VWR Gel Imager Mode d’emploi

Gel Imager and Gel Imager Plus
Installation and User Instructions
81-0317-01 Rev A
Gel Imager

Gel Imagers 2
System Introduction
Gel Imagers and Gel Imagers Plus
The Gel Imagers enables simple documentation of fluorescent and non-fluorescent gels,
membranes, blots, film plates and assays with the ability to save images to a USB storage device for later
quantitative analysis or enhancement for publication.
Images are saved in 8-bit TIFF or JPEG formats allowing the images to be read by most PC or MAC
programs. The Gel Imager system is a cost effective solution for archiving quality images in a compact
standalone package. No computer is required.
The models comes in two configurations:
Gel Imager Plus, a high resolution 1.3 megapixel camera, 8-48mm zoom lens or
Gel Imager - a 0.3 megapixel camera and a 5.7-34.2mm zoom lens.
System Components
Components
Refer to the packing slip and pictured components for parts included with the system.
Gel Imager Plus
8” LCD touch screen
with tilt capability
White light and darkroom
p
ower switches
Wide access
door
USB drive
Filter tray (located under the
camera) with EtBr filter
Main s
y
stem
p
ower switch
View port
Transilluminator
Camera and zoom Lens
Stylus pen
Side access door (Gel
Imager Plus models only)

Gel Imagers 3
System Specifications
Configurations
Part No. Description
89131-380 Gel Imager, 0.3 Camera, 20 x 20cm Transilluminator, 302nm UV, 115V
89131-382 Gel Imager, 0.3 Camera, 20 x 20cm Transilluminator, 302nm UV, 230V
89131-384 Gel Imager, 0.3 Camera, 21 x 26cm Transilluminator, 302nm UV, 115V
89131-386 Gel Imager, 0.3 Camera, 21 x 26cm Transilluminator, 302nm UV, 230V
89131-388 Gel Imager, 0.3 Camera, 20 x 20cm Transilluminator, 302/365nm UV, 115V
89131-390 Gel Imager, 0.3 Camera, 20 x 20cm Transilluminator, 302/365nm UV, 230V
89131-392 Gel Imager, 0.3 Camera, 21 x 26cm Transilluminator, 302/365nm UV, 115V
89131-394 Gel Imager, 0.3 Camera, 21 x 26cm Transilluminator, 302/365nm UV, 230V
89131-396 Gel Imager Plus, 1.3 Camera, 20 x 20cm Transilluminator, 302nm UV, 115V
89131-398 Gel Imager Plus, 1.3 Camera, 20 x 20cm Transilluminator, 302nm UV, 230V
89131-400 Gel Imager Plus, 1.3 Camera, 21 x 26cm Transilluminator, 302nm UV, 115V
89131-402 Gel Imager Plus, 1.3 Camera, 21 x 26cm Transilluminator, 302nm UV, 230V
89131-404 Gel Imager Plus, 1.3 Camera, 20 x 20cm Transilluminator, 302/365nm UV, 115V
89131-406 Gel Imager Plus, 1.3 Camera, 20 x 20cm Transilluminator, 302/365nm UV, 230V
89131-408 Gel Imager Plus, 1.3 Camera, 21 x 26cm Transilluminator, 302/365nm UV, 115V
89131-410 Gel Imager Plus, 1.3 Camera, 21 x 26cm Transilluminator, 302/365nm UV, 230V
Camera
System will be configured with one of the two camera
options: 1.3mpx (high resolution) or 0.3mpx .
Both cameras use a USB 2.0 PC connection. All camera
settings are factory pre-set for optimum performance when
viewing gels, films, or membranes under low light level
conditions. UVP Technical Support should be contacted
before making any adjustments to the camera settings.
Lens
The Gel Imager Plus with the 1.3mpx camera is equipped
with an 8-48mm zoom lens fitted with a close-up diopter and
step-up ring. The camera can then be held in a fixed
position. The diopter is for focusing on objects at the focus
length of the Gel Imager.
The Gel Imager with the 0.3mpx camera is equipped with a
5.7-34.2mm zoom lens fitted with a close-up diopter and
step-up ring. The camera can then be held in a fixed
position. The diopter is for focusing on objects at the focus

Gel Imagers 4
length of the Gel Imager.
Processor
Internal Storage: 75GB or more
USB 2.0 ports: 1 front, 4 in the back
USB storage device capacity: 1GB (minimum)
Software: Gel Imager software
Ethidium Bromide Filter (EtBr)
The Ethidium Bromide (EtBr) Filter (50mm sq.) UV blocking
band pass interference filter blocks UV and IR radiation
emitted from the transilluminator. The filter is placed in the
filter tray below the camera assembly. The filter allows
visualization of fluorophores from 580-630nm, targeting the
EtBr emission which peaks at 605nm. The EtBr filter can be
substituted for other specific fluorophore filters or removed
when imaging non-fluorescent media (protein gels, colony
plates, etc.) in order to produce brighter images.
Darkroom
The darkroom features:
Gel viewer window
Overhead white light
Wide access doors with UV safety switch
Transilluminator
A transilluminator is included with the Gel Imager. VWR offers a large variety of transilluminators, from
models with 8-watt, single wavelength and single intensity to models with multiple wavelengths and high/low
intensities.
Note: For UV protection, there is a 10-minute timer installed in the Gel Imager. The timer shuts off the
transilluminator after 10 minutes of operation. A switch on the front of the darkroom allows the user to reset
the timer. Opening the darkroom door automatically resets the timer.
LCD Touch Screen
The display is an 8-inch VGA color touch screen connected to the darkroom cabinet with tilt adjustment. The
touch screen allows the user to preview, snap, save and print images, as well as select certain preference
options, without the need of an external mouse or keyboard.
A stylus pen is included for increased pointer control for selecting software options on the touch screen.
Focus Target
The focus target fluoresces when placed on a transilluminator or when exposed to overhead UV. The target
provides sharp, fluorescent images to aid in adjusting the focus of the camera.
Removable USB Stick
The removable USB stick has 1 GB memory (minimum) included with the system which connects to the
system allowing transfer of images.

Gel Imagers 5
Printer
UV/White Converter Plate
Optional Equipment
Refer to the ordering information section of this manual for optional
equipment part numbers.
Thermal Printer
The thermal printer provides archive quality, 256 grayscale prints, and
five optional cost-effective print sizes.
Converter Plates
The UV/White Converter Plate allows imaging of non-fluorescent
stained media with an ultraviolet transilluminator. The converter plate
is specially coated to convert the 302nm UV output to white light,
rather than using a separate white light box.
The Visi-Blue Converter Plate converts UV to 460-470nm designed
for use with blue excitation samples and SYBR green, SYPRO orange
and GFP stains.
Set-up Instructions
Components
Camera with zoom lens, bracket, filter, cables
Gel Imager darkroom
USB storage device, 1 GB (minimum)
Transilluminator
WARNING: DO NOT ATTEMPT TO CONNECT ANY WIRING WHILE THE EQUIPMENT IS CONNECTED
TO ANY POWER SUPPLY
CAUTION
Do not install the system in places with high moisture, dust, or high temperature. Do not use any oil or
petroleum based cleaner for the cabinet. Use only mild soap or detergent solution for cleaning. Ensure that
the system is turned OFF during cleaning. Keep the equipment away from motors or other large magnetic
equipment apparatus.

Gel Imagers 6
Installing the Filter
Systems include a 50mm sq Ethidium Bromide (EtBr) filter. To install the filter in the Gel Imager darkroom:
Carefully remove the filter from the protective
plastic case, holding the filter at the edges to
prevent fingerprints.
The filter tray is located under the camera
bracket on the right side top of the darkroom.
Slide the tray open.
Place the filter inside the filter tray so that the
text on the edge of the filter is positioned so
the letters are right side up when facing you.
Additional filters are available. Refer to Accessories for ordering information.
Camera Set-up
The camera and zoom lens are assembled at the
factory.
Note: The zoon lens shipped may appear
different than the one pictured at right.
Remove the cap from the lens.
Attach the step up ring and diopter to the
lens. The step up ring and diopter will only
fit one way.
Using the four thumb screws provided,
secure the bracket to the base.
Slide the camera and lens assembly into the
camera bracket. Slide the lens through the
center hole in the gasket. The gasket forms
a seal around the lens.
Camera and lens
assembl
y
Camera cable
Thumb screws
secure the bracket
to the base
Ste
p
u
p
rin
g
Dio
p
te
r

Gel Imagers 7
Insert and tighten the knob/washer in the
central hole in the camera to secure the
camera to the bracket.
Plug the camera cable into the top the
camera and the other end into a UBS port on
back of the darkroom.
Connect the jumper cord from the darkroom
to the transilluminator.
Connect the main power cord from the back
of the darkroom to a surge protector power
outlet.
Printer Set-up
The thermal printer is an optional accessory. Proceed to the next section if a printer was not included with
the shipment.
Stylus and Holder
The screen is touch sensitive. Use the stylus included with the system to operate the software controls. For
easy access to the stylus when operating the system, remove the tape from the back of the holder and
adhere it to the Gel Imager. Slide the stylus into the holder.
Connection
for jumper cord
to the
transilluminator
Connection for the
darkroom power
USB port

Gel Imagers 8
Using the System
Powering Up the System
Power up the system by pushing once on the
MAIN POWER button located on the right side
of the unit. All internal components will now
have power.
Turn on the transilluminator and darkroom
power switches.
To power down the system, push the main
power button.
Adjusting the Screen Angle
To adjust the angle of the screen, loosen
the knobs on each side of the monitor.
Twist each knob counter-clockwise (Note
that the knobs will be turning opposite
directions.)
Tilt the screen to the desired angle.
Turn each knob clockwise to tighten.

Gel Imagers 9
Operating the Gel Imager Software
The system will go through the boot-up process. When ready, the LCD will display the Gel Imager software
screen similar to the one below:
To shut down the Gel Imager software, touch the X on the bottom left corner of the screen.
To turn the system off, press the MAIN POWER button or power down from the Windows icon on the bottom
right corner of the screen.
Refer to the Gel Imager Software User Manual for further instructions on using the software.
Using the Transilluminator
Once the Gel Imager darkroom power is turned on, power is supplied to all components including the jumper
power cable that supplies power to the transilluminator. To use the transilluminator, turn the transilluminator
power on.
For UV protection, there is a 10-minute timer installed in the system. The timer shuts off the transilluminator
after 10 minutes of operation. A switch on the front of the darkroom allows the user to reset the timer.
Opening the darkroom door will automatically reset the timer. When the timer runs out, press the timer again.
This prevents the transilluminator from being on for too long and prevents damage to gels.
Refer to the Transilluminator Manual for additional instructions on using the transilluminator.

Gel Imagers 10
Using the Gel Viewing Window
The Gel Viewer has a pressure-sensitive clasp. Press firmly to open the viewer. The glass is UV blocking
while providing a clear view to the transilluminator surface for visibility of samples without opening the door.
Image Focusing
Prior to capturing any images, prepare the image focus capabilities of the system:
Turn on the transilluminator and place a fluorescent gel on the transilluminator surface. NOTE: The Gel
Imager darkroom has a UV safety switch that turns the transilluminator off when the door is open. After
closing the door, be sure the overhead white light switch is turned off.
While watching the LCD screen, rotate the lens f-stop adjustment so that the image is bright enough to
see on the screen.
Rotate the lens focus adjustment on the lens. Adjust this so that the image appears in clear focus on the
LCD monitor. Note: Once the proper zoom range is set, the lens will have to be refocused when
zoomed in completely.
Rotate the zoom lens adjustment on the lens, so that the image is as big as possible. Readjust the
focus ring on the lens, making the image clear. Adjust the zoom so that the object of interest is within
the picture on the LCD monitor.
Ce manuel convient aux modèles suivants
17
Table des matières
Manuels Imprimante tout-en-un populaires d'autres marques

Triumph Adler
Triumph Adler DC 2016 Manuel

Epson
Epson WorkForce WF-2510 Guide de l'utilisateur

Epson
Epson Stylus Scan 2000 Manuel d'installation et d'exploitation

Epson
Epson PictureMate Snap PM 240 Mode d’emploi

Konica Minolta
Konica Minolta bizhub C35 Manuel utilisateur

Brother
Brother MFC-J430w Manuel utilisateur