VICI DBS MISTRAL Manuel utilisateur

1
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
TECH SUPPORT
Europe, Asia, Africa, Australia & New Zealand:
techsupport@vicidbs.com
USA, Canada, Central & South America:
techsupportus[email protected]
Italy
service@vicidbs.com
VICI AG INTERNATIONAL
Parkstrasse 2
CH-6214 Schenkon
Switzerland
tel: + 41 41 925-6200
fax: + 41 41 925-6201
web: www.vicidbs.com
APPLICABLE MODELS
This manual is applicable to the following models:
• Mistral Gas Station
Version 2.00.0000

1
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
TABLE OF CONTENTS
1. VERSION HISTORY................................................................................................................................ 2
2. INTRODUCTION.................................................................................................................................... 3
2.1. Symbol Denitions...............................................................................................................................3
3. SAFETY.................................................................................................................................................. 4
3.1. Personal Safety Instructions.................................................................................................................4
3.2. Instrument Safety Instructions ............................................................................................................4
4. SPECIFICATIONS................................................................................................................................... 5
5. PRESENTATION OF DEVICE ................................................................................................................. 6
5.1. Front View............................................................................................................................................6
5.2. Back View............................................................................................................................................6
5.3. Lateral View.........................................................................................................................................7
5.4. Connections ........................................................................................................................................7
5.5. Power Supply Settings ........................................................................................................................8
5.6. Transport and Storage ........................................................................................................................8
6. SCHEMATICS......................................................................................................................................... 9
6.1. Fluidics.................................................................................................................................................9
6.2. Electrical............................................................................................................................................ 10
7. STARTING THE GENERATOR ......................................................................................................... 11-13
8. SOFTWARE.......................................................................................................................................... 14
8.1. Overview of the Menus.....................................................................................................................14
8.2. Parameters ........................................................................................................................................14
8.3. Diagnostic .........................................................................................................................................15
9. OPTIONAL EXTERNAL TANK.............................................................................................................. 15
9.1. Size of Tank .......................................................................................................................................15
9.1.1. Tank Size Calculation................................................................................................................16
9.2. Where to Get the Tank......................................................................................................................16
9.3. Recommended Set-Up......................................................................................................................16
10. ALARMS & PRE-ALARMS .................................................................................................................. 17
10.1. Pre-Alarms.......................................................................................................................................17
10.2. Alarms .............................................................................................................................................17
11. TROUBLESHOOTING......................................................................................................................... 18
12. MAINTENANCE................................................................................................................................. 19
12.1. 4000h Maintenance ......................................................................................................................... 19
12.2. 8000h Maintenance......................................................................................................................... 19
13. SPARE PARTS ............................................................................................................................... 19-20

2
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
1. VERSION HISTORY
VERSION DATE COMMENT/CHANGE
1.00.0100 01/10/2017 Layout and initial release, VICI DBS
1.01.0000 03/29/2019 Update
2.00.0000 07/13/2021 New layout, images and graphics, VICI Gig Harbor Group

3
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
2. INTRODUCTION
The VICI DBS® Mistral Gas Station includes an oil-free air compressor and delivers a continuous ow of
nitrogen with a purity of 99.5% at a rate of 25 L/min and simultaneously 35 L/min of ltered air.
The Mistral Gas Station uses pressure swing adsorption technology (PSA) to remove oxygen, carbon dioxide
and water from compressed air. The resulting pure nitrogen is ideal for laboratory applications like LC-MS and
other applications requiring inert gas.
2.1. Symbol Denitions
SYMBOL DESCRIPTION
General warning. Caution is necessary when operating the device or control
close to where the symbol is placed, or to indicate that the current situation
needs operator awareness or operator action in order to avoid undesirable
consequences.
The manual must be consulted in all cases where this symbol is marked on
the device.
Risk of asphyxia. Caution is necessary when operating the device in a not
sufciently ventilated area.
Oxidant material. Caution is necessary due to the presence of oxidant material.
WEEE symbol. When discard this product, it must be sent to separate collection
facilities for recovery and recycling.
Fuse. Notify replaceable fuse place.
OR
OR
OR

4
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
3. SAFETY
3.1. Personal Safety Instructions
When using your Mistral Gas Station, basic safety precautions should always be followed, including
the following:
• The quantity of dust in the ambient Air can be very different from one place to another.
We recommend to replace the dust lters every 4000H. In some dusty areas, this
maintenance need to be performed more frequently.
• Clogged dust lter may damage the device or give a bad Nitrogen quality.
• The power supply main cable connector is the disconnect device in case of
emergency, it must be easily accessible at any time.
• If the Mistral Gas Station is used in a way not specied by VICI DBS, the
protection provided by the equipment could be compromised.
3.2. Instrument Safety Instructions
• The Mistral Gas Station is a nitrogen generator. Do not let it run with the outlet open to
the air in a low volume room.
• The increase of the Nitrogen (and decrease of oxygen) may cause human disease.
It is the responsibility of the user to check if the room is appropriate or if an O2
measurement is necessary.
• The Mistral Gas Station rejects oxygen as waste.
• Oxygen ratio in the air could change around the device in a low volume or not
sufcient ventilated room.
• If fuel gases are present around, their explosive or ammable limits could be
affected. It is the responsibility of the user to check if the room is appropriate or
if an O2 measurement is necessary.

5
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
4. SPECIFICATIONS
MISTRAL GAS STATION
Nitrogen ow rate L/min 25 @ 5.5 barg (80 psig)
Air 1 - ow rate L/min 35 @ 5.5 barg (80 psig)
Nitrogen purity >99.5%
Dewpoint oC (oF) -40 (-40)
Outlet pressure barg (psig) 7 (100)
Technology Carbon molecular sieve
Warm up time (minutes) 20
Electrical supply 110-120V 60 Hz / 220-240V 50Hz
Power consumption (watts) 2700
Noise level 55 dBA @ 1 meter
Dimensions mm (in) 482W x 641H x 1235D (18.9W x 25H x 48D)
Weight kg (lbs) 110 (243)
Shipping dimensions mm (in) 550W x 800H x 940D (21.6W x 31.4H x 37D)
Shipping weight kg (lbs) 141 (311)
Operating temp oC (oF) 15 to 35 (59 to 95)
Outlet connection 2 x 6mm OD compression or 1/4”
Certication CE, FCC, MET (UL and CSA compliant)
MODELS & SPECS

6
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
5. PRESENTATION OF THE DEVICE
5.1. Front View
Command Buttons (x3)LED Indicator Lights
Backlit Display
5.2. Back View
Relay Alarm
Outlet Compressor
Box Cooling System
Dust Silencer Filter
Outlet Water
Purge
Outlet N2
with Valve
Air Power Supply
and Fuses
Nitrogen Power
Supply and Fuses
Outlet Air
with Valve
Fuse Holder F3

7
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
Line Pressure
Flow Set
Pressure
Air Outlet
Pressure
5.3. Lateral View
5.4. Connections
Power Supply: C-20 Outlet N2 with Valve: G1/4” female
Outlet Water Purge: Ø10mm tube Relay Alarm: 2 pos universal Mate-N-Lock
connector female
N2 Outlet
Pressure
Service Box

8
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
5.5. Power Supply Settings
The current power supply setting of the generator is indicated on the ID Sticker at the back of the generator.
The MISTRAL GAS Station standard setting is 210-230V /50-60Hz.
If other voltages are required, this should be done when ordering. In this case a transformer will be installed
to transform internally the requested voltage to 110-115 V /50-60Hz. Below you will nd the list with possible
voltage- and frequency settings. In any case the voltage of the power supplied must stay within +/- 10% of the
nominal value.
VOLTAGE SETTING FREQUENCY FUSE F3 (5X20 OR 6.3 x 32 mm) COUNTRY
110-120 VAC 60 Hz 20 A Japan/U.S.A.
220-240 VAC 50 Hz 10 A Korea/Europe
Please see service manual for the connection of the transformer. General earth connection is done via power
supply main cable.
5.6. Transport and Storage Conditions
Temperature Minimum 4ºC (39ºF), Maximum 45ºC (113ºF)
Humidity Inferior to 90% Rh, non-condensing
Shocks and vibrations during transport Shocks <1g; Vibrations <0.2g/5Hz
Instruments position Horizontal
The above mentioned values are considered for an instrument transported in its original new packing.
The device contains castors to be moved easily. Use the breaks to make the device stable. The Mistral Gas
Station could be elevated from any points under the chassis base.
Recommendations for Transport
When the instrument is shipped back to the factory for recalibration or repair, the Mistral Gas Station should
be packed in its original packing with the inlets and outlets protected against any accidental penetration of
dust or particles of the package.

9
MISTRAL GAS STATION
USER MANUAL www.vicidbs.com
Version 2.00.0000
6.1. Fluidics
6. SCHEMATICS
System:
Schematic no:
Date:
Created:
Checked:
Released:
THIS DOCUMENT AND THE INFORMATION WHICH
IT CONTAINS SHALL NOT BE USED, EXPLOITED
OR SOLD, AND SHALL NOT BE REVEALED OR
DISCLOSED TO OTHERS WITHOUT THE EXPRESS
WRITTEN PERMISSION OF VICI. THIS DOCUMENT
SHALL REMAIN THE PROPERTY OF VICI AND
SHALL BE RETURNED UPON DEMAND.
R
Rev.:
Sheet:
By:
Description:
ECN no:
Mistral Evolution Gas Station
air flow diagram 11.06.21 SBU
1
1/1
Table des matières
Autres manuels VICI DBS Générateur portable

VICI DBS
VICI DBS NM Plus 100 FID Tower Manuel utilisateur

VICI DBS
VICI DBS HK PLUS Series Manuel utilisateur

VICI DBS
VICI DBS PG Plus Rack 100 Manuel utilisateur

VICI DBS
VICI DBS MISTRAL EVOLUTION 10 Manuel utilisateur

VICI DBS
VICI DBS GT ULTRA ZERO AIR GENERATOR Manuel utilisateur

VICI DBS
VICI DBS MISTRAL EVOLUTION HYBRID Manuel utilisateur

VICI DBS
VICI DBS NM PLUS Manuel utilisateur

VICI DBS
VICI DBS HK PLUS Series Manuel d'instructions

VICI DBS
VICI DBS NM PLUS Series Manuel utilisateur

VICI DBS