any direct sunlight or strong light sources. Additionally
the instrument should be kept away from corrosive gas,
magnetic fields and dusty environments. Keep the instrument
and the air vents at least 4 inches from walls, other instruments
or any obstructions.
Shut the power off when you finish your work. Unplug the
machine if it will not be operated for a long time period and cover
it with a cloth or plastic paper for protection from dust particles
Immediately unplug the instrument from the outlet and
contact the vendor in the event of:
Fire or smoke.
Liquid has gotten into the Instrument;
Abnormal operation: such as abnormal sound or smell.
Instrument has been dropped or outer case is damaged.
The function has visibly changed.
Always shut the power off when cleaning the Instrument.
When cleaning the block wells, do not apply the cleaning liquid
directly to the block, use a dampened cloth. Do not use strong or
corrosive cleaning agents.
The Main power switch is on the rear of the device,
push “I” to power on the instrument, and push “O” to
power off.

- 4 -
Contents
Chapter1:Introduction 5
Chapter2:Specifications 5
1. Normal operating conditions 5
2.Technicaldata 6
Chapter 3: Preparations 8
1. Physical Description 8
2. Keyboard and Display panel 9
3. Key Functions 10
Chapter4:OperationGuide 11
1. Single Temperature, Speed and Time setting 11
2. Sequence Link Setting 12
3. Disabling the Temperature Speed or Time 12
4.Pulse 13
5. Temperature Calibration 13
6.ChangingtheBlock 16
Chapter 5: Failure analysis and troubleshooting 17

H5000 Series MultiTherm Shakers
5
Chapter 1: Introduction
The MultiTherm shaker is an ideal instrument for intensive mixing of samples under
temperature regulated conditions.
Mixing and temperature control modes can be used both simultaneously and
independently i.e, the device can work as shaker and/or as an incubator. The base
instrument can be used with a variety of different blocks. The instrument is applicable
for DNA analysis, extraction of lipids and other cell components, DNA library creation,
PCR amplification, pre-denaturation in electrophoresis, serum solidification etc.
Features of this product as follows:
1. Because it can be used with various optional mixing blocks, the instrument can
adapt to different tubes and plates to accommodate many experimental needs.
Customized blocks are available to suit your specific demands.
2. LCD display
3. Simultaneous display of set and actual time, temperature and speed
4. Over-heat protection system ensures safety & reliability
5. Temperature can be calibrated to meet the user's needs
6. End-of-run signal alerts the user of cycle completion
Chapter 2: Specifications
1. Normal operating conditions
Ambient temperature: 5C ~30C / Relative humidity: ≤70%
Note: Do not attempt to operate the instrument at temperatures or humidity levels
outside of this range.

H5000 Series MultiTherm Shakers
6
2. Technical data, H5000-H
Model
Parameter
H5000-H
Mixing rate* 200~1500 rpm
Orbit 2mm
Temperature setting range 5°C to 100°C
Temperature control range* Ambient +5°C to 100°C
Timing range 1min to 99h59min
Accuracy of the temperature ≤0.5°C
Heating time ≤15min (from 20°C to 100°C)
Cooling time N/A
Block Configuration
A-BLOCK: 96 well Plate or 96 x 0.2ml;
B-BLOCK: 0.5ml x 54 C-BLOCK: 1.5ml x 35
D-BLOCK: 2.0ml x 35 E-BLOCK: 15ml x 12
F-BLOCK: 12mm x 24 G-BLOCK: 50ml x 6
H-BLOCK: 0.5ml x 20 & 0.5ml x15
Power Supply
AC100-120~2.5A
50-60Hz
AC200-240~1.5A
50-60Hz
Fuse 250V 4.0A Ф5×20mm 250V 3.0A Ф5×20mm
Dimension (mm) 300(D)×2 12(W)×210(H)
Net weight (kg) 8.5
*Specification is valid when used with “standard” blocks only. Larger blocks (Blocks E, F and G)
have different speed and temperature specifications.

H5000 Series MultiTherm Shakers
7
Technical data, H5000-HC
Model
Parameter
H5000-HC
Mixing rate* 200~1500 rpm
Orbit 2mm
Temperature setting range 0° to 100°C
Temperature control range*
0° to 100° at Room temp. ≤20°C
4° to 100° at Room temp.≤25°C
10° to 100° at Room temp.≤30°C
Timing range 1min ~ 99h59min
Accuracy of the temperature ≤0.5°C
Heating time ≤15min (from 20°C to 100°C)
Cooling time ≤30min from room temp. to -20°C below room temp.
Block Configuration
A-BLOCK: 96 well Plate or 96 x 0.2ml;
B-BLOCK: 0.5ml x 54 C-BLOCK: 1.5ml x 35
D-BLOCK: 2.0ml x 35 E-BLOCK: 15ml x 12
F-BLOCK: 12mm x 24 G-BLOCK: 50ml x 6
H-BLOCK: 0.5ml x 20 & 0.5ml x15
Power Supply
AC100-120~2.5A
50-60Hz
AC200-240~1.5A
50-60Hz
Fuse 250V 4.0A Ф5×20mm 250V 3.0A Ф5×20mm
Dimension (mm) 300(D)×2 12(W)×210(H)
Net weight (kg) 8.5
*Specification is valid when used with “standard” blocks only. Larger blocks (Blocks E, F and G)
have different speed and temperature specifications.

H5000 Series MultiTherm Shakers
─ 8─
Chapter 3: Preparations
This chapter provides a physical description, location of controls, function of
individual keys, as well as some preparations before the initial power-on. You
should be familiar with this chapter before the Thermo Shaker is first operated.
1. Physical Description
Standard Block
Keyboard
Power Switch
Power connector
Fuse

H5000 Series MultiTherm Shakers
─ 9─
2. Keyboard and Display panel
Display Panel:
P1100.0120010:30
P1100.0 120015:00
Current temperature Current speed Remaining time
Selected tem
p
erature Selected s
p
eed Selected time
Current program
Selected
p
ro
g
ram