HI
-The blood alcohol content above the display range.
-Display range: 0.200%BAC
Important notes
•Before testing, wait at least 15 minutes after your last drinking.
•The blowing time must be enough until beeping stops.
•It may show ERR (error) if none, interrupt or shorter blowing.
•If the device changes temperature suddenly, wait some minutes to use.
•
Only use the device in a temperature 10°C -40°C or 50°F -104°F.
•Recalibration interval: 12 months
•Never store the device in a corrosive gas room.
•Never store the device under sun exposure area.
•Avoid dropping or subjecting the device to a strong shock.
Specifications
•Electron-chemical sensor (Fuel Cell Sensor)
•Blowing time: 6 seconds
•Blowing flow: 10L/min
•Display range: 0.000-2.000‰BAC
•Show “HI” if above the display range
•Accuracy: +/-0.05‰at 0.5‰
•Memory: 5 records
•5 mouthpieces included
•Warm up time: 15 seconds
•Testing Time: 10 seconds
•Repeat test: less than 60 seconds.
•Audible sound alarm if above 0.50‰BAC
•4 digits display with status and backlight
•Power supply: 9V battery
•Battery for standard testing is about 800 times
•Low battery indication and auto power off function
•
Operation: 10°C -40°C or 50°F -104°F.
•Store: -10ºC ~ 60ºC or 14°F-140°F.
•Size: 115x60x23mm
•Recalibration interval suggest: 12 months
•Blowing interrupt detection
Calibration and Record
•You should have your alcohol tester calibrated every 12 months. Failure to do so may cause your
readings to be inaccurate.