PRODUCT DESCRIPTION
The P
PP
POSI
OSIOSI
OSIFLOW
Sensor will verify that your
metering pump is truly moving liquid into the
system and has not lost prime. Utilizing
proprietary circuitry coupled with advanced
pressure sensing technology, the P
PP
POSI
OSIOSI
OSIFLOW
Sensor provides the most accurate and non-
intrusive flow verification device to date.
Without any moving parts in the sensor, the
standard flow path of the pump is preserved
and the liquid does not move through any gears
or lift a magnet to actuate a pulse in the sensor.
This eliminates any concern of plugging,
leaking or jamming of the flow path or sensing
mechanism.
The output signal of the P
PP
POSI
OSIOSI
OSIFLOW Sensor is
an open-collector contact closure
corresponding with each flow pulse.
Additionally, a visual green LED is mounted
on the topside of the sensor. This is always
ON to indicate power and will briefly flash
OFF in sync with each output signal.
The P
PP
POSI
OSIOSI
OSIFLOW will send out a signal with every stroke as long as there is liquid pumping against
pressure. If the pump becomes either air-locked or in a dead head condition, the sensor will no longer
output a signal. And unlike the competition, if the discharge line breaks or becomes disconnected,
pressure will be lost and the P
PP
POSI
OSIOSI
OSIFLOW sensor will stop signaling flow.

W
A L C H E M
SAFETY & CAUTION NOTES
Wear Protection
When working on or around a metering pump,
always wear proper protective clothing and
equipment as recommended by the supplier of the
chemical being pumped.
Remove Power
Disconnect the pump and sensor from their electrical
power sources before performing any maintenance.
If the pump starts operation during maintenance,
chemical may be sprayed & cause injury.
Application of Product
Use the PosiFlow within the specified range.
Using the PosiFlow in any other application than its
intended purpose may result in personal injury or
damage to the product.
Power Source
Use only stable DC voltage within the specified range.
Voltage outside the range may cause damage or fire.
Do Not Modify the Product
Never attempt to modify the PosiFlow. Alteration of
the product may produce a dangerous situation and
will void the warranty.
Location
Do not use the PosiFlow Sensor in an aggressive
environment. Exposure to liquid or excess humidity
may cause failure or electrical shock.
Do Not Use a Damaged Sensor
Using a damaged PosiFlow may cause chemical
leakage or electrical shock.
Not observing precautions may cause injury or damage to the product.
SPECIFICATIONS
1. Identification:
Part Number Description Material
FCP-1VC PosiFlow Sensor, VC VC
FCP-1VE PosiFlow Sensor, VE VE
FCP-1PC PosiFlow Sensor, PC PC
FCP-1PE PosiFlow Sensor, PE PE
2. Applicable Models and Ordering Information:
The P
PP
POSI
OSIOSI
OSIFLOW Sensor is available for the 10, 15 & 20 sizes of EH and EZ
pumps with matching liquid end codes. It will mount into either the manual air
vent valve or the Multi
MultiMulti
MultiFunction valve, but at this time it is not compatible with
auto air vent valve (VCA or VCC codes).
3. Electrical
Power Source Voltage 12VDC +2VDC
Current 25mA max (@12VDC)
Output Type Non-Isol. Open Collector (NPN)
Rating 24VDC max (from source)
Wave Form 100mS +15mS
Indicator LED (Green)
Cable
Length 9.75 Feet
Cross Section 3.1x10-4 in2
Termination #6 Fork Terminal
4. Materials of Construction
Housing PVC or GFRPP
O-Rings FKM or EPDM
Sensor 96% Ceramic (Al2O3)
5. Output Pressure Range*: 30+5PSI to 150+15PSI
6. Environment:
Ambient Temperature 0-120°F
Ambient Humidity 35-90% RH
Storage Temperature 32-120°F
Liquid Temperature 32-105°F (PVC models)
32-140°F (GFRPP models)
*The P
PP
POSI
OSIOSI
OSIFLOW Sensor requires a minimum of 30PSI
total back pressure. If this is not available, the injection
valve supplied with the pump can be retrofitted with an
optional high-pressure spring, part number E90375.
This spring will increase the back pressure at the pump
approximately 50PSI.
3.82” 1.89”
1.89”

W
A L C H E
M
INSTALLATION
Disconnect the pump from the electrical power source before performing maintenance.
Release the pressure in the discharge tubing before disconnecting or performing any maintenance on the pump.
Plumbing
1. Unscrew and remove the air vent knob from the manual air vent valve (or Multi
MultiMulti
MultiFunction Valve) by turning counter-
clockwise.
2. Insert the PosiFlow Sensor directly where the air vent knob was located (turn clockwise to secure).
CAUTION: The wires on the sensor will want to twist as it is being screwed into the
valve. While installing the sensor, also turn the cable to avoid wire twisting and
damage. If this is not possible, pre-twist the wire in the opposite direction about 5
turns to relieve stress during sensor installation.
3. The function of the manual air vent is not lost when using the PosiFlow Sensor.
Simply use the sensor itself as the manual air vent valve when necessary, turning it
one-half to one turn counter clockwise to bleed out air, prime the pump or release the
discharge pressure.
Electrical
Wiring
The PosiFlow Sensor requires 12VDC power to operate. Attach the RED wire to +12VC and the BLACK wire to GND
(common). The output signal is a closed contact between the WHITE and BLACK wires.
CAUTION: Do not run the PosiFlow cable near other power lines carrying high current or high power equipment as line
noise can cause sensor damage or abnormal operation.
Output Signal
The output signal (white wire) is normally high and drops to zero when the PosiFlow senses a pulse from the discharge of
the pump. The time of the closed contact is 100mS. See the output wave form below.
This type of signal acts similar to a contact closure from a reed switch device between the signal (white) and the common
(black) wires. The only difference is that the receiving device will see a trigger on the rise of the output instead of the fall
(reverse of typical).
Ω
Installing the PosiFlow Sensor into the manual air vent valve.
Installing the PosiFlow
Sensor into the
Multi
MultiMulti
MultiFunction Valve
PosiFlow Outpu
t

W
A L C H E M
Member of the Walchem Group
Five Boynton Road Hopping Brook Park Holliston, MA 01746 USA TEL 508-429-1110 FAX 508-429-7433
WEBSITE: www.walchem.com
OPERATION
Back Pressure Requirement
In order for the PosiFlow Sensor to function, a minimum of 30PSI system pressure is required at the sensor. In cases
where there is little or no system pressure, the standard injection check valve supplied with each pump can be modified
with a high-pressure spring that adds approximately 50PSI. Contact your local distributor to order Part Number E90375.
Operation
Turn on the pump. The PosiFlow will not output a signal until the pump actually is moving liquid. For priming of the
pump, the PosiFlow sensor may have to be turned one-half or one full turn counter-clockwise to vent the air out the vent
connection.
CAUTION: Tubing should always be connected from the air vent back into the supply tank or suitable drain. Do not
submerge the air return tubing below the chemical surface level in the supply tank.
Once fully primed, close the air vent by tightening down the PosiFlow Sensor. The green LED on the PosiFlow should
now begin to flash OFF with each pump stroke and it will output a pulse simultaneously. The PosiFlow will continue to
output pulses as long as it senses pressure changes. If the pump loses prime or is air locked, the compression of air with
each pump stroke will not generate enough pressure for the PosiFlow to sense and it will not output pulses. Similarly,
excessive pressures will cause the pump to begin to stall and not generate enough pressure differential for the PosiFlow
to sense.
NOTE: If there is any trapped air, long lengths of flexible tubing, or any area of relief or expansion during a
deadhead situation, the PosiFlow may still keep sensing flow. This is more likely in C15/20 sized pumps.
Output Signal
The output signal from the PosiFlow simulates a contact closure. It can be used in any device that takes a non-powered
contact closure as a switch. It can also be used to drive another EH series pump.
The Walchem Web
WebWeb
WebMaster is designed to accept the PosiFlow Sensor signal. It both powers and reads the signal.
The Web
WebWeb
WebMaster will alarm a no flow condition after a customer-set time limit and will also totalize pumped volume
once the pump is calibrated.
In a similar manner, a PLC or other device can be programmed to interpret the PosiFlow signal in any number of ways.
TROUBLESHOOTING & MAINTENANCE
CAUTION: Do NOT disassemble the PosiFlow Sensor! There are no user serviceable parts inside and the components
can easily be damaged. Disassembly of the Sensor will void the manufacturer’s warranty.
Problem Possible Cause Corrective Action
No output pulses from the
PosiFlow Sensor
(Output pulse is not synchronous
with the pump stroke)
oIncorrect or disconnected wiring
oPump has lost prime (air-lock)
oPump is drawing in air
oDisconnected or cut tubing
oPinched or clogged tubing
oForeign material lodged in pump
oPosiFlow Sensor not tightened
down (in the air vent condition)
oNot enough back pressure
oDouble check wire connections
oUnscrew PosiFlow one-half turn to
vent air
oCheck the tubing and fitting
connections
oCorrect/Replace the tubing
oCorrect/Replace Tubing
oDisassemble/inspect/clean pump
head assembly
oTighten the PosiFlow Sensor by
turning clockwise by hand.
oCheck Injection Valve location and
system pressure. Install the High
Pressure Spring (E90375) if there
is little or no system pressure
Liquid Leakage from around the
PosiFlow Sensor
oO-ring seals on PosiFlow missing,
dislocated, or worn
oRe-install or replace the o-rings
E00098 PosiFlow Sensor
Instruction Manual
REVISION A 4/10/02